Induztr
Engineering Digest Document Viewer Related Models

HiPace 300 Plus — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiPace 300 Plus turbomolecular vacuum pump — a single-model 3-page brochure (15.09.2025). No ordering codes are given; it lists standard communication interfaces, mounting orientation, flange options, and a range of accessories/options (fieldbus systems, power supply kit, safety valve, venting accessories, vibration damper, etc.).

Section-by-Section Technical Breakdown

pp. 1
Product Highlights

Positions the HiPace 300 Plus as a turbomolecular pump with a low magnetic stray field, suited to high-sensitivity applications such as high-resolution electron microscopes. Reliability is attributed to hybrid bearing technology; quiet operation to an embedded bearing with patented laser balancing.

  • Low magnetic stray field — ideal for high-resolution electron microscopes
  • Best-in-class hybrid bearing technology, long life cycle, maximum uptime
  • Low noise and vibration via embedded bearing and patented laser balancing
pp. 2
Technical Data

Pumping speeds, compression ratios, gas throughputs, and general operating data for N₂, H₂, He, and Ar, plus mechanical/acoustic parameters. Key figures include 260 l/s N₂ pumping speed and 3.5 min run-up time.

  • Pumping speed: N₂ 260 l/s, H₂ 220 l/s, He 255 l/s, Ar 255 l/s
  • Compression ratio: N₂ 1·10¹¹, H₂ 9·10⁵, He 1·10⁸, Ar 1·10¹¹
  • Gas throughput at final rotational speed: N₂ 5 hPa·l/s, H₂ 14 hPa·l/s, He 8 hPa·l/s, Ar 2 hPa·l/s
  • Max. forevacuum pressure (N₂): 15 hPa
  • Run-up time: 3.5 min
  • Sound pressure level: 50 dBA (EN ISO 2151)
  • Weight: approx. 5.8 kg
pp. 1,3
Interfaces, Connections and Accessories

Standard communication interfaces and flange/vacuum connections are specified, along with the full accessories and options list. Mounting orientation is restricted to horizontal with the HV flange on top.

  • Standard interfaces: RS-485, remote
  • Mounting orientation: 0° (HV flange top)
  • Gas inlet: DN 100 ISO-K/ISO-F/CF-F; Gas outlet: DN 16 ISO-KF / G 1/4"
  • Accessories/options: electrical adapter, mains cable, connecting cable, various fieldbus systems, power supply kit and display unit, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed for N₂ is 260 l/s (H₂ 220 l/s, He 255 l/s, Ar 255 l/s). p.2 ↗ Size the forevacuum pump and process budget around ~260 l/s effective N₂ speed; H₂ performance is ~15% lower. HiPace 300 Plus
Compression ratio for N₂ is 1·10¹¹ and for H₂ only 9·10⁵. p.2 ↗ Ultimate pressure for light gases is limited by fore-vacuum pressure: with a 1 hPa fore-line, H₂ ultimate is ~10⁻⁶ hPa; plan H₂-sensitive applications accordingly. HiPace 300 Plus
Maximum permissible forevacuum pressure for N₂ is 15 hPa. p.2 ↗ The backing pump must maintain <15 hPa at full gas throughput; select a fore pump with adequate staging margin. HiPace 300 Plus
Gas throughput at final rotational speed: N₂ 5 hPa·l/s, H₂ 14 hPa·l/s, He 8 hPa·l/s, Ar 2 hPa·l/s. p.2 ↗ Defines the maximum continuous gas load before the pump decelerates; throttle process gas admission to stay below these values. HiPace 300 Plus
Run-up time to full speed is 3.5 min. p.2 ↗ Interlock processes or gas admission until ~3.5 min after start; do not admit full gas load during run-up. HiPace 300 Plus
Sound pressure level is 50 dBA measured to EN ISO 2151. p.2 ↗ Suitable for laboratory environments (e.g. electron microscopy rooms) without additional acoustic enclosures. HiPace 300 Plus
Pump weight is approx. 5.8 kg. p.2 ↗ Light enough for single-person mounting; verify chamber/flange load limits are not an issue. HiPace 300 Plus
Mounting orientation is restricted to 0°, i.e. HV flange on top (horizontal inlet flange up). p.3 ↗ Do not specify horizontal-shaft or inverted installations; system layout must place the high-vacuum flange facing upward. HiPace 300 Plus
Gas inlet flange options: DN 100 ISO-K, ISO-F, or CF-F; outlet is DN 16 ISO-KF or G 1/4". p.3 ↗ Choose CF-F variant for UHV/electron-microscope applications; foreline can use small ISO-KF hardware or a 1/4" threaded connection. HiPace 300 Plus
Standard communication interfaces are RS-485 and remote (contact/plug); fieldbus systems are optional accessories. p.3 ↗ For PROFIBUS/PROFINET/EtherCAT-style integration, order the appropriate fieldbus accessory; base RS-485 supports the Pfeiffer protocol. HiPace 300 Plus
A power failure venting unit is available as an accessory. p.1 ↗ Recommended where a sudden power loss could contaminate the pump or chamber; it vents the pump automatically on mains failure. HiPace 300 Plus (option)
Low magnetic stray field is a design feature of the HiPace 300 Plus. p.1 ↗ Can be mounted close to electron-optical columns and other magnetic-field-sensitive instruments without heavy magnetic shielding. HiPace 300 Plus
Bearing concept is a hybrid bearing (with embedded bearing design) and laser balancing for low vibration. p.1 ↗ Low vibration benefits high-resolution imaging; if further reduction is needed, the optional vibration damper can be fitted. HiPace 300 Plus
A safety valve and venting accessories are offered for the pump. p.1 ↗ Include the safety valve in the backing-line design to protect against oil backstreaming on power failure; specify venting valves for controlled venting cycles. HiPace 300 Plus (options)