Induztr
Engineering Digest Document Viewer Related Models

c10g-e097

Technical Reference Summary

Covered Models & Document Scope

Application brochure on oligonucleotide characterization and quality control using Shimadzu analytical instruments: MALDI-MS (MALDI-8030, MALDI-8020, MALDImini-1), LC-MS (LCMS-2050, LCMS-2020, LCMS-9050, LCMS-9030, LCMS-QTOF), Nexera XS inert / Nexera UC / Nexera Prep HPLC, and LabSolutions / LabSolutions MD / LabSolutions Insight Biologics software. No product model datasheet with ordering codes is included.

Technical Data

Parameter Value Condition
Flow rate 0.3 mL/min p.14 ↗ LCMS single quadrupole, 70-mer DNA analysis
Mobile phase A composition 95.4 mM HFIP, 7.1 mM TEA in water p.14 ↗ LCMS single quadrupole, 70-mer DNA analysis
Detection wavelength range 200-400 nm p.14 ↗ PDA detection, 70-mer DNA analysis
Ionization mode ESI/APCI (DUIS), Negative mode p.14 ↗ LCMS single quadrupole, 70-mer DNA analysis
Interface voltage -2.0 kV p.14 ↗ LCMS single quadrupole, 70-mer DNA analysis
Flow rate 0.3 mL/min p.16 ↗ LCMS single quadrupole, FLP impurity analysis
UV detection wavelength range 190-400 nm p.16 ↗ LCMS single quadrupole, FLP impurity analysis
Injection volume 6 µL p.16 ↗ LCMS single quadrupole, FLP impurity analysis
Scan mode Scan (m/z 550-2000) p.16 ↗ LCMS single quadrupole, FLP impurity analysis
Nebulizing gas flow 3.0 L/min p.16 ↗ LCMS single quadrupole, FLP impurity analysis
Mobile phase A composition 100 mmol/L HFIP, 10 mmol/L TEA in water p.18 ↗ Nexera XS inert, oligonucleotide impurity analysis
Column temperature 60 °C p.18 ↗ Nexera XS inert, oligonucleotide impurity analysis
Drying gas flow 5.0 L/min p.18 ↗ LCMS single quadrupole, oligonucleotide impurity analysis
Column temperature 60 °C p.13 ↗ Nexera XS inert, co-injection pretreatment
Co-injection volumes 1, 3, 5 µL p.13 ↗ Nexera XS inert, co-injection pretreatment
Column temperature 40 °C p.20 ↗ Reversed-phase ion-pair chromatography optimization
Flow rate 0.35 mL/min p.20 ↗ Reversed-phase ion-pair chromatography optimization