Italian safety, installation, operation and maintenance instructions
Procedurepp. 43-72
German safety, installation, operation and maintenance instructions
Procedurepp. 73-99
French safety, installation, operation and maintenance instructions
Procedurepp. 100-126
Spanish safety, installation, operation and maintenance instructions
Procedurepp. 127-156
Chinese safety, installation, operation and maintenance instructions
Procedurepp. 157-183
Japanese safety, installation, operation and maintenance instructions
Procedurepp. 184-210
English safety, installation, operation and maintenance instructions
Specpp. 211-249
Technical data, graphs, dimensions and operating parameters
Orderingpp. 250-252
Accessories, spare parts and ordering information
pp. 253-257
Vacuum Products Division contact and legal information
p. 219table, 17 rows
p. 211table, 16 rows
p. 220table, 13 rows
p. 251table, 11 rows
p. 255table, 5 rows
p. 188table, 8 rows
p. 205table, 4 rows
p. 214table, 8 rows
p. 235table, 2 rows
p. 257table, 11 rows
Tab. 1 Technical Specification
Technical Specification
ISO 100/ CF 6" / ISO 160 / CF 8"
Pumping speed N2
250 l/s
Pumping speed He
255 l/s
Pumping speed H2
220 l/s
Pumping speed Ar
250 l/s
Max gas flow rate (N2):
250 sccm
Compression ratio N2
>1 × 1011
Compression ratio He
>1 × 108
Compression ratio H2
1.5 × 106
Compression ratio Ar
> 1 × 1011
Base pressure with recommended forepump
1 x 10-10 mbar (1 x 10-10 Torr)
Inlet flange
CFF 8" O.D. / ISO 160 / CFF 6" O.D. / ISO 100
Foreline flange
KF16 NW (KF25 – optional)
Max Rotation Speed
60600 rpm (1010 Hz driving frequency)
Start-up time
3 minutes (it can be longer in case of soft start active)
Recommended forepump
Mechanical pump: DS102, DS302 Dry Pump: IDP-3 (no gas flow), IDP-7, IDP10
Operating position
Any
Max foreline pressure tolerance N2
12 mbar
According to standard DIN 28 428, the base pressure is that measured in a leak-free test dome, 48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange and using the recommended pre-vacuum pump. Note: foreline Tolerance defined as the pressure at which the turbopump still produces a compression of 100. For continuous operation, water cooling is recommended (water temperature between 15°C and 20°C). Note: value refer to water-cooling pump version with: water temperature between 15°C and 20°C (non condensing, refer to Figure 13); backing pump with pumping speed equal or above 5 m3/h
Содержание
Стр.
Описание
213
Description of the TwisTorr 305 FS
219
Technical Specification
222
TwisTorr 305 FS Outline
233
Interconnections
233
Earth (Ground) Connection
234
Pump Operations
234
Inlet screen installation
237
Air Cooling Kit Installation
239
Water Cooling Kit Connection (only for water cooling pumps)
242
Vent Accessories
245
Pump Used with Corrosive and Oxidizing Gases
247
Purge Valve Installation
248
Pump Used in Presence of Magnetic Fields Vibration
248
Isolator Installation
249
Controller Operations
250
Accessories and Spare Parts
Technical Specification
Parameter
Value
Operating ambient temperature
+5 °C to +35 °C
Relative humidity of air
From graph in Figure 13 (non condensing) **
Bakeout temperature
ISO flange: 75°C at inlet flange max
CFF flange: 100°C at inlet flange max
Note: Measure a point close to the sealing element.
Lubricant
Permanent lubrication
Air cooling requirements
Natural convection (only with no gas load)
Forced air (5- 35 °C ambient temperature)
Coolant water
Minimum flow: 50 l/h (0.22 GPM)
Temperature: +15 °C to +30 °C
Max pressure: 5 bar (75 psi)
Noise Pressure level
41 dB(A) at 1 meter
Installation category
II
Pollution degree
2
Storage temperature
-40° C to +70° C
Weight kg (lbs): (*)
Pump ISO 100 K
Pump CFF 6’’
Pump ISO 160 K
Pump CFF 8’’
Note: versions with water cooling kit
ISO 100 K
5.84 (12.87)
CFF 6"
8.16 (17.98)
ISO 160 K
6.28 (13.84)
CFF 8"
10.43 (22.99)
Pwr supply :
Max input pwr:
Pump stand-by average pwr:
Pump max oper. pwr:
300 VA
10 W
150 W
Max operating altitude
3000 m
Note: mean values based on a significative sample (Ar and N2 compression ratio estimated); standard deviation per test: pumping speed: below ± 7%; noise pressure level: ± 10%” (only pump). The maximum magnetic field strengths allowed for Agilent turbo pumps are: • 100 Gauss (10 mT) in the axial direction • 50 Gauss (5 mT) in the transveral direction
Vent Valve N.O. 0,5 mm Orifice (for remote controller only)
9699844
Vent Valve N.O. 1,2 mm Orifice (for remote controller only)
9699845
Vent Valve N.C. 0,5 mm Orifice (for remote controller only)
9699847
Vent Valve N.C. 1,2 mm Orifice (for remote controller only)
9699846
5 m Turbopump extension cable (for remote controller only)
9699949
Mains Cable CHINA plug (3 m long)
8121-0723
The communication S/W between the controller and the pump (A-Plus) is available on the Agilent website: https://www.agilent.com/en/products/vacuum-technologies/vacuum-leak-detection-software/a-plus-software The connection is provided through serial communication or via RS232 SERIAL to B/T Adapter (X3514-68003).
RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION
EUROPE
NORTH AMERICA
PACIFIC RIM
Fax:
00 39 011 9979 330
1 781 860 9252
Fax Free:
00 800 345 345 00
Toll Free:
00 800 234 234 00
800 882 7426, Option 3
vpt-customercare@agilent.com
vpl-ra@agilent.com
Please visit our website for individual office information
http://www.agilent.com
Warning Symbols
Symbol
Description
Protective Ground Terminal
Do Not Touch
J1
Main connector
CE certification
CSA certification
Restriction of Hazardous Substances certification
RoHS China certification
Waste Electrical and Electronic Equipment
Символы нанесены на прибор; изображения символов в тексте не переданы.
Controller MPNs
Controller MPNs
Controller Descritpion
Ref.Instruction Manual
X3507-64130
Twistorr 305 FS onboard 24V
87-901-060-01
X3507-64131
Twistorr 305 FS onboard 110/230V
87-901-060-01
X3506-64130
Twistorr 305 FS AG Remote RS232/485
87-901-059-01
X3506-64131
Twistorr 305 FS AG Remote Profibus
87-901-059-01
The TwisTorr 305 FS models are:
Model
Description
X3513-64008
TWISTORR 305-K FS ISO100K STD AIR COOLING
X3513-64009
TWISTORR 305-K FS CFF6" STD AIR COOLING
X3513-64010
TWISTORR 305-K FS ISO160K STD AIR COOLING
X3513-64011
TWISTORR 305-K FS CFF8" STD AIR COOLING
X3513-64012
TWISTORR 305-K FS ISO100K STD WATER COOLING
X3513-64013
TWISTORR 305-K FS CFF6" STD WATER COOLING
X3513-64014
TWISTORR 305-K FS ISO160K STD WATER COOLING
X3513-64015
TWISTORR 305-K FS CFF8" STD WATER COOLING
Reduction in pumping speed (%)
Ar
N2
He
H2
X3500-68000
X3500-68001
20%
18%
7%
5%
Health and safety
Substances (please refer to MSDS forms)
Trade name
Chemical name
Chemical Symbol
CAS Number
Toxic
YES
NO
Harmful
YES
NO
Corrosive
YES
NO
Reactive
YES
NO
Flammable
YES
NO
Explosive (*)
YES
NO
Radioactive (*)
YES
NO
Biological (*)
YES
NO
Oxidizing
YES
NO
Sensitizer
YES
NO
Other dangerous substances
YES
NO
* Agilent will not accept delivery of any product that is exposed to radioactive, biological, explosive substances or dioxins, PCB’s without written evidence of decontamination.
Engineering Facts
13 facts · every claim linked to a page
Installation
The pump must be mounted so that the high-vacuum flange connection can withstand a torque of 1000 Nm around its axis.
The mounting hardware and chamber flange must be rated for this torque to prevent the pump from detaching if the rotor seizes.
Source
“with a connection capable of withstanding a torque of 1000 Nm around its axis”
Installation
Only high-resistance class clamps or bolts of grade 8.8 or higher may be used for the high-vacuum flange.
Lower-grade fasteners may fail under the sudden torque generated by a rotor stop, causing the pump to tear off.
Source
“Use only high resistance class clamps (or bolts) ≥ 8.8.”
Installation
Mixing different flange types is not permitted; the pump must be fastened to flanges recommended by Agilent.
Non-recommended flange combinations can lead to the pump twisting or tearing off if the rotor suddenly stops.
Source
“The combination of different flange types is not allowed.”
Installation
For ISO-K flanges, use 4 claw clamps for ISO 100 or ISO 160 and tighten to 22 Nm.
Correct clamp count and torque ensure a secure vacuum seal and mechanical retention.
Source
“Tighten the claw clamps with a torque of 22 Nm.”
Installation
For ConFlat flanges, use 16 screws for CFF 6-inch or 20 screws for CFF 8-inch and tighten to 20 Nm.
Proper screw count and torque are required to compress the copper gasket and hold the flange under rotor-failure torque.
Source
“Tighten the screws with a torque of 20 Nm.”
Operation
The pump must always be soft-started on first use and after 60 days of inactivity.
Soft start prevents excessive wear and stress on the rotor bearings after long idle periods.
Source
“the turbomolecular pumps must be always soft-started after a period of 60 days of inactivity of the pump.”
Operation
The pump can only pump non-corrosive and non-oxidizing gases or gas compounds.
Using the pump with corrosive or oxidizing gases without proper purge flow voids warranty and damages the pump.
Source
“puede bombear cualquier tipo de gas o compuesto de gas no corrosivo y oxidante”
Operation
The maximum rotational speed of the TwisTorr 305 FS is 60,600 rpm.
Knowing the maximum speed helps assess rotor energy and the required flange torque rating.
Source
“max. 60600 rpm”
Operation
Emergency stop by removing power must be used only in an emergency; improper use can damage the pump and void warranty.
Frequent emergency stops can degrade bearings and electronics, so normal stop commands should be used.
Source
“This mode is to be used only in an emergency. The use of this stop mode could lead to faults and/or damages at the pump.”
Safety
The pump must be installed indoors only and must not be exposed to rain, ice, snow, dust, aggressive gases, or explosive environments.
Outdoor or hazardous environments can cause condensation, corrosion, or ignition risk, leading to pump failure or safety incidents.
Source
“Install this device indoors only, and do not use the pump in an environment exposed to atmospheric agents (rain, ice, snow), dust, aggressive gases, or in explosive environments or those with a high fire risk.”
Safety
Operating environment must be kept at +5 °C to +35 °C and 0–90 % relative humidity non-condensing to avoid condensation.
Condensation inside the pump can cause electrical shorts and corrosion; the specified range prevents this.
Source
“temperature: from +5 °C to +35 °C”
Maintenance
The rotor fatigue life is estimated at 405,000 start-stop cycles; Agilent recommends returning the pump for maintenance after 17 years of use.
This gives a preventive maintenance interval to avoid unexpected rotor failure and costly downtime.
Source
“305 FS涡轮分子泵转子的抗疲劳性为从停止到全速再回到停止的405000次循环。”
Maintenance
The shelf life of a turbomolecular pump is 12 months from the shipping date; exceeding it requires return to factory.
Long storage can degrade bearings and lubricants; knowing the shelf life prevents commissioning a degraded pump.
Source
“the shelf life of a turbomolecular pump is 12 months from the shipping date.”