Engineering Digest Document Viewer Related Models

SFC-40P Instruction Manual - Back Pressure Regulator

Technical Reference Summary

Covered Models & Document Scope

This manual (228-92879) covers the SFC-40P Back Pressure Regulator (BPR) for Shimadzu Nexera UC Prep Series preparative supercritical fluid chromatography. It is intended for use with the LC-40P SF CO2 Solvent Delivery Module or similar. The document describes basic operation, accessories, and options; system-level details are in the Nexera UC Prep Series System Guide (228-92882).

Section-by-Section Technical Breakdown

pp. 8
Component Parts and Accessories

Lists the parts supplied with the SFC-40P, including the instrument, Safety Guideline, DVD-ROM, drain tubing, fittings, wrenches, joints, drain adapters, and stop plug. Used for unpacking verification.

  • SFC-40P main unit (no part number listed)
  • Safety Guideline booklet, part 228-92880
  • Instruction Manual / System Guide DVD-ROM, part 228-97175-41
  • DRAIN TUBE SI (L1000), part 228-25162-03
  • FITTING, L-TYPE 2005-10, part 035-61561-12
  • WRENCH, 13 x 17, part 086-03047-10 (qty 2)
  • STRAIGHT JOINT, part 228-28163
  • DRAIN ADAPTER, part 228-42204
  • DRAIN OUT, part 228-42205
  • DRAIN CTO, part 228-42206
  • STOP PLUG, PEEK, part 228-53071
pp. 9
Parts Identification: Front

Identifies front-panel features: operation area, front cover, and power button. The operation area is touch-sensitive and displays operation keys when its central part is touched.

  • Operation area: touch central part to display operation keys
  • Front cover: covers the flow line
  • Power button: turns power on/off
pp. 10
Parts Identification: Behind Front Cover

Describes internal components behind the front cover, including the BPR pressure sensor, BPR cover, BPR block, thermo-sensor, BPR 2nd heater inlet, makeup inlet, leak sensor, and BPR 1st heater.

  • BPR pressure sensor monitors pressure
  • BPR cover covers the BPR block and first heater
  • BPR block is the pressure control block
  • Thermo-sensor measures BPR 1st heater temperature
  • BPR 2nd heater inlet
  • Makeup inlet connects to makeup pump
  • Leak sensor detects liquid leaks
  • BPR 1st heater warms tubing from BPR block
pp. 12
Parts Identification: Back

Identifies rear-panel connectors and switches: REMOTE, external I/O terminals, COOLER, HEX, PC, power cord connector, and main power switch.

  • [REMOTE] connector for system controller
  • External input/output terminals for external equipment
  • [COOLER] connector for CO2 cooler
  • [HEX] connector for heat exchanger
  • [PC] connector for service personnel (normally unused)
  • Power cord connector
  • Main power switch: normally keep on; turn off if not used for a long time
pp. 13-15
Operation Panel: Display and Keys

Describes the operation panel display area, status LED, link LED, power button, direct keys, and keypad functions. The panel has arrow-key and numeric-key input modes switched by [Enter].

  • Status LED colors: Green = analysis ready; Red = error; Blue = analysis in progress; Yellow = preparation for analysis in progress; Orange = sleep
  • Link LED illuminates when controlled by system controller
  • Power on: press and hold power button at least 3 seconds
  • Power off: hold power button 3 seconds or more, then hold again 1 second on confirmation screen
  • Direct keys: [PRESS], [HEAT], [COOL], [HEX]
  • [CE] clears input, initializes screen, clears error messages and cancels alarms
  • [del] deletes input values
  • [Enter] confirms input values and switches input modes
  • Arrow keys move cursor or switch screens
  • Numeric keys [0]-[9] input values
pp. 16-17
Basic Operation Settings: BPR Pressure and Temperature

Provides procedures to set BPR pressure and BPR temperature via the FUNCTION > PARAMETER menu. BPR pressure range is 1.00 to 40.00 MPa (initial 10.00 MPa); BPR temperature range is 0 to 70 °C (initial 50 °C).

  • BPR pressure [BPR PRS]: initial 10.00 MPa, settable 1.00 to 40.00 MPa
  • BPR temperature [BPR TEMP]: initial 50 °C, settable 0 to 70 °C
  • Access: press [→] twice on initial screen, select PARAMETER, press [Enter]
  • Use numeric keypad and [Enter] to confirm; [CE] clears entered value
pp. 18
Operation: Starting Pressure and Temperature Control

Describes starting operation: verify temperature and pressure increase after starting heaters and CO2 pump, and stopping pressure control with [PRESS].

  • Check that displayed temperature starts increasing after starting temperature control
  • Check that displayed pressure starts increasing after CO2 pump starts pumping
  • Press [PRESS] to stop pressure control
pp. 19-20
Screen Types and Monitoring Screens

Describes the initial screen, monitor screen, auxiliary functions screen (FUNCTION), and VP functions screen. Monitoring screens display pump head temperatures, CO2 temperature, BPR pressure, and heat exchanger temperature.

  • Initial screen: press [↓] for monitoring screens; press [→] to select FUNCTION or VP
  • PUMP HEAD: displays right and left pump head temperatures
  • CO2: displays CO2 temperature drawn into CO2 pump and cooling block temperature
  • BPR PRS: displays BPR pressure; [PZ] and [MT] normally unused
  • HEX: displays heat exchanger temperature
  • When turning pump head cooling off and on again, wait until cooling block temperature reaches 5 °C or higher to avoid freezing in piping
pp. 21-30
Auxiliary Functions Screen (FUNCTION): Parameter, Control, System, Utility

Lists and describes the FUNCTION setting groups: PARAMETER (analysis parameters), CONTROL (instrument control), SYSTEM (system configuration), and UTILITY (display and buzzer). Includes setting ranges and initial values.

  • PARAMETER: BPR PRS (1.00-40.00 MPa, init 10.00), BPR TEMP (0.0-70.0 °C, init 50.0), HEX TEMP (0.0-70.0 °C, init 40.0)
  • CONTROL: FORCE PRESS OFF, ZERO ADJUST, BPR HOME, BPR BACK
  • SYSTEM: LOCAL MODE (0=Remote, 1=Local, init 0), LINK ADDRESS (1-12, init 4), USE COOL UNIT (0=Use, 1=Not Use, init 0), USE HEX (0=Use, 1=Not Use, init 1)
  • SYSTEM: BPR RDY CHECK (0=Off, 1=On, init 1), BPR RDY RNG (0.5-1.0 MPa, init 0.5), BPR WAIT TIME (0-10 min, init 1)
  • SYSTEM: COLUMN PROTECT (0=Off, 1=On, init 1), DEPRESS TIME (0.1-60.0 min, init 5.0)
  • SYSTEM: COOL RDY CHECK (0=Off, 1=On, init 1), COOL RDY RNG (1-10 °C, init 5), COOL W.TIME (0-10 min, init 5)
  • SYSTEM: PRESSURE UNIT (0=MPa, 1=kgf/cm2, 2=bar, 3=psi, init 0)
  • UTILITY: KEY CLOSE, BRIGHTNESS (1-4, init 4), BEEP MODE (0=On, 1=Alarm, 2=Off, init 0), VOLUME (1-3, init 2), DISP OFF TIME (0-10 min, init 1), DIRECT KEY MODE (0=Confirm, 1=Direct, init 0)
pp. 32-48
VP Functions Screen: Product Info, Maintenance, Validation, Calibration

Describes VP functions for product information, maintenance logs and usage timers, validation tests, and calibration settings. Includes password-protected calibration group.

  • PRODUCT INFO: serial numbers and software versions for instrument, CO2 cooler, CO2 pump, heat exchanger
  • MAINTENANCE: TOTAL OP TIME, BPR BLOCK USED, BPR DIAPHRAGM USED, COOL PUMP USED, PART REPLACEMENT, MAINTENANCE LOG (20 records), OPERATION LOG (10 records), ERROR LOG (10 records), COOL/HEX logs
  • VALIDATION: DATE, TIME, MEMORY CHECK, LEAK SENSOR CHECK, PULSE CHECK (acceptance criterion 0.05-1.00 MPa), BPR HEATER1 TEST, BPR HEATER2 TEST, PUMP COOLER TEST, HEX HEATER TEST
  • CALIBRATION: password (default 00000), BPR BLOCK replacement time (0-9999 h, init 2000), BPR DIAPHRAGM replacement time (0-9999 h, init 2000), COOLANT PUMP replacement time (0-9999 h, init 2000), COOL TEMP (0-10 °C), CANCEL COVER SW (0=detect, 1=do not detect), DATE FORMAT, INITIALIZE PARAM, PARAMETER LOCK, PASSWORD SETTING, CBM PARAMETER, LEAK THOLD, PRESS COMP FACTOR
pp. 44-45
Connection with System Controller or Workstation

Describes setting the instrument for connection to a system controller or workstation, including local/remote mode and link address. Also covers external input/output terminals and remote cable connection.

  • LOCAL MODE: 0=Remote (via system controller), 1=Local (independent)
  • LINK ADDRESS: 1 to 12
  • External I/O terminals on rear panel
  • Remote cable connection described in section 3.8.2
pp. 46-57
Maintenance: Periodic Inspection, Diaphragm Replacement, Filters, Cleaning

Covers periodic inspection and maintenance, removing the right panel cover and filter holder, replacing the BPR diaphragm (and BPR block), resetting usage time, replacing the air filter, cleaning the leak tray, and cleaning the exterior.

  • Periodic inspection and maintenance list
  • Removing right panel cover and filter holder
  • Replacing diaphragm (replacing BPR block)
  • Resetting usage time of diaphragm/BPR block
  • Replacing air filter
  • Cleaning leak tray
  • Cleaning exterior
pp. 58-59
Technical Information: Specifications and Maintenance Parts

Provides instrument specifications and lists consumable and replacement parts.

  • Specifications section
  • Consumable parts list
  • Replacement parts list

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
BPR pressure setting range is 1.00 to 40.00 MPa; initial value is 10.00 MPa. p.16 ↗ Select operating pressure within this range; default is 10.00 MPa. SFC-40P
BPR temperature setting range is 0 °C to 70 °C; initial value is 50 °C. p.17 ↗ Set BPR temperature within 0-70 °C to prevent freezing and maintain control. SFC-40P
Heat exchanger temperature setting range is 0 °C to 70 °C in increments of 1 °C; initial value is 40.0 °C. p.24 ↗ HEX temperature can be set in 1 °C steps within 0-70 °C. SFC-40P with heat exchanger
When turning pump head cooling off and then on again, wait until the cooling block temperature reaches 5 °C or higher to avoid freezing in the piping. p.20 ↗ Prevents ice formation and flow blockage in the CO2 pump head cooling circuit. SFC-40P with CO2 pump
BPR ready check range (BPR RDY RNG) is 0.5 to 1.0 MPa; initial value 0.5 MPa. A pressure within set pressure ± range is considered stable. p.27 ↗ Adjusts the tolerance for declaring BPR pressure ready. SFC-40P
BPR WAIT TIME is 0 to 10 min; initial value 1 min. If set to 0, READY is displayed unconditionally. p.27 ↗ Controls delay before READY after pressure reaches setpoint. SFC-40P
Column protection depressurizing time (DEPRESS TIME) is 0.1 to 60.0 min; initial value 5.0 min. p.28 ↗ Gradual depressurization time to protect columns after pressure control is turned off. SFC-40P
COOL RDY RNG is 1 to 10 °C; initial value 5 °C. COOL W.TIME is 0 to 10 min; initial value 5 min. p.28 ↗ Sets pump head temperature stability criteria for READY status. SFC-40P with CO2 pump
Pressure unit conversion: 1 kgf/cm2 = 0.098 MPa = 0.98 bar = 14.2 psi. p.29 ↗ Use for converting displayed pressure units. SFC-40P
PULSE CHECK acceptance criterion range is 0.05 to 1.00 MPa; measurement duration is 1 minute. p.41 ↗ Used to verify pressure control accuracy. SFC-40P
Estimated replacement time for BPR block, BPR diaphragm, and refrigerant pump is 2000 hours (initial value); settable 0 to 9999 hours. p.43 ↗ Maintenance scheduling based on usage time; reset after replacement. SFC-40P
COOL TEMP setting range for pump head temperature is 0 °C to 10 °C. p.43 ↗ Sets the CO2 pump head cooling temperature. SFC-40P with CO2 pump
Default calibration password is 00000 (five digits). p.42 ↗ Required to access CALIBRATION group settings. SFC-40P
CANCEL COVER SW setting: 0 = automatically detect BPR cover (initial), 1 = do not detect. Setting 1 keeps temperature control ON without cover, causing dangerous situation and possible burns. p.44 ↗ Do not change without special reason; safety hazard. SFC-40P
LINK ADDRESS range is 1 to 12; initial value 4. p.26 ↗ Sets channel number for system controller connection. SFC-40P
USE COOL UNIT: 0 = Use (initial), 1 = Not Use. USE HEX: 0 = Use, 1 = Not Use (initial). p.26 ↗ Must be set according to whether CO2 cooler and heat exchanger are connected. SFC-40P
Maintenance log stores 20 records; operation log and error log store 10 records each. p.36 ↗ Log capacity for tracking maintenance and errors. SFC-40P
Power on: press and hold power button for at least 3 seconds. Power off: hold 3 seconds or more, then hold again 1 second on confirmation screen. p.13 ↗ Correct power sequencing to avoid abrupt shutdown. SFC-40P

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
SENSOR NO GOOD p.41 ↗ Leak sensor detection level not adjusted properly. Adjust the detection level with [LEAK THOLD] function in the Calibration Support Group. 3.6.4 / p.41
PASSWORD WRONG p.45 ↗ Incorrect password input when changing password. Password has not been changed; re-enter correct password. 3.6.5 / p.45