Technical Reference Summary

Covered Models & Document Scope

Shimadzu GC-2014 Gas Chromatograph; document covers the GC-2014 main unit with capillary and packed column capability, single or dual AFC, and optional detectors (FID, TCD, ECD, FPD, FTD) and injection units.

Technical Data

Parameter Value Condition
Column oven temperature range (Ambient + 10°C) ~ 400 °C p.13 ↗ using liquid CO2
Column oven temperature setting 1 °C steps p.13 ↗
Column oven capacity 15.8 L p.13 ↗
Column oven temperature accuracy Set value (K) ± 1% p.13 ↗
Column oven temperature variation coefficient 0.01 °C/°C p.13 ↗
Column oven temperature program steps Up to 20 steps p.13 ↗ cooling
Column oven total program time 9999.99 minutes max. p.13 ↗
Column oven linear heating rate 7 °C/min p.13 ↗ up to 400°C, 100V/120V
Column oven linear heating rate 15 °C/min p.13 ↗ up to 400°C, 200V/230V
Column oven cooling rate 300°C ~ 50°C in 6 min max. p.13 ↗ at 25°C ambient temperature
Detector temperature range 400 °C max. p.13 ↗ FID, TCD, ECD, FTD
Detector temperature range 350 °C max. p.13 ↗ FPD
Detector types FID, TCD, ECD, FPD, FTD p.13 ↗ for capillary/packed columns
Detector flow control system Dual flow rate differential system p.13 ↗
FID dynamic range 10^7 p.13 ↗
TCD dynamic range 10^5 p.13 ↗
ECD dynamic range 10^4 p.13 ↗ 63Ni 370 MBq radioactive source
FPD dynamic range P: 10^4 p.13 ↗
FTD dynamic range N: 10^3 p.13 ↗
Injection unit types Dual packed p.13 ↗
Flow rate setting range 0 ~ 100 mL/min p.13 ↗
Programmable flow steps 7 steps p.13 ↗
Programmed rate setting range -400 ~ 400 p.13 ↗
Pressure setting range 0 ~ 970 kPa p.13 ↗
Dimensions 250 (W) × 360 (H) × 175 (D) mm p.13 ↗

Operating Limits

  • Max detector temperature (FID, TCD, ECD, FTD): 400°C p.13 ↗
  • Max detector temperature (FPD): 350°C p.13 ↗
  • Max column oven temperature: 400°C p.13 ↗
  • Max pressure setting: 970 kPa p.13 ↗
  • Max total program time: 9999.99 minutes p.13 ↗