Induztr
Engineering Digest Document Viewer Related Models

321-78451

Technical Reference Summary

Covered Models & Document Scope

WBC-102 Glass Windbreak, part number 321-78451, instruction manual dated Mar. 2023. Compatible only with small-pan models (capacity 1,020 g or less) of Shimadzu UP and UW/UX electronic balance series. The document covers the windbreak assembly, installation on balances with or without screw holes, operation, cleaning, transport, and maintenance parts.

Section-by-Section Technical Breakdown

pp. 5, 8, 9
Installation

Describes mounting the WBC-102 windbreak on compatible Shimadzu UP and UW/UX balances. The procedure covers removal of the pan and pan supporter caps, handling of optional screw holes and sealing stickers, fixing the windbreak holder with two knobs, and securing the windbreak with the sliding knob set to FIX.

  • Remove the pan and four pan supporter caps from the balance before installation.
  • If the balance has two screw holes plugged with rubber caps, remove the rubber caps.
  • If the balance has screw holes, peel off the two sealing stickers on the floor of the windbreak.
  • Fix the windbreak holder by tightening two fixing knobs against the two screw holes on the back of the balance.
  • Lower the windbreak onto the top of the balance and slide the sliding knob to the FIX position.
  • If the balance has screw holes on the top, open the side glass doors and tighten two fixing knobs to secure the windbreak to the balance top.
  • Return the four pan supporter caps and then the pan.
  • Check that the bubble of the level indicator at the rear of the balance is centered in the red circle; if not, perform level adjustment per the balance manual.
pp. 9
Operation

Covers basic use of the windbreak during weighing. All glass doors must be closed before connecting the AC adapter and turning power on, and before reading the display. The most convenient door may be opened when placing or removing samples.

  • Close all glass doors before connecting the AC adapter and turning the power ON.
  • Open the most convenient door when placing or removing the sample or container.
  • Close all doors before reading the display.
pp. 10, 11
Maintenance and Transport

Provides cleaning instructions, glass door removal procedure, transport guidance, and a list of maintenance parts. Cleaning must use a soft cloth with neutral detergent; organic detergents and chemicals must not be used. Glass doors can be removed for cleaning or replacement.

  • Clean by wiping with a soft cloth soaked in neutral detergent and wrung tightly.
  • Never use organic detergents, chemicals, or chemical wiping cloths, as they may damage the coating.
  • To remove a glass door: remove the pan and four pan supporter caps, unscrew the inside knob on the glass door, and slide the glass door out backwards.
  • When moving the balance by hand, do not hold it by the windbreak; hold the balance main body.
  • For other transport methods, dismount the windbreak from the balance and use the delivery box of each product.
  • Handle glass doors carefully to prevent breakage and injury; be careful not to touch the pan supporter shaft when removing the inside knob.
pp. 11
Maintenance Parts

Lists the available maintenance parts for the WBC-102 with their part numbers and remarks.

  • WINDBREAK HOLDER: S321-62779
  • SIDE DOOR ASSY (RIGHT): S321-62790-01, includes knob
  • SIDE DOOR ASSY (LEFT): S321-62790-02, includes knob
  • TOP DOOR ASSY: S321-62791, includes knob
  • KNOB: S321-62787, for door

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
WBC-102 can be used only with small-pan models (capacity 1,020 g or less) of UP and UW/UX series balances. p.4 ↗ Verify balance model and capacity before selecting the windbreak; larger-capacity or non-listed models are not supported. WBC-102
The windbreak is fixed to the balance back using two fixing knobs against two screw holes, if present. p.8 ↗ Balances without screw holes rely on the sliding knob and top fixing knobs; installation method depends on balance version. WBC-102
If the balance has screw holes, two sealing stickers on the floor of the windbreak must be peeled off. p.8 ↗ Sealing stickers must be removed to allow proper mounting when screw holes are present. WBC-102
If the balance has screw holes on the top, two fixing knobs are tightened to secure the windbreak to the balance top after opening the side glass doors. p.9 ↗ Top screw holes require additional fixing knobs; ensure the correct number of knobs (four fixing knobs supplied) are available. WBC-102
The sliding knob must be set to the FIX position to fix the windbreak to the balance. p.8 ↗ Confirm the sliding knob is in FIX position after installation to prevent movement. WBC-102
The bubble of the level indicator at the rear of the balance must be centered in the red circle; otherwise level adjustment is required per the balance manual. p.9 ↗ Leveling is mandatory after windbreak installation to maintain weighing accuracy. WBC-102
All glass doors must be closed before connecting the AC adapter and turning power ON, and before reading the display. p.9 ↗ Operating with doors open may affect weighing results; close all doors for stable readings. WBC-102
Cleaning must use a soft cloth soaked in neutral detergent and wrung tightly; organic detergents, chemicals, and chemical wiping cloths must never be used. p.10 ↗ Use only neutral detergent to avoid damaging the coating on the windbreak. WBC-102
Glass doors (two side doors and one top door) can be removed for cleaning and replacement by unscrewing the inside knob and sliding the door out backwards. p.10 ↗ Doors are serviceable; handle with care to avoid breakage and injury. WBC-102
When moving the balance by hand, do not hold it by the windbreak; hold the balance main body. p.10 ↗ Holding the windbreak during transport may cause damage or injury. WBC-102
For transport methods other than by hand, dismount the windbreak from the balance and use the delivery box of each product. p.10 ↗ Always separate the windbreak from the balance for non-manual transport. WBC-102
Maintenance parts: WINDBREAK HOLDER S321-62779; SIDE DOOR ASSY (RIGHT) S321-62790-01; SIDE DOOR ASSY (LEFT) S321-62790-02; TOP DOOR ASSY S321-62791; KNOB S321-62787. p.11 ↗ Use these exact part numbers when ordering replacements. WBC-102
The windbreak is a precision instrument; subjecting it to impacts could cause failure. p.6 ↗ Handle with care during installation and use to avoid damage. WBC-102
Do not use the windbreak outdoors or anywhere exposed to water. p.5 ↗ Indoor use only; avoid wet environments. WBC-102
Avoid locations with extreme temperature changes, vibration, direct sunlight, corrosive or flammable gases, dust, electromagnetic waves, or magnetic fields. p.6 ↗ Install in a stable, controlled environment to ensure proper performance. WBC-102
Install the balance on a strong and stable flat table or floor. p.6 ↗ An unstable site could lead to injury or trouble with the windbreak. WBC-102
If the windbreak has to be stored for a long time, store it in the packaging box in which it was delivered. p.6 ↗ Use original packaging for long-term storage to protect the instrument. WBC-102

Related models