Induztr
Engineering Digest Document Viewer Related Models

HiPace 30 Neo — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Single model datasheet: Pfeiffer Vacuum HiPace 30 Neo turbomolecular vacuum pump, with a list of optional accessories (electrical adapter, mains cable, connecting cable, CIP adapter, power supply kit/display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper).

Section-by-Section Technical Breakdown

pp. 1
Product Highlights and Accessories

The HiPace 30 Neo is positioned as a compact turbomolecular pump with extended thermal operating range, long maintenance intervals, high compression in a compact design, and flexible integration into small analytical systems. It can be installed in any orientation. A range of accessories, spare parts and options is offered.

  • Extended thermal operating range and long maintenance intervals
  • Maximum compression in a compact design
  • Easy integration even into small analytical systems; installation in any orientation
  • Accessories/options: electrical adapter, mains cable, connecting cable, CIP adapter, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper
pp. 2
Performance and Interface Data

Key vacuum performance figures: pumping speeds for Ar, H2, He and N2, ultimate pressure, gas throughput at full rotational speed, compression ratios, run-up time, noise level and interfaces. Supply voltage is 24 V DC ±10%.

  • Pumping speed: N2 32 l/s, Ar 31 l/s, He 30 l/s, H2 24 l/s
  • Ultimate pressure: 1·10⁻⁷ hPa (mbar)
  • Gas throughput at full rotational speed (short term): H2 20, Ar 3, He 12.5, N2 7.5 hPa·l/s
  • Compression ratio: Ar 1·10¹¹, N2 1·10¹¹, He 2·10⁵, H2 1·10⁴
  • Run-up time 65 s; sound pressure level ≤48 dB(A) (EN ISO 2151); protection class IP54 type 12
  • I/O interfaces: RS485, remote; input voltage 24 V DC (±10%)
pp. 3
Mechanical and Connection Data

Physical and mounting specifications: hybrid bearing, 90,000 min⁻¹ rotational speed (±2%) with variable speed 50–100%, arbitrary mounting orientation, dimensions and weight, and inlet/outlet flange options.

  • Hybrid bearing; max fore-vacuum pressure for N2: 20 hPa (mbar)
  • Rotational speed 90,000 min⁻¹ (±2%), variable 50–100%
  • Mounting orientation: arbitrary
  • Dimensions Ø x H: 86 x 106.4 mm; weight 1.5 kg
  • Gas inlet/outlet: DN40 / DN63 ISO-K / ISO-KF / CF-F / G 1/4" / DN16 / DN25

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Ultimate pressure is 1·10⁻⁷ hPa (mbar). p.2 ↗ Sufficient for high-vacuum analytical instruments; process base pressure must be well above this to retain pumping margin. HiPace 30 Neo
Pumping speeds: 32 l/s (N2), 31 l/s (Ar), 30 l/s (He), 24 l/s (H2). p.2 ↗ Size conductance and chamber pump-down calculations; note the notable drop for hydrogen. HiPace 30 Neo
Compression ratios: 1·10¹¹ for Ar and N2, 2·10⁵ for He, 1·10⁴ for H2. p.2 ↗ Low H2 compression means hydrogen-bearing backstreaming/fore-vacuum pressure must be managed carefully; low fore-vacuum pressure needed for lowest ultimate pressures with H2. HiPace 30 Neo
Maximum fore-vacuum pressure (N2) is 20 hPa. p.3 ↗ Fore pump must maintain ≤20 hPa backing pressure; affects backing pump sizing. HiPace 30 Neo
Gas throughput at full speed, short term: H2 20 hPa·l/s, He 12.5, N2 7.5, Ar 3 hPa·l/s. p.2 ↗ Short-term throughput limits define permissible gas loads; do not exceed for continuous operation at full speed. HiPace 30 Neo
Run-up time is 65 s. p.2 ↗ Vent/process start sequences must allow ~65 s before full-speed operation; can be used as interlock threshold. HiPace 30 Neo
Rotational speed 90,000 min⁻¹ (±2%), variable in the range 50–100%. p.3 ↗ Speed can be reduced to 50% for process control or load limitation; speed monitoring tolerance is ±2%. HiPace 30 Neo
Input voltage 24 V DC with ±10% tolerance (21.6–26.4 V DC). p.2 ↗ Power supply design must hold this window; select PSU and cable losses accordingly. HiPace 30 Neo
I/O interfaces: RS485 and remote. p.2 ↗ RS485 enables digital control/monitoring integration; remote interface for hardwired control. HiPace 30 Neo
Protection class IP54 type 12. p.2 ↗ Suitable for dust-protected, splash-resistant cabinet environments; not for washdown or outdoor exposure. HiPace 30 Neo
Sound pressure level ≤48 dB(A) (EN ISO 2151). p.2 ↗ Quiet enough for laboratory/analytical environments without additional acoustic measures. HiPace 30 Neo
Hybrid bearing type. p.3 ↗ Oil-free hybrid bearing supports long maintenance intervals and clean vacuum; relevant for contamination-sensitive applications. HiPace 30 Neo
Mounting orientation is arbitrary. p.3 ↗ Can be integrated vertically or horizontally without derating, simplifying compact system layouts. HiPace 30 Neo
Dimensions Ø86 x 106.4 mm, weight 1.5 kg. p.3 ↗ Very compact and lightweight; mounting structures and vibration isolation (optional vibration damper) can be minimal. HiPace 30 Neo
Gas inlet/outlet options: DN40, DN63 ISO-K/ISO-KF/CF-F, G 1/4", DN16, DN25. p.3 ↗ Multiple flange options allow matching to existing chamber interfaces; select correct ordering variant for the desired connection. HiPace 30 Neo
A power failure venting unit is available as an accessory. p.1 ↗ Enables automatic pump venting on power loss, protecting the pump and process; recommended for unattended systems. HiPace 30 Neo (option)