HiPace 30 Neo — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Single model datasheet: Pfeiffer Vacuum HiPace 30 Neo turbomolecular vacuum pump, with a list of optional accessories (electrical adapter, mains cable, connecting cable, CIP adapter, power supply kit/display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper).
Section-by-Section Technical Breakdown
The HiPace 30 Neo is positioned as a compact turbomolecular pump with extended thermal operating range, long maintenance intervals, high compression in a compact design, and flexible integration into small analytical systems. It can be installed in any orientation. A range of accessories, spare parts and options is offered.
- Extended thermal operating range and long maintenance intervals
- Maximum compression in a compact design
- Easy integration even into small analytical systems; installation in any orientation
- Accessories/options: electrical adapter, mains cable, connecting cable, CIP adapter, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper
Key vacuum performance figures: pumping speeds for Ar, H2, He and N2, ultimate pressure, gas throughput at full rotational speed, compression ratios, run-up time, noise level and interfaces. Supply voltage is 24 V DC ±10%.
- Pumping speed: N2 32 l/s, Ar 31 l/s, He 30 l/s, H2 24 l/s
- Ultimate pressure: 1·10⁻⁷ hPa (mbar)
- Gas throughput at full rotational speed (short term): H2 20, Ar 3, He 12.5, N2 7.5 hPa·l/s
- Compression ratio: Ar 1·10¹¹, N2 1·10¹¹, He 2·10⁵, H2 1·10⁴
- Run-up time 65 s; sound pressure level ≤48 dB(A) (EN ISO 2151); protection class IP54 type 12
- I/O interfaces: RS485, remote; input voltage 24 V DC (±10%)
Physical and mounting specifications: hybrid bearing, 90,000 min⁻¹ rotational speed (±2%) with variable speed 50–100%, arbitrary mounting orientation, dimensions and weight, and inlet/outlet flange options.
- Hybrid bearing; max fore-vacuum pressure for N2: 20 hPa (mbar)
- Rotational speed 90,000 min⁻¹ (±2%), variable 50–100%
- Mounting orientation: arbitrary
- Dimensions Ø x H: 86 x 106.4 mm; weight 1.5 kg
- Gas inlet/outlet: DN40 / DN63 ISO-K / ISO-KF / CF-F / G 1/4" / DN16 / DN25
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Ultimate pressure is 1·10⁻⁷ hPa (mbar). p.2 ↗ | Sufficient for high-vacuum analytical instruments; process base pressure must be well above this to retain pumping margin. | HiPace 30 Neo |
| Pumping speeds: 32 l/s (N2), 31 l/s (Ar), 30 l/s (He), 24 l/s (H2). p.2 ↗ | Size conductance and chamber pump-down calculations; note the notable drop for hydrogen. | HiPace 30 Neo |
| Compression ratios: 1·10¹¹ for Ar and N2, 2·10⁵ for He, 1·10⁴ for H2. p.2 ↗ | Low H2 compression means hydrogen-bearing backstreaming/fore-vacuum pressure must be managed carefully; low fore-vacuum pressure needed for lowest ultimate pressures with H2. | HiPace 30 Neo |
| Maximum fore-vacuum pressure (N2) is 20 hPa. p.3 ↗ | Fore pump must maintain ≤20 hPa backing pressure; affects backing pump sizing. | HiPace 30 Neo |
| Gas throughput at full speed, short term: H2 20 hPa·l/s, He 12.5, N2 7.5, Ar 3 hPa·l/s. p.2 ↗ | Short-term throughput limits define permissible gas loads; do not exceed for continuous operation at full speed. | HiPace 30 Neo |
| Run-up time is 65 s. p.2 ↗ | Vent/process start sequences must allow ~65 s before full-speed operation; can be used as interlock threshold. | HiPace 30 Neo |
| Rotational speed 90,000 min⁻¹ (±2%), variable in the range 50–100%. p.3 ↗ | Speed can be reduced to 50% for process control or load limitation; speed monitoring tolerance is ±2%. | HiPace 30 Neo |
| Input voltage 24 V DC with ±10% tolerance (21.6–26.4 V DC). p.2 ↗ | Power supply design must hold this window; select PSU and cable losses accordingly. | HiPace 30 Neo |
| I/O interfaces: RS485 and remote. p.2 ↗ | RS485 enables digital control/monitoring integration; remote interface for hardwired control. | HiPace 30 Neo |
| Protection class IP54 type 12. p.2 ↗ | Suitable for dust-protected, splash-resistant cabinet environments; not for washdown or outdoor exposure. | HiPace 30 Neo |
| Sound pressure level ≤48 dB(A) (EN ISO 2151). p.2 ↗ | Quiet enough for laboratory/analytical environments without additional acoustic measures. | HiPace 30 Neo |
| Hybrid bearing type. p.3 ↗ | Oil-free hybrid bearing supports long maintenance intervals and clean vacuum; relevant for contamination-sensitive applications. | HiPace 30 Neo |
| Mounting orientation is arbitrary. p.3 ↗ | Can be integrated vertically or horizontally without derating, simplifying compact system layouts. | HiPace 30 Neo |
| Dimensions Ø86 x 106.4 mm, weight 1.5 kg. p.3 ↗ | Very compact and lightweight; mounting structures and vibration isolation (optional vibration damper) can be minimal. | HiPace 30 Neo |
| Gas inlet/outlet options: DN40, DN63 ISO-K/ISO-KF/CF-F, G 1/4", DN16, DN25. p.3 ↗ | Multiple flange options allow matching to existing chamber interfaces; select correct ordering variant for the desired connection. | HiPace 30 Neo |
| A power failure venting unit is available as an accessory. p.1 ↗ | Enables automatic pump venting on power loss, protecting the pump and process; recommended for unattended systems. | HiPace 30 Neo (option) |