HiPace 700 Plus with TC 400 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiPace 700 Plus turbomolecular vacuum pump — a single model, presented as a 3-page product leaflet/brochure covering key performance data, mounting/interface details, and available accessories/options (no ordering codes or variants listed).
Section-by-Section Technical Breakdown
Introduces the HiPace 700 Plus turbomolecular pump, highlighting its low magnetic stray field for high-sensitivity applications such as high-resolution electron microscopes. Emphasizes reliability via hybrid bearing technology, and low noise/vibration from an embedded bearing with patented laser balancing.
- Low magnetic stray field — suited to high-resolution electron microscopes
- Best-in-class hybrid bearing technology for long life cycle and maximum uptime
- Low noise and vibration via embedded bearing plus patented laser balancing technology
Lists pumping speeds for N₂, H₂, He and Ar, compression ratios, gas throughput at final rotational speed, forevacuum limit, run-up time, sound pressure level and weight. Core performance figures for pump selection.
- Pumping speed N₂: 685 l/s; H₂: 555 l/s; He: 655 l/s; Ar: 665 l/s
- Compression ratio N₂: 1·10¹¹; H₂: 4·10⁵; He: 3·10⁷; Ar: 1·10¹¹
- Gas throughput at final speed: N₂ 6.5 hPa·l/s, H₂ 14 hPa·l/s, He 20 hPa·l/s, Ar 3.5 hPa·l/s
- Forevacuum max. (N₂): 11 hPa; Run-up time: 4 min
- Sound pressure level (EN ISO 2151): 50 dBA; Weight approx. 11.5 kg
Specifies the standard communication interfaces, permissible mounting orientation, and gas inlet/outlet flange dimensions. A short data block on page 3.
- Standard communication interfaces: RS-485, remote
- Mounting orientation: 0° (HV flange top)
- Gas inlet: DN 160 ISO-K/ISO-F/CF-F; gas outlet: DN 25 ISO-KF / G 1/4"
Lists the available accessories and options for the HiPace 700 Plus, covering electrical connectivity, fieldbus integration, venting and mounting hardware.
- Electrical adapter, mains cable, connecting cable
- Various fieldbus systems; power supply kit and display unit
- Mounting kit, safety valve, venting accessories, power failure venting unit
- Vibration damper
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed for N₂ is 685 l/s (H₂ 555 l/s, He 655 l/s, Ar 665 l/s). p.2 ↗ | Size backing pumps and conductance calculations around ~685 l/s effective N₂ speed at the inlet flange. | HiPace 700 Plus |
| Compression ratio for H₂ is only 4·10⁵, versus 1·10¹¹ for N₂ and Ar. p.2 ↗ | Ultimate pressure for light gases is strongly limited by forevacuum pressure; use a clean, low-pressure backing pump for low H₂ partial pressure. | HiPace 700 Plus |
| Maximum permissible forevacuum pressure for N₂ is 11 hPa. p.2 ↗ | Backing pump must maintain ≤11 hPa at maximum gas throughput to keep the turbopump operational. | HiPace 700 Plus |
| Run-up time to final rotational speed is 4 minutes. p.2 ↗ | Plan process start-up delays accordingly; also relevant for interlock logic gating process start on full speed. | HiPace 700 Plus |
| Gas throughput at final rotational speed ranges from 3.5 hPa·l/s (Ar) to 20 hPa·l/s (He). p.2 ↗ | Defines max continuous gas load per gas species; do not exceed with process gas dosing, else pump stalls or overheats. | HiPace 700 Plus |
| Sound pressure level is 50 dBA per EN ISO 2151. p.2 ↗ | Low enough for laboratory environments (e.g., electron microscopy rooms) without additional acoustic enclosures in most cases. | HiPace 700 Plus |
| Pump weight is approx. 11.5 kg. p.2 ↗ | Single-person handling feasible; check load ratings of mounting flanges/brackets accordingly. | HiPace 700 Plus |
| Gas inlet is DN 160 ISO-K/ISO-F/CF-F; outlet is DN 25 ISO-KF / G 1/4". p.3 ↗ | Select matching high-vacuum flange hardware (incl. CF for UHV/microscopy applications) and a DN 25-class foreline. | HiPace 700 Plus |
| Permissible mounting orientation is 0° only (HV flange top). p.3 ↗ | The pump must be installed vertically with the inlet flange up; no horizontal or inverted mounting permitted. | HiPace 700 Plus |
| Standard communication interfaces are RS-485 and remote. p.3 ↗ | Plan control/monitoring integration on RS-485; fieldbus variants available as options if other protocols are needed. | HiPace 700 Plus |
| Low magnetic stray field is a stated design feature. p.1 ↗ | Suitable for mounting near beam-path instruments (electron microscopes) where magnetic fields would degrade image resolution. | HiPace 700 Plus |
| Hybrid bearing technology with embedded bearing and patented laser balancing is used. p.1 ↗ | Reduced vibration levels support vibration-sensitive applications; a vibration damper accessory is available for further isolation. | HiPace 700 Plus |
| A power failure venting unit is offered as an accessory. p.1 ↗ | Recommended where sudden pump shutdown could contaminate the chamber; integrates automatic venting on power loss. | HiPace 700 Plus |