mo305
Technical Reference Summary
Covered Models & Document Scope
AXIMA Confidence MALDI mass spectrometer; linear and reflectron modes; positive and negative ion modes; MS/MS with precursor ion selection; sample targets: 96 or 384 well microtitre plate format, FlexiMass microscope slide-format targets (plain or 48 well), adaptors for unconventional layouts
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Laser repetition rate | 50 Hz p.2 ↗ | N2 laser |
| Laser type | N2 p.2 ↗ | |
| Resolution | >15,000 FWHM p.2 ↗ | ACTH fragment 7-38 |
| Ion source | MALDI p.2 ↗ | near-normal (on-axis) laser irradiation |
| Mass range | 55000-80000 m/z p.2 ↗ | linear mode, bovine serum albumin |
| Detector types | patented beam blanker p.2 ↗ | optional removal of unwanted low mass ions |
| Sample target format | 96 or 384 well p.2 ↗ | standard microtitre plate-format |
| Sample target format | plain or 48 well p.2 ↗ | FlexiMass microscope slide-format |