Technical Reference Summary

Covered Models & Document Scope

This manual (349-09011E, Apr. 2020) covers software installation for the Shimadzu Micro Vickers Hardness Tester HMV-G Series (Model HMV-G21 referenced). It applies to the HMV-G software on Windows 10 (64-bit edition) and covers optional configurations: electric XY stage, electric XYZ stage, motored micrometer, and high-resolution color camera models.

Section-by-Section Technical Breakdown

pp. 13-16
Software Installation

Describes installing the HMV-G software from the installation CD using setup.vbs, including handling of ShimadzuWordAddinTools and Word edition detection. The default installation folder is C:\Program files(x86)\Shimadzu\HMV-G, and language selection (English, Japanese, Chinese) is required.

  • Do not connect PC and HMV-G until software installation is finished.
  • Log in with an administrator account and insert the HMV-G installation CD.
  • Double-click setup.vbs (may appear as 'Setup' in Explorer).
  • ShimadzuWordAddinTools installation begins; click Yes for Office Desktop Apps, No for Office Store App or if Word 2010+ is not installed.
  • If HMV-G was previously installed, click 'Yes (Overwrite)' when prompted.
  • Default installation folder: C:\Program files(x86)\Shimadzu\HMV-G; can be changed via [Change].
  • Choose language (English, Japanese or Chinese) and click [Next].
  • If the USB driver installer for HMV-G is displayed, click [Next].
  • Restart the PC if prompted; HMV-G software is ready after restart.
pp. 17
Shortcut Setting

Explains how to set the HMV desktop shortcut to run as administrator via the Advanced properties of the shortcut, and warns not to use the Compatibility tab privilege level setting.

  • Set 'Run as administrator' in the Advanced Properties dialog of the shortcut.
  • Do not check 'Privilege Level: Run this program as an administrator' in the Compatibility tab.
pp. 18
Folder Configuration

Lists the default installation folder and the folders created by the HMV-G software for data, patterns, common configuration, and user-specific configuration, plus Start Menu and Desktop shortcuts.

  • Default installation folder: C:\Program Files(x86)\SHIMADZU\HMV-G
  • Data folder: X:\shimadzu\HMV-G\Data (X is application installation drive).
  • Pattern folder: X:\shimadzu\HMV-G\Pattern (default folder when pattern CSV is opened).
  • Common configuration file storage folder: X:\ProgramData\shimadzu\HMV-G\
  • User individual configuration file storage folder: X:\Users\<User name>\AppData\Local\shimadzu\HMV-G\
  • Start Menu shortcuts: HMV-G (Run as administrator), HMV-G (Standard user), Pattern Setting.
  • Desktop shortcut: HMV-G (Run as administrator).
pp. 19-20
Uninstallation and Program Modification

Covers uninstalling HMV-G via Control Panel and modifying or repairing the installation. Modify installs language resources and pattern Excel file without updating the registry; Repair reinstalls broken/deleted files and always overwrites the registry.

  • Uninstall: Start > Control Panel > Uninstall a program, select 'HMV-G', delete per instructions (administrator authority required).
  • Modify: installs Dell (RC.dll) for application and foreign language resource, and Excel file for pattern setting; configuration file not installed unless language-dependent; registry not updated.
  • Repair: installs broken or deleted files according to previously selected language; files with version information overwritten if installer version is larger; files without version information overwritten if installer date is newer; registry always overwritten.
  • After Modify or Repair, write device information to the registry according to the manual.
pp. 20
Machine Configuration Disc

Describes loading hardware preference and objective lens information from the machine configuration disc using the Save/Load Machine Configuration function.

  • HMV-G comes with an installation disc and a machine configuration disc.
  • Insert the machine configuration disc, start HMV-G, select [File]-[Save/Load Machine Configuration].
  • Click [OK] on the confirmation message, then click [Load machine configuration].
  • Select the drive with the disc and click [OK].
pp. 21-26
Camera Driver Installation

Details installing the Sentech camera driver (StCamSWare) from the HMV installation CD, including 64-bit and 32-bit executables, and configuring camera power settings via StPowerCam.

  • Do not connect PC and HMV-G until software installation is finished.
  • For Windows 10 or other 64-bit OS: run Driver\SENTECH\StCamSWare_x64.exe; for 32-bit OS: Driver\SENTECH\StCamSWare_x86.exe.
  • Select 'Complete' setup type; default folder C:\Program files\StCamSWare\.
  • If Windows security prompts, select 'Install this driver software anyway'.
  • Confirm StCamSWare appears in installed programs.
  • Run [Sentech] > [StPowerCam], uncheck all checkboxes, click 'Read', then 'Exit'.
pp. 27-31
Motion Driver Installation

Covers driver installation for optional electric XY/XYZ stage (CONTEC board) and motored micrometer (Melec UC-7660). The stage driver must be installed before inserting the board; the micrometer driver is installed via Device Manager and a registry file.

  • For electric XY/XYZ stage: run Driver\CONTEC\INF\WDM\SMC_ForWin10\Setup.exe (64-bit) or Driver\CONTEC\INF\WDM\SMC\Setup.exe (32-bit).
  • Install device driver before installing the board; turn PC off before inserting the board.
  • For motored micrometer: connect USB cable of control box, turn on control box, update driver for 'UC-7660' in Device Manager.
  • Driver folder: Windows10: Driver\Melec\64bit\Windows10; other OS 32-bit: Driver\Melec\32bit\Windows8; 64-bit: Driver\Melec\64bit\Windows8.
  • After driver installation, double-click MotoredMicroMeter_x64.reg (64-bit) or MotoredMicroMeter_x86.reg (32-bit) and confirm warnings.
pp. 33
Cable Connection and Operation Check

Describes connecting the PC and HMV-G with USB cables, connecting the stage control box if an electric XY stage is installed, powering on, and verifying that drivers install automatically and equipment works in the HMV-G software.

  • Connect PC and HMV-G with 2 USB cables; for high-resolution color camera models, the camera USB cable connects directly to HMV-G.
  • If optional electric XY stage is installed, connect motion control board in PC to stage control box with a cable.
  • Turn on PC and HMV-G; if electric XY stage or motored micrometer is installed, turn on the control box also.
  • When Windows is ready, drivers for HMV-G and USB camera install automatically.
  • Start HMV-G software and check that each equipment correctly works.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Correspondence OS is Windows 10 (64-bit edition). p.5 ↗ Install only on Windows 10 64-bit; change OS if different before installation. Whole HMV-G series
Do not connect PC and HMV-G until software installation is finished. p.13 ↗ Prevents driver conflicts and ensures correct enumeration during first connection. Whole HMV-G series
Default installation folder is C:\Program files(x86)\Shimadzu\HMV-G. p.16 ↗ Use this path unless a custom folder is specified via [Change]. Whole HMV-G series
Data folder default is X:\shimadzu\HMV-G\Data, where X is the application installation drive. p.18 ↗ Plan disk space and backup on the installation drive. Whole HMV-G series
Common configuration file storage folder is X:\ProgramData\shimadzu\HMV-G\. p.18 ↗ Shared settings are stored here; ensure write permissions for all users. Whole HMV-G series
User individual configuration file storage folder is X:\Users\<User name>\AppData\Local\shimadzu\HMV-G\. p.18 ↗ User-specific settings are per Windows profile. Whole HMV-G series
Camera driver executables: StCamSWare_x64.exe for Windows 10 or other 64-bit OS; StCamSWare_x86.exe for 32-bit OS. p.21 ↗ Select the correct executable matching the OS bitness. Whole HMV-G series
Default camera driver installation folder is C:\Program files\StCamSWare\. p.22 ↗ Use default unless a custom folder is selected via [Browse]. Whole HMV-G series
For electric XY/XYZ stage, install device driver before installing the board. p.27 ↗ Prevents hardware detection issues; turn PC off before inserting the board. Models with electric XY/XYZ stage
CONTEC stage driver path: Driver\CONTEC\INF\WDM\SMC_ForWin10\Setup.exe for Windows 10 or other 64-bit OS; Driver\CONTEC\INF\WDM\SMC\Setup.exe for other 32-bit OS. p.27 ↗ Use the correct path for the OS version. Models with electric XY/XYZ stage
Motored micrometer driver folder: Windows10: Driver\Melec\64bit\Windows10; other OS 32-bit: Driver\Melec\32bit\Windows8; 64-bit: Driver\Melec\64bit\Windows8. p.30 ↗ Select the folder matching the OS. Models with motored micrometer
Registry file for motored micrometer: MotoredMicroMeter_x64.reg for Windows 10 or other 64-bit OS; MotoredMicroMeter_x86.reg for 32-bit OS. p.31 ↗ Must be applied after driver installation to complete setup. Models with motored micrometer
Connect PC and HMV-G with 2 USB cables; for high-resolution color camera models, the camera USB cable connects directly to HMV-G. p.33 ↗ Verify cabling per model to ensure camera and communication function. Whole HMV-G series; high-resolution color camera models
If optional electric XY stage is installed, connect motion control board in PC to stage control box with a cable. p.33 ↗ Required for stage motion control. Models with electric XY stage
After Modify or Repair, the registry is always overwritten; write device information to the registry according to the manual. p.20 ↗ Device-specific settings must be re-entered after Modify/Repair. Whole HMV-G series
Set the HMV shortcut to 'Run as administrator' via Advanced properties; do not use Compatibility tab privilege level. p.17 ↗ Ensures proper elevation without conflicting settings. Whole HMV-G series
Local security policy settings: set 'User Account Control: Behavior of the elevation prompt for administrators in Admin Approval Mode' to 'Elevate without prompting' and 'User Account Control: Behavior of the elevation prompt for standard users' to 'Automatically deny elevation requests'. p.12 ↗ Omits elevation prompts when starting the program. Whole HMV-G series
Power management: choose [High performance] plan and disable [Turn on fast startup(recommended)]. p.8 ↗ Required for proper operation of the HMV-G software. Whole HMV-G series