Induztr
Engineering Digest Document Viewer Related Models

MVP 003-2 DC — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum MVP 003-015 series oil-free diaphragm vacuum pumps, covering six variants: MVP 003-2 DC, MVP 003-2 DC Vario, MVP 010-3 DC Neo, MVP 015-4 DC Neo, MVP 015-4 AC (also available with 120 V AC 60 Hz), and MVP 015-2 AC. Includes accessories/options such as RS-485/USB interface cables, gas-ballast valve kits, purge gas nozzle, OmniControl, handheld programming unit, silencer G1/8", service/overhaul kits, variable speed drive, and cleanroom option.

Section-by-Section Technical Breakdown

pp. 1
Product Features, Accessories and Options

The leaflet highlights oil- and particle-free vacuum for process safety, compact footprint for easy system integration, and low noise/vibration. A list of available accessories, spare parts and options is provided, including interface cables, gas-ballast kits, controllers (OmniControl), and a cleanroom option.

  • Oil- and particle-free vacuum ensures contamination-free, stable process conditions
  • Compact footprint and smart DC features simplify integration into instruments and vacuum systems
  • Low noise and low vibration for lab work and stable analytical conditions
  • Accessories: connecting cable, RS-485 interface cable, USB RS-485 converter, handheld programming unit
  • Options: gas-ballast valve kit / optional gas-ballast valve, purge gas nozzle with filter, flange and hose connection, silencer G1/8"
  • Control options: OmniControl with or without integrated power supply, variable speed drive
  • Maintenance: service kit, overhaul kit; cleanroom option available
pp. 2-3
Technical Data (Performance, Electrical, Interface)

A per-variant specification table covers pumping speed, ultimate pressure, gas ballast, rotational speeds (50/60 Hz), pressure limits, supply voltage, current, I/O interface, cooling, ambient temperature limits and certification. DC Neo variants carry RS-485 I/O and cTUVus certification (UL/CSA 61010-1); AC variants run on mains voltages with forced-air cooling.

  • MVP 003-2 DC / Vario: 0.25 m³/h pumping speed, 7.0 hPa ultimate pressure, 24 V DC ±5%, max 0.5 A, 3000 min⁻¹ (Vario: 1000–3000 min⁻¹), Vario has 0–5 V DC analog I/O
  • MVP 010-3 DC Neo: 0.6 m³/h, 1.0 hPa ultimate pressure, gas ballast (DC Neo single) / no gas ballast (second listing), 540–3000 min⁻¹, 24 V DC ±10%, max 2.5 A, RS-485
  • MVP 015-4 DC Neo: 0.8 m³/h, 0.5 hPa, gas ballast optional, 750–1800 min⁻¹, 24 V DC ±10%, max 6.8 A, RS-485
  • MVP 015-4 AC: 0.55 m³/h (50 Hz), 0.5 hPa, 230 V AC (-10/+5%) 50 Hz, max 1.1 A, forced-air cooling; also available with 120 V AC 60 Hz
  • MVP 015-2 DC Neo: 1.0 m³/h, 3.5 hPa, 750–2600 min⁻¹, 24 V DC ±10%, max 4.6 A, RS-485, gas ballast standard
  • MVP 015-2 AC: 0.50 m³/h (50 Hz) / 0.70 m³/h (60 Hz), 3.5 hPa, 100–115 / 208–236 V AC ±10% 50/60 Hz, max 1.1/0.6 A
  • Max intake pressure: 1000 hPa (MVP 003) / 1100 hPa (others); max exhaust pressure 1300 hPa (MVP 003/010) or 1100 hPa (MVP 015)
  • All variants: natural convection (DC models) or forced-air (AC models), ambient temperature 5–40 °C
  • DC Neo and MVP 015-4 DC Neo variants: cTUVus certified to UL/CSA 61010-1
pp. 2-4
Mechanical Data: Sound Level, Weight, Dimensions, Connections

Acoustic, mass and envelope data plus gas connections are given for each variant. All variants use G 1/8" gas inlet/outlet with silencer. Weights range from 0.7 kg (MVP 003-2 DC) to 7.5 kg (MVP 015-4 AC).

  • Sound pressure levels: 48 dB(A) (MVP 003-2 DC/Vario), 50 dB(A) (MVP 010-3 DC Neo), 49 dB(A) (MVP 015 DC Neo models), 52 dB(A) (AC variants)
  • Weights: 0.7 kg (MVP 003-2 DC/Vario), 2.0 kg (MVP 010-3 DC Neo), 4.4 kg (MVP 015-4 DC Neo), 7.5 kg (MVP 015-4 AC), 2.7 kg (MVP 015-2 DC Neo), 6.5 kg (MVP 015-2 AC)
  • Dimensions: 116×116×62 mm (MVP 003-2 DC/Vario), 173×105×103 mm (MVP 010-3 DC Neo), 260×140×130 mm (MVP 015-4 DC Neo), 332×145×154 mm (MVP 015-4 AC), 171×140×130 mm (MVP 015-2 DC Neo), 287×163×154 mm (MVP 015-2 AC)
  • Gas inlet/outlet on all models: G 1/8" / G 1/8" with silencer
  • A dimensional drawing and pumping speed curves are provided (p.2)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
All MVP 003-015 pumps are oil- and particle-free diaphragm pumps p.3 ↗ No contamination of process or pumped media; no oil management/oil capacity needed in system design whole series
Pumping speed ranges from 0.25 m³/h (MVP 003-2 DC) to 1.0 m³/h (MVP 015-2 DC Neo) p.3 ↗ Series is suitable only for small-volume evacuation/backing duty, not high-throughput applications whole series
Ultimate pressure: 7.0 hPa (MVP 003-2 DC), 1.0 hPa (MVP 010-3 DC Neo), 0.5 hPa (MVP 015-4), 3.5 hPa (MVP 015-2) p.3 ↗ Selection must match required base pressure; MVP 015-4 variants give the lowest ultimate pressure (0.5 hPa) whole series, per variant
DC variants run on 24 V DC: ±5% tolerance for MVP 003-2 DC/Vario, ±10% for Neo models p.3 ↗ Power supply must meet the tighter ±5% window on MVP 003-2 DC/Vario; Neo DC models allow wider supply tolerance DC variants
Maximum current draw: 0.5 A (MVP 003-2 DC), 2.5 A (MVP 010-3 DC Neo), 6.8 A (MVP 015-4 DC Neo), 4.6 A (MVP 015-2 DC Neo) p.3 ↗ Size 24 V DC supply and cabling for worst-case inrush/current, especially 6.8 A on MVP 015-4 DC Neo DC variants
AC variants: MVP 015-4 AC runs on 230 V AC (-10/+5%) 50 Hz (120 V AC 60 Hz version available); MVP 015-2 AC on 100–115 / 208–236 V AC ±10%, 50/60 Hz p.3 ↗ Direct mains connection possible for AC models; verify supply voltage band before ordering MVP 015-4 AC, MVP 015-2 AC
MVP 010-3 DC Neo and MVP 015-x DC Neo feature RS-485 I/O; MVP 003-2 DC Vario has 0–5 V DC analog I/O; MVP 003-2 DC and AC variants have no I/O interface p.3 ↗ Remote speed/process control is only possible via RS-485 or analog variants; USB RS-485 converter accessory enables PC integration MVP 010-3 DC Neo, MVP 015-x DC Neo, MVP 003-2 DC Vario
Variable speed control ranges: MVP 010-3 DC Neo 540–3000 min⁻¹, MVP 015-4 DC Neo 750–1800 min⁻¹, MVP 015-2 DC Neo 750–2600 min⁻¹, MVP 003-2 DC Vario 1000–3000 min⁻¹ p.3 ↗ Pumping throughput and standby behavior can be throttled via speed control on Vario/Neo DC models Vario and DC Neo variants
Maximum intake pressure 1000 hPa for MVP 003-2 DC/Vario and 1100 hPa for all other variants; exhaust back-pressure limit 1300 hPa (MVP 003/010) or 1100 hPa (MVP 015) p.3 ↗ Do not block or over-pressurize the exhaust line; MVP 015 variants tolerate less exhaust back-pressure whole series
Ambient operating temperature range 5–40 °C for all variants; cooling by natural convection (DC models) or forced air (AC models) p.3 ↗ Installation must allow free convection airflow around DC pumps; forced-air AC models need fan clearance whole series
Gas ballast: standard on MVP 003-2 DC/Vario, MVP 010-3 DC Neo (one listing) and MVP 015-2; optional on MVP 015-4; absent on one MVP 010-3 DC Neo listing p.3 ↗ For water-vapor-laden processes select variants with gas ballast or order the optional gas-ballast valve kit whole series
Gas inlet/outlet on all models is G 1/8" with silencer; a G1/8" silencer is offered as an accessory p.3 ↗ Standardize plumbing on G 1/8" fittings; use the silencer for noise reduction at the exhaust whole series
Sound pressure levels 48–52 dB(A) depending on variant (AC models 52 dB(A), DC models 48–50 dB(A)) p.3 ↗ Suitable for benchtop/instrument integration in noise-sensitive lab environments without acoustic enclosures whole series
Weights range from 0.7 kg to 7.5 kg; smallest footprint 116×116×62 mm (MVP 003-2 DC/Vario), largest 332×145×154 mm (MVP 015-4 AC) p.3 ↗ All models can be mounted inside instruments or compact OEM enclosures; check mounting surface for up to 7.5 kg whole series
DC Neo and MVP 015-4 DC Neo variants are cTUVus certified according to UL/CSA 61010-1; MVP 015-4 AC carries no listed certification p.3 ↗ For North American safety compliance, prefer the certified DC Neo variants or verify certification status of AC models MVP 010-3 DC Neo, MVP 015-x DC Neo
AC 60 Hz versions of MVP 015-4 AC have slightly higher pumping speed values; MVP 015-2 AC delivers 0.70 m³/h at 60 Hz vs 0.50 m³/h at 50 Hz p.3 ↗ Pumping performance depends on line frequency; check local mains frequency when predicting throughput AC variants
Service kit and overhaul kit accessories are offered, plus purge gas nozzle with filter and cleanroom option p.3 ↗ Plan preventive maintenance intervals using the service kit; overhaul kit supports full rebuilds; purge nozzle protects against corrosive/particulate process media whole series
OmniControl (with or without integrated power supply) and a handheld programming unit are available as control accessories p.3 ↗ System integration can use Pfeiffer's own controller; the handheld unit allows parameter programming without a PC DC variants with I/O