Technical Reference Summary

Covered Models & Document Scope

This manual covers the Shimadzu LC-30ADSF solvent delivery module (CO2 pump) for SFC/SFE systems. It describes the standard pump unit, its keypad/display operation, auxiliary and VP functions, time programs, gradient operation (master/slave pump units), external I/O, troubleshooting, maintenance, and specifications. Optional accessories mentioned include the liquid CO2 shutoff valve, automatic rinsing kit (standard), SFC-30A back-pressure regulator, FCV flow line switching valve, and degassing unit.

Section-by-Section Technical Breakdown

pp. 9-10
Configuration and Accessory Pack

Lists the standard components and accessory pack shipped with the LC-30ADSF, including part numbers and quantities. It also describes the instrument overview and key features such as built-in pump head cooling, high-pressure CO2 delivery up to 66 MPa, low pulsation, and automatic plunger rinsing.

  • Instrument: LC-30ADSF pump unit.
  • Accessory pack includes optical cable 070-92025-51, clamp 046-05522-02, spiral tube wrap 018-26020-02, pipe clip 037-60177-05, silicone tubing 228-25162-03, elbow tubing 228-44611-91, L-type tubing connector 035-61561-12, washing bottle 228-58181-42, refrigerant bottle 228-58181-41, bottle holder 228-51593, flow review jig 228-61054-41, RS-232C cable 070-03623-02.
  • Built-in pump head cooling mechanism for stable CO2 delivery.
  • Maximum pressure up to 66 MPa.
  • Discharge volume per plunger stroke: 10 µL for low pulsation.
pp. 11-15
Parts Identification and Function

Describes the front panel, rear panel connectors, display panel, and keypad. It explains the functions of the display indicators (flow, pressure, p.max, p.min, pump, prog.run, remote, G.E., status indicator) and the 24-key keypad.

  • Rear connectors: external I/O terminals, CO2 injection connector, PUMP PRESS (0–5 to 0–100 MPa in 20 steps), DGU PRESS, SFC, REMOTE, OPTION, power cord.
  • Status indicator: green = power ON, red = error, orange = sleep mode.
  • Pump indicator illuminates when pump runs; blinks when pump stops and liquid CO2 shutoff valve is open.
  • G.E. indicator blinks in Fast LC mode.
  • Keypad includes pump, run, valve, VP, sleep, delete, edit, clear, back, concentration, function, enter, numeric keys.
pp. 16-19
Preparation: Power, Pressure Check, Pump Head Cooling

Covers turning power ON/OFF, checking the pressure sensor zero before operation, and cooling the pump head. The pressure reading should be within -0.3 to 0.3 MPa with no plumbing connected; otherwise use ZERO ADJ.

  • Power ON: press power switch in; OFF: press again.
  • Before operation, remove all plumbing from pressure sensor and verify pressure display is -0.3 to 0.3 MPa.
  • If pressure is outside range, perform zero adjustment with [ZERO ADJ].
  • Pump head cooling requires connecting the RS-232C cable provided with LC-30ADSF to SFC-30A.
pp. 20-23
Basic Operation: Setting Parameters

Explains how to set flow rate, maximum pressure limit (p.max), and minimum pressure limit (p.min). It lists parameter ranges, steps, and initial values, and describes automatic pump stop and alarms when limits are exceeded.

  • Flow rate range: 0 to 5.0000 mL/min, step 0.0001 mL/min, initial 0 mL/min.
  • p.max range: 1.0–66.0 MPa for flow 0–3.0000 mL/min; 1.0–44.0 MPa for flow 3.0001–5.0000 mL/min; step 0.1 MPa; initial 10 MPa.
  • p.min range: 0.0–66.0 MPa for flow 0–3.0000 mL/min; 0.0–44.0 MPa for flow 3.0001–5.0000 mL/min; step 0.1 MPa; initial 0 MPa.
  • If pressure exceeds p.max, pumping stops and ERROR P-MAX alarm appears.
  • If pressure remains below p.min for more than 3 minutes, pumping stops and ERROR P-MIN alarm appears.
  • Setting p.min to 0 disables automatic stop on pressure drop.
pp. 25-26
Gradient Mode

Describes high-pressure gradient operation using two pump units (pump A master, pump B slave). It covers preparatory settings (SYS, ADRS), setting initial flow and concentration, and operation notes.

  • Two control methods: all pumps via system controller (Fast LC or Normal mode), or pump A controls pump B (Normal mode).
  • Set SYS parameter: pump A = 2, pump B = 1.
  • Set ADRS link address for both pumps (e.g., 85).
  • Set p.max and p.min individually for each pump; limits apply only to that pump.
  • Mobile phase B concentration range 0–100%, minimum setting unit 0.1%.
  • Do not press RUN on pump B during gradient operation controlled by pump A.
pp. 27-28
Automatic Rinsing Kit Operation

Provides the procedure for operating the automatic rinsing kit, including setting flow rate, starting the pump, checking rinse solution flow, and maintenance guidance.

  • Set flow rate to about 3 mL/min.
  • Press PUMP to start; pump indicator illuminates.
  • Check liquid flow from PTFE tubing at rinsing kit outlet.
  • Replace rinse solution with 10% 2-propanol solution once a day as a rough guide.
  • If air bubbles are released from the pump, CO2 may be leaking from a plunger seal; remove rinsing kit and drain rinse solution.
pp. 25-38
Auxiliary Functions

Lists and describes auxiliary function parameters grouped into Parameter Settings, File Operation, Control Settings, System Settings, and Monitor Settings. Includes settings such as EVENT output, compressibility compensation, P-SET, ZERO ADJ, MODE CHANG, LOCAL, ADRS, KEY CLOSE, BRIGHT, EXT-S, SYS, S-PROT, PRS-UNIT, RANGE, BEEP MODE, and monitoring functions.

  • Parameter Settings Group: EVENT, COMP.
  • File Operation Group: FILE NUM, FILE CPY, FILE DEL.
  • Control Settings Group: P-SET, ZERO ADJ, MODE CHANG.
  • System Settings Group: LOCAL, ADRS, KEY CLOSE, BRIGHT, EXT-S, SYS, S-PROT, PRS-UNIT, RANGE, BEEP MODE.
  • Monitor Settings Group: MON TIME, MON FLOW, MON ID, MON REV, CONDITION, DEGAS PRS.
pp. 39-53
VP Functions

Describes VP (validation/product) functions including mobile phase monitoring, product information, maintenance information, validation support, and calibration support. Covers serial number, software version, total operating time, seal delivered volumes, maintenance log, error log, operation log, date/time, memory check, pulse check, flow check, gradient test program, and CO2 leakage test.

  • Mobile Phase Monitor Group: MOBILE PHASE, ALARM LEVEL.
  • Product Information Group: SERIAL NUMBER, S/W ID.
  • Maintenance Information Group: TOTAL OP TIME, L SEAL DELIVERED, R SEAL DELIVERED, DGU OP TIME, PART REPLACEMENT, MAINTENANCE LOG, ERROR LOG, OPERATION LOG.
  • Validation Support Group: DATE, TIME, MEMORY CHECK, PULSE CHECK, FLOW CHECK, GE TEST PROGRAM, LEAKAGE TEST.
  • Calibration Support Group: Input PASSWORD, FLOW COMP FACT, PRESS COMP FACT, SEAL REPLACEMENT, FAST LC MODE, INITIALIZE PARAM, CHANGE PASSWORD, CBM PARAMETER.
pp. 57-66
Creating Time Program

Explains how to create, edit, delete, start, and stop time programs, including gradient programs and special commands STOP, LOOP, and GOTO.

  • Time program commands include flow, pressure, concentration, EVENT, etc.
  • Gradient programs can be created using BCNC command.
  • STOP command sets stop of time program.
  • LOOP command sets loop count.
  • GOTO command switches file.
  • RUN key starts/stops time program; inoperative if no program registered.
pp. 67-69
Connection to System Controller and External I/O

Covers preparation for connecting to a system controller and connection of external input/output terminals and event cable.

  • Connect to system controller via REMOTE connector.
  • External I/O terminals allow event signal input/output.
  • Event cable connection described in section 5.6.2.
pp. 70-74
Troubleshooting and Corrective Action

Provides tables of symptoms, probable causes, and corrective actions for common problems such as power failure, no pumping, unstable pumping, leaks, pressure issues, and gradient pressure differences. Also includes error message table with alarm types, causes, and actions.

  • Symptoms covered: power does not turn ON, key operation not possible, no liquid pumped, unstable pumping, leaks, flow rate below set value, retention time not reproducible, pressure does not rise, pressure rises too much, air bubbles in rinse solution, rinse solution volume decreases.
  • Error messages: ERROR P-MAX, ERROR P-MIN, ERROR HOME POS, ERROR EXTERNAL, WARN M-PHASE.
  • Corrective actions include cleaning/replacing check valves, replacing plunger seals, retightening male nuts, cleaning line filter, zero adjustment of pressure sensors, replacing CO2 cylinder if remaining <10 kg.
pp. 76-99
Maintenance

Describes periodic inspection and maintenance, replacement of plunger seal, cleaning and inspection of plungers and diaphragms, inspection and ultrasonic cleaning of check valves, inspection/replacement of line filter, exterior cleaning, and cleaning of automatic rinsing kit.

  • Periodic inspection list and check after maintenance.
  • Plunger seal replacement procedure including pump head removal and installation.
  • Cleaning/inspection of plungers and diaphragms.
  • Check valve inspection, ultrasonic bath cleaning, or replacement.
  • Line filter inspection/replacement.
  • Exterior cleaning and automatic rinsing kit cleaning.
pp. 100-102
Specifications

Provides technical specifications for the LC-30ADSF pump unit and gradient system, including pressure, flow rate, and control methods.

  • Maximum pressure: 66 MPa (for flow up to 3.0000 mL/min); 44 MPa (for flow 3.0001–5.0000 mL/min).
  • Flow rate range: 0 to 5.0000 mL/min.
  • Gradient system can be controlled by system controller or by pump unit.
  • Maintenance parts and consumable parts listed.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Maximum pressure limit p.max range is 1.0–66.0 MPa for flow rates 0–3.0000 mL/min and 1.0–44.0 MPa for flow rates 3.0001–5.0000 mL/min. p.20 ↗ When selecting operating pressure, ensure it does not exceed the limit corresponding to the set flow rate; higher flow rates reduce the allowable maximum pressure. LC-30ADSF
Minimum pressure limit p.min range is 0.0–66.0 MPa for flow rates 0–3.0000 mL/min and 0.0–44.0 MPa for flow rates 3.0001–5.0000 mL/min; setting 0 disables automatic stop on pressure drop. p.23 ↗ Set p.min to protect against leaks or CO2 depletion; setting 0 removes protection. LC-30ADSF
If pressure remains below p.min for more than 3 minutes, pumping stops automatically and ERROR P-MIN alarm occurs. p.23 ↗ The 3-minute delay prevents nuisance trips from short pressure fluctuations; ensure p.min is set appropriately. LC-30ADSF
Before operation, with no plumbing connected to the pressure sensor, the pressure display should read between -0.3 and 0.3 MPa. p.18 ↗ If outside this range, perform zero adjustment using [ZERO ADJ] to ensure accurate pressure readings. LC-30ADSF
The PUMP PRESS connector outputs voltage for external pressure monitoring, adjustable from 0–5 MPa up to 0–100 MPa in 20 steps. p.13 ↗ Select the appropriate range for the external device to match expected pressure levels. LC-30ADSF
In gradient mode, the two pump units may show pressure differences of up to 2.0 MPa or 5% as normal. p.71 ↗ Do not attempt to correct small pressure differences; only investigate if exceeding this tolerance. Gradient system with two LC-30ADSF units
For gradient operation controlled by pump A, set SYS parameter to 2 on pump A and 1 on pump B; set ADRS link address on both. p.25 ↗ Correct master/slave configuration is required for proper gradient control. Gradient system with two LC-30ADSF units
Mobile phase B concentration can be set from 0 to 100% with minimum setting unit 0.1%. p.26 ↗ Gradient precision is 0.1%; mobile phase A concentration is automatically 100% minus B concentration. LC-30ADSF in gradient mode
Replace the rinse solution in the automatic rinsing kit with 10% 2-propanol solution once a day as a rough guide. p.28 ↗ Regular replacement maintains rinsing effectiveness and extends plunger seal life. LC-30ADSF with automatic rinsing kit
If the remaining quantity of CO2 in the cylinder is less than 10 kg, replace the cylinder with a new one. p.71 ↗ Low CO2 can cause pressure fluctuations and unstable pumping; monitor cylinder weight. LC-30ADSF
The discharge volume per plunger stroke is 10 µL, and high-speed drive reduces flow pulsation and pulsation period. p.9 ↗ Provides precise delivery suitable for high-resolution separations. LC-30ADSF
The maximum allowable pressure of the instrument is 66 MPa; when connecting to other equipment, set the maximum pressure limit according to the pressure limit of all system components. p.22 ↗ Do not exceed the lowest pressure rating among all connected components. LC-30ADSF
The pump head cooling requires connection of the provided RS-232C cable to the SFC-30A back-pressure regulator. p.19 ↗ Ensure this cable is connected for proper temperature control of the pump head. LC-30ADSF
The status indicator shows green when power is ON, red when an error is generated, and orange during sleep mode. p.14 ↗ Use the indicator color for quick status diagnosis. LC-30ADSF
The pump indicator blinks when the pump stops and the liquid CO2 shutoff valve is open (valve optional). p.14 ↗ A blinking pump indicator with pump stopped indicates the shutoff valve is open; close it if not needed. LC-30ADSF with optional liquid CO2 shutoff valve
The G.E. indicator blinks when the analysis mode is set to Fast LC mode. p.14 ↗ Indicates Fast LC mode is active; ensure method compatibility. LC-30ADSF
The flow rate range is 0 to 5.0000 mL/min with a step of 0.0001 mL/min. p.20 ↗ Very fine flow adjustments are possible; ensure flow rate is not set to 0 when pumping is desired. LC-30ADSF
The pressure sensor zero adjustment range is -0.3 to 0.3 MPa; if outside, use ZERO ADJ. p.18 ↗ Regular zero checks ensure accurate pressure control and limit protection. LC-30ADSF

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Power does not turn ON even after switching ON power. p.78 ↗ Power plug is disconnected. Connect plug correctly. P. 2
Power does not turn ON even after switching ON power. p.78 ↗ Power cord internal wires are cut. Replace with a new cord of the same type. P. 2
Power does not turn ON even after switching ON power. p.78 ↗ Power supply does not meet specifications for this instrument. Use power supply that meets specifications for this instrument.
Key operation is not possible. p.78 ↗ DISPLAY was not pressed. Press DISPLAY. Operation keys are shown to operate. P. 7
No liquid is pumped. (The instrument does not run.) p.78 ↗ PUMP was not pressed. (Pump indicator does not illuminate.) Press PUMP. The pump indicator will illuminate. P. 16
No liquid is pumped. (The instrument does not run.) p.78 ↗ System controller did not send start signal. Have the system controller send a signal to start pumping. *1
No liquid is pumped. (The instrument does not run.) p.78 ↗ Flow rate is set to [0]. Set a flow rate other than 0. P. 13
No liquid is pumped. (The instrument does not run.) p.78 ↗ System controller set flow rate to 0. Have the system controller set a flow rate. *1
No liquid is pumped. (The instrument does not run.) p.78 ↗ Pump head cooling has stopped or cooling is instable. Press PUMP of SFC-30A to start cooling. Check that the refrigerant bottle is filled with 30% ethylene glycol.
No liquid is pumped. (The instrument does not run.) p.78 ↗ Error message (P.MAX, P.MIN etc.) appears. Press CLEAR to clear the message and take corrective actions for errors. P. 73
The instrument is running, but no liquid is pumped. p.78 ↗ Check valve is not working properly. Push in 2-propanol etc. from the inlet check valve. *2 *2: Connect the provided syringe tubing to the inlet check valve and push in 2-propanol etc.
Pumping is unstable, with large pump pulsation. p.79 ↗ Check valve is fouled. Push in 2-propanol etc. from the inlet check valve. *2; Pump 2-propanol to clean check valve. If cleaning check valve is not effective, clean the valve in an ultrasonic bath or replace it. P. 94
Pumping is unstable, with large pump pulsation. p.79 ↗ Liquid is leaking from gap between pump head and head holder, or from rinse flow lines. Replace the plunger seal. If carbon dioxide is still leaking after replacing the plunger seal, replace the plunger. P. 80, P. 88
Pumping is unstable, with large pump pulsation. p.79 ↗ Mobile phase is leaking from a flow line connection. Retighten the male nuts. If retightening the male nuts is not effective, replace the tubing, male nuts and ferrules.
Pumping is unstable, with large pump pulsation. p.79 ↗ Flow line is (partially) clogged. Clean the line filter in an ultrasonic bath or replace it. Identify the clogged parts and replace it. P. 96
Pumping is unstable, with large pump pulsation. p.79 ↗ Little CO2 remains in the cylinder. Measure the remaining quantity (weight) of CO2 in the cylinder and, if the remaining quantity is less than 10 kg, replace the cylinder with a new CO2 cylinder.
Flow rate is below set value. p.79 ↗ Check valve does not work properly. Pump 2-propanol to clean check valve. Otherwise, clean the valve in an ultrasonic bath or replace it. P. 94
Retention time cannot be reproduced. p.79 ↗ Check valve does not work properly. Pump 2-propanol to clean check valve. If cleaning check valve is not effective, replace the valve or clean it in an ultrasonic bath. P. 94
In a gradient system, the two pump units show different pressures. (Difference of up to 2.0 MPa or 5% is normal.) p.79 ↗ Zero points of the pressure sensors of two pump units are not the same. Use ZERO ADJ auxiliary function to execute zero adjustment of the pressure sensors. P. 30
In a gradient system, the two pump units show different pressures. (Difference of up to 2.0 MPa or 5% is normal.) p.79 ↗ Line filter of one of the pump units is clogged. Clean the line filter in an ultrasonic bath, or replace it. P. 96
In a gradient system, the two pump units show different pressures. (Difference of up to 2.0 MPa or 5% is normal.) p.79 ↗ Flow line(s) is/are clogged at the junction of the two pump unit flow lines. Identify the clogged lines and replace it.
Pressure does not rise. p.79 ↗ Liquid is leaking from flow line connections. Retighten the male nuts. If retightening the male nuts is not effective, replace the tubing, male nuts and ferrules.
Pressure rises too much. (Remove the column to check.) p.80 ↗ Line filter is clogged. Clean the line filter in an ultrasonic bath or replace it. P. 96
Pressure rises too much. (Remove the column to check.) p.80 ↗ Flow lines are clogged. Identify the clogged parts and replace them.
Pressure rises too much. (Remove the column to check.) p.80 ↗ Inner diameter of tubing is too small. Use tubing with the appropriate inner diameter.
Air bubbles leaked from the pump are mixed into the rinse solution in the automatic rinsing kit. p.80 ↗ Liquid is leaking from the plunger seal. Replace the plunger seal. If carbon dioxide is still leaking after replacing the plunger seal, replace the plunger. P. 88, P. 80
The volume of the automatic rinsing kit's rinse solution decreases. p.80 ↗ Rinse solution is leaking from the back of the head holder. Inspect (or replace) the diaphragm.
ERROR P-MAX p.82 ↗ Pump discharge pressure has exceeded the set maximum pressure limit, [P.MAX] or the working pumping pressure. Pumping will automatically stop, unless [S-PROT] function has been activated, in which case pumping will continue at half flow rate. Inspect for clogging in the flow line. If no clogging is found, set the appropriate [P.MAX] value or decrease the flow rate. "4.1 Setting Parameters" P. 12, "Setting System Protection [S-PROT]" P. 33, "4.1.3 Setting Maximum Pressure Limit" P. 14
ERROR P-MIN p.82 ↗ Pump discharge pressure has fallen below the set min. pressure limit, [P.MIN]. Pumping will automatically stop. Inspect for leakage in the flow line. If no leakage is found, set the appropriate [P.MIN] value. Note that [P.MIN] function is not activated for first one minute of pumping. "4.1.4 Setting Minimum Pressure Limit" P. 15
ERROR HOME POS p.82 ↗ The motor home position cannot be detected, the motor does not run, or the motor is slipping. Turn the power OFF and ON again, and press PUMP. If this message still appears, turn the power OFF and contact your Shimadzu representative.
ERROR EXTERNAL p.82 ↗ An error signal was sent from the external device connected to the external input/output terminal [ERR IN]. Inspect the external device, and eliminate the cause of the error.
WARN M-PHASE p.82 ↗ The remaining volume of mobile phase is less than the alarm level. Replace the mobile phase and reset the amount of mobile phase. "Showing/Setting Mobile Phase Volume [MOBILE PHASE]" P. 42