HiPace 700 H with TC 400 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiPace 700 H turbomolecular vacuum pump product leaflet. Covers a single model (HiPace 700 H) with its performance data, interface/mounting specifications, and the range of available accessories and options (adapter, mains/connection cables, fieldbus systems, power supply kit, display unit, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper).
Section-by-Section Technical Breakdown
Highlights three key characteristics: high performance with best compression ratio for light gases suited to high and ultra-high vacuum, reliable hybrid bearing technology for long life and maximum uptime, and quiet operation via patented laser balancing technology. Lists available accessories, spare parts, and options.
- Best compression ratio for light gases; suited for high and ultra-high vacuum applications
- Best-in-class hybrid bearing technology, long life cycle, maximum uptime
- Low noise and vibration levels thanks to patented laser balancing technology
- Accessories/options: electrical adapter, mains cable, connection cable, various fieldbus systems, power supply kit and display unit, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper
Pumping speeds, compression ratios, and gas throughputs at final rotational speed for N₂, H₂, He, and Ar, plus maximum forevacuum pressure. Shows very high compression ratios for N₂ and Ar and strong H₂ pumping (555 l/s).
- Pumping speed: N₂ 685 l/s, H₂ 555 l/s, He 655 l/s, Ar 665 l/s
- Compression ratio: N₂ 1·10¹³, H₂ 2·10⁷, He 1·10⁹, Ar 1·10¹³
- Gas throughput at final rotational speed: N₂ 3.5 hPa·l/s, H₂ 14 hPa·l/s, He 18 hPa·l/s, Ar 2 hPa·l/s
- Max. forevacuum pressure (N₂): 22 hPa
Standard communication interfaces, mounting orientation, flange connections, run-up time, sound pressure level, and weight. Confirms arbitrary mounting orientation and DN 160 ISO-K/ISO-F/CF-F inlet.
- Communication interfaces standard: RS-485, remote
- Mounting orientation: arbitrary
- Gas inlet / outlet: DN 160 ISO-K/ISO-F/CF-F / DN 25 ISO-KF / G 1/4"
- Run-up time: 2 min
- Sound pressure level (EN ISO 2151): 50 dBA
- Weight approx.: 11.8 kg
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed is 685 l/s for N₂ and 555 l/s for H₂. p.2 ↗ | Strong hydrogen pumping makes the pump suitable for UHV systems with H₂-dominated residual gas; size forepump accordingly. | HiPace 700 H |
| Compression ratio for H₂ is 2·10⁷ and for He 1·10⁹, versus 1·10¹³ for N₂ and Ar. p.2 ↗ | Even for light gases the compression is high enough that base pressure is largely determined by the forevacuum side; select a good forepump for UHV work. | HiPace 700 H |
| Maximum permissible forevacuum pressure for N₂ is 22 hPa. p.2 ↗ | Forepump and backing-line design must keep backing pressure below 22 hPa during operation, including during gas load events. | HiPace 700 H |
| Gas throughput at final rotational speed: 14 hPa·l/s for H₂ and 18 hPa·l/s for He, but only 3.5 hPa·l/s for N₂ and 2 hPa·l/s for Ar. p.2 ↗ | The pump tolerates much higher continuous gas loads of light gases than heavy gases; process gas composition strongly affects permissible throughput. | HiPace 700 H |
| Run-up time to final speed is 2 minutes. p.2 ↗ | Process sequencing and interlocks (e.g., opening the high-vacuum valve) should account for ~2 min start-up delay. | HiPace 700 H |
| Sound pressure level is 50 dBA per EN ISO 2151. p.2 ↗ | Low noise level permits installation near workstations without additional acoustic enclosure in most cases. | HiPace 700 H |
| Weight is approximately 11.8 kg. p.2 ↗ | Mounting flanges and chamber interfaces must support ~12 kg; relatively easy manual handling for installation. | HiPace 700 H |
| Gas inlet is DN 160 ISO-K/ISO-F/CF-F; outlet is DN 25 ISO-KF / G 1/4". p.3 ↗ | High-vacuum side requires DN 160 flange hardware; forevacuum side connects via DN 25 ISO-KF or a G 1/4" threaded port. | HiPace 700 H |
| Mounting orientation is arbitrary. p.3 ↗ | The pump can be installed in any orientation, simplifying integration into constrained system layouts (verify with full installation manual for oil/bearing details). | HiPace 700 H |
| Standard communication interfaces are RS-485 and remote. p.3 ↗ | Integration into control systems uses RS-485 as standard; fieldbus options are available as accessories for other protocols. | HiPace 700 H |
| Hybrid bearing technology with patented laser balancing is used. p.1 ↗ | Reduced vibration and long bearing life; a vibration damper accessory is available for vibration-sensitive applications. | HiPace 700 H |
| A power failure venting unit and venting accessories are offered. p.1 ↗ | For unattended or safety-critical systems, specify automatic/power-failure venting to protect the pump and process on shutdown. | HiPace 700 H (optional accessories) |
| Fieldbus systems, power supply kit, and display unit are available as options. p.1 ↗ | Non-RS-485 plant networks require ordering the appropriate fieldbus option and power supply/display kit for local monitoring. | HiPace 700 H (optional accessories) |