Waters Xevo TQ-GC Site Preparation Guide
Technical Reference Summary
Covered Models & Document Scope
Covers the Waters Xevo TQ-GC GC/MS-MS system site preparation, including the Xevo TQ-GC mass spectrometer, GC oven, data system, and external backing pump options (Edwards RV5 or Vacuubrand RE6). Optional equipment addressed includes the CTC PAL3 autosampler, Waters Hydrogen Control Box, CI option, and PAL3 headspace/SPME options. Document number 715005566, VER. 07.
Section-by-Section Technical Breakdown
Specifies the weights of the system components and the lifting equipment required to move the instrument onto the laboratory bench. The instrument and GC must be lifted with suitable equipment, not manually, and the bench must support the combined weight of the mass spectrometer, data system, and GC oven.
- Xevo TQ-GC weight: 84 kg (185.2 lbs)
- Data system weight: <50 kg (110 lbs)
- Rotary pump weight: 15.7 kg (34.6 lbs)
- Gas chromatograph weight: 50 kg (110 lbs)
- Use a forklift truck or A-frame hoist; the supplied lifting harness must be used to lift the instrument from the crate onto the bench
- Do not use the GC interface as a lifting handle
- Note the additional height of the CTC PAL3 autosampler, which may involve working above shoulder height
Provides plan and front-view layouts and dimensional data for the Xevo TQ-GC, rotary pumps, GC oven, and sampler. Defines minimum service clearances and bench leveling requirements.
- Xevo TQ-GC: width 352 mm (13.86 in) without GC interface, 500 mm (19.69 in) with GC interface; depth 913 mm (35.94 in); height 593 mm (23.35 in)
- Minimum service clearance: 600 mm (23.6 in) front, back, and left side; temporary clearance of 1000 mm (39.4 in) required for the left side
- Mass spectrometer must be installed on a surface level to within ±1° in any direction
- Instrument supplied with a 2.5-m (8-ft) power cord
- Waters recommends an additional 150 mm (6 in) behind the workbench to accommodate vacuum tubing
- Edwards RV5 rotary pump: 158 mm W x 430 mm L x 225 mm H, 25 kg without oil
- Vacuubrand RE6 rotary pump: 142 mm W x 370 mm L x 207 mm H, 15.8 kg with oil and 15.4 kg without oil
- Rotary pump must be positioned on the floor within 1 m (3.3 ft) of the rear of the chassis and elevated 150 to 200 mm (6 to 8 in) above the floor
- GC oven and sampler: width 580 mm (22.8 in), depth 530 mm (20.9 in), height 1040 mm (40.9 in) including turret/injector or 1430 mm (56.3 in) with the CTC PAL3 autosampler
Describes data system placement and cabling, and states the electrical safety standards and pollution degree for the Xevo TQ-GC. The data system may be placed on the same bench or a separate desk.
- A 3-m (10-ft) X-wire network cable connects the computer to the mass spectrometer
- Data system power cords for PC and monitor are approximately 2.5 m (8 ft) in length
- Complies with IEC 61010-1:2010 and EN 61010-1:2010 (European Low Voltage Directive 2014/35/EU)
- For Australia and New Zealand, building installation must comply with AS3000
- Suitable for Pollution Degree 2 environments (only non-conductive pollution allowed)
Lists nominal voltages, currents, power consumption, connector types, and number of sockets for each system component. Specifies voltage fluctuation limits and protective device recommendations.
- Xevo TQ-GC: 200 to 240 V, 50/60 Hz, 4.5 A, 900 W, IEC 60320 C19, 1 socket
- GC oven: 220 to 240 V, 50/60 Hz, 13 to 16 A, 2.2 kW, IEC 60320 C19, 1 socket
- PC: 100 to 240 V, 50/60 Hz, 10 to 5 A, 110 W, IEC 60320 C13, 1 socket
- Monitor: 100 to 240 V, 50/60 Hz, 1.5 A, 25 W, IEC 60320 C13, 1 socket
- Vacuubrand RE6 rotary pump: 200 to 240 V, 50/60 Hz, 5.4 A, 300 W, IEC 60320 C19, 1 socket
- Edwards RV5 pump: 220 to 240 V/50 Hz, 3.4 A, 450 W; 230 to 240 V/60 Hz, 3.0 A, 550 W; 110 V/50 Hz, 6.8 A, 450 W; 115 to 120 V/60 Hz, 6.9 A, 550 W; IEC 60320 C13, 1 socket
- LAN box: 100 to 240 V, 50/60 Hz, 3 A, 25 W, IEC 60320 C7, 1 socket
- Waters Hydrogen Control Box (option): 200 to 240 V, 50/60 Hz, 3 A, 10 W, 9-mm jack plug, 1 socket
- CTC PAL3 autosampler (option): 100 to 240 V, 50/60 Hz, 3 A, 200 W, IEC 60320 C13, 1 socket
- Main power supply voltage fluctuations must not exceed ±10% of nominal device voltages
- Power sockets must be located within 2 m (6.5 ft) of the instrument
- Do not use an uninterruptible power supply with a GC
- Waters recommends RCDs for UK and Europe, GFCIs for the rest of the world
Covers the use of transformers when supply voltages are outside the specified range, and lists the power cord plug types supplied by Waters for various countries. Also addresses UPS recommendations and dimensions.
- If supply voltages may not meet the specified operating range, a transformer must be used
- For instruments fitted with a transformer, the RCD or GFCI must be fitted on the primary (supply) side of the transformer
- Power cords are shipped as requested at order; they must only be used with the Waters system
- In North America, if L6-15 outlets are not available, L6-20 outlets and cord sets may be used instead
- Waters recommends a UPS; Waters supplies UPS units configured for its MS systems that step up single-phase line voltage to 230 V AC
- For North America, the UPS system requires one L6-30 (30 amp) wall socket
- UPS approximate maximum dimensions: 838 mm (33 in) height, 305 mm (12 in) width, 813 mm (32 in) depth
Specifies the laboratory environmental conditions required for normal operation, including temperature, humidity, altitude, vibration, magnetic fields, and radio emissions. Also covers ventilation and exhaust requirements.
- Ambient temperature range: 15 to 28 °C (59 to 82 °F)
- Ambient temperature fluctuation must be less than 2 °C (3.5 °F) peak-to-peak in 1.5 hours
- Relative humidity: 20% to 80%, noncondensing
- Altitude: below 2000 m (6562 ft)
- Must be positioned away from magnetic fields greater than 10 Gauss
- Must not be placed within an RF field greater than 1.0 V/m
- Active exhaust vent must provide a minimum vacuum of 0.030 psi (2 mbar) below atmospheric pressure and support a maximum instrument exhaust gas load of 15 mL/min (900 mL/hour)
- Standard GC configuration releases up to 100 mL/min of carrier gas into the laboratory
Details the required purity, pressure, and connection specifications for GC carrier gas, CI reagent gas, collision gas, and PAL3 headspace/SPME gas. Includes piping material requirements and safety warnings for hydrogen and ammonia.
- Helium carrier gas: 99.999% or better with <1 ppm water, dual-stage regulator up to 200 psi (14 bar), regulated at 56 to 74 psi (4 to 5 bar) maximum 145 psi (10 bar)
- Hydrogen carrier gas: 99.999% or better with <1 ppm water, hydrogen-specific dual-stage regulator up to 200 psi (14 bar), maximum supply pressure to GC 145 psi (10 bar)
- Hydrogen requires the Hydrogen Compatibility Kit (205001743)
- CI reagent gas: methane required for CI performance tests; dry, high purity (>99.9%), regulated at 14.7 psi (1 bar) with a two-stage regulator
- Collision gas: argon, dry, high purity (99.997%), regulated at 10 ±2 psi (0.69 ±0.1 bar) using a two-stage high purity regulator
- Typical argon usage during operation is less than 0.5 mL/hr
- PAL3 headspace/SPME gas: helium or nitrogen, minimum grade 55 (99.9995% purity), grade 59 or 60 preferable; pressure to CTC PAL3 inlet regulator between 1 and 6 bar
- Stainless steel tubing type 316L with compression fittings or orbital welded joints; medical grade, chemically cleaned and passivated or electropolished
- Copper tubing: no soldered joints, no fluxes; brazed joints with flux-free copper-phosphorous-silver brazing rod under oil-free dry nitrogen
Specifies requirements for venting rotary pump exhaust gases outside the laboratory. Provides tubing size and length details.
- Rotary pump exhaust must be vented to atmosphere outside the laboratory using a user-supplied fume hood or industrial vent
- Five meters (16 ft) of 12-mm ID PVC tubing is supplied
- If extra distance is needed, supply an adapter and tubing with an internal diameter of at least 51 mm (2 inches)
- The fume hood/industrial vent must be equipped with an extraction fan system
Describes the test samples, solvents, reagents, sample preparation equipment, and cleaning facilities required for installation and routine operation. Emphasizes the need for high-purity solvents and clean glassware.
- A test sample kit is supplied with the instrument for installation setup
- High-purity solvents (GC-MS grade or better) are required
- Sample preparation equipment includes calibrated syringes (1 µL to 1 mL), calibrated analytical balance, nitrile gloves, lint-free tissue
- Ultrasonic bath required for routine cleaning: at least 300 mm x 150 mm x 100 mm deep (12 in x 6 in x 4 in)
- Surfactants must not be used for cleaning glassware or other components
- Cleaning vessels must have a diameter of at least 120 mm (5 in) and be approximately 120 mm (5 in) high
Table 4 lists the waste and gas connections required for installation, indicating which items are supplied with the instrument and which must be supplied by the customer. Also includes delivery pressures for inlet types.
- Rotary pump exhaust: 12-mm OD barbed fitting on system; 5-m (16-ft) PVC tube, 12-mm ID supplied; industrial vent or fume hood to be supplied by customer
- Collision gas supply: 1/8-inch Swagelok fitting; 3-m (10-ft) of 1/8-inch OD stainless steel tubing supplied; argon supply regulated to 10 ±2 psi through a 1/8-inch adapter to be supplied by customer
- GC carrier gas supply: 1/8-inch Swagelok male fitting; 1/8-inch OD stainless steel tubing to a regulated supply supplied
- CI reagent gas supply: 1/8-inch Swagelok male fitting; 1/8-inch Swagelok female fitting supplied; 1/8-inch OD stainless steel or copper tubing to a regulated gas supply to be supplied by customer
- Purge gas supply (for CTC PAL3 headspace or SPME options): 1/8-inch Swagelok fitting; 1/8-inch OD stainless steel tubing and 1/8-inch Swagelok nuts and ferrules supplied; 1/8-inch tee-piece and inert purge gas at 14.5 to 87.0 psi (1 to 6 bar) to be supplied by customer
- Delivery pressures for inlet types: Split/splitless 150 psi; Multimode 100 psi; PTV 100 psi
- Carrier maximum: Split/splitless 1,172 kPa (170 psi); Multimode 827 kPa (120 psi); PTV 827 kPa (120 psi)
- Carrier minimum: 137.9 kPa (20 psi) above maximum pressure used in method
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Xevo TQ-GC weight is 84 kg (185.2 lbs); GC weight is 50 kg (110 lbs); rotary pump weight is 15.7 kg (34.6 lbs); data system weight is <50 kg (110 lbs). p.6 ↗ | Bench must support the combined weight of the mass spectrometer, data system, and GC oven; lifting equipment must be rated for these loads. | Whole series |
| Minimum service clearance is 600 mm (23.6 in) for the front, back, and left side of the instrument; a temporary clearance of 1000 mm (39.4 in) is required for the left side. p.8 ↗ | Laboratory layout must provide these clearances for service access. | Whole series |
| Mass spectrometer must be installed on a surface level to within ±1° in any direction. p.9 ↗ | Bench or floor must be leveled accordingly to ensure proper operation. | Whole series |
| Rotary pump must be positioned on the floor within 1 m (3.3 ft) of the rear of the chassis and elevated 150 to 200 mm (6 to 8 in) above the floor. p.9 ↗ | Provides easy access for routine maintenance such as changing pump oil. | Whole series |
| Main power supply voltage fluctuations must not exceed ±10% of the nominal device voltages. p.10 ↗ | If supply is unstable, a transformer or power conditioner/stabilizer must be used. | Whole series |
| Do not use an uninterruptible power supply with a GC. p.10 ↗ | An unsafe condition can result if the GC remains powered-on during a power loss. | GC oven |
| Ambient temperature range required for normal operation is 15 to 28 °C (59 to 82 °F) with fluctuation less than 2 °C (3.5 °F) peak-to-peak in 1.5 hours. p.16 ↗ | Laboratory air conditioning must maintain this stability. | Whole series |
| Relative humidity must be 20% to 80%, noncondensing. p.16 ↗ | Humidity control may be required in some laboratories. | Whole series |
| Instrument must be operated below 2000 m (6562 ft) altitude. p.16 ↗ | Site altitude must be verified. | Whole series |
| Active exhaust vent must provide a minimum vacuum of 0.030 psi (2 mbar) below atmospheric pressure and support a maximum instrument exhaust gas load of 15 mL/min (900 mL/hour). p.16 ↗ | Exhaust system must be sized accordingly. | Whole series |
| Helium carrier gas must be 99.999% or better with <1 ppm water content, supplied through a dual-stage regulator capable of regulating up to 200 psi (14 bar); gases are generally regulated at 56 to 74 psi (4 to 5 bar) maximum 145 psi (10 bar). p.18 ↗ | Gas supply system must meet purity and pressure requirements. | Whole series |
| Hydrogen carrier gas requires the Hydrogen Compatibility Kit (205001743), which shuts off the supply of carrier gas in the event that the Xevo TQ-GC loses vacuum. p.18 ↗ | Must be purchased separately when using hydrogen as carrier gas. | Whole series |
| Argon collision gas must be dry, high purity (99.997%), regulated at 10 ±2 psi (0.69 ±0.1 bar) using a two-stage high purity gas regulator. p.19 ↗ | Regulator must be suitable for high purity and low pressure. | Whole series |
| Typical argon usage during operation is less than 0.5 mL/hr. p.19 ↗ | Argon supply will last a long time; cylinder size can be small. | Whole series |
| PAL3 headspace/SPME gas must be helium or nitrogen of minimum grade 55 (99.9995% purity); pressure to CTC PAL3 inlet regulator must be between 1 and 6 bar. p.19 ↗ | Gas supply must meet purity and pressure range. | PAL3 headspace and SPME options |
| Ultrasonic bath for cleaning must be at least 300 mm x 150 mm x 100 mm deep (12 in x 6 in x 4 in). p.22 ↗ | Laboratory must have an appropriately sized ultrasonic bath. | Whole series |
| Delivery pressures for inlet types: Split/splitless 150 psi; Multimode 100 psi; PTV 100 psi. Carrier maximum: Split/splitless 1,172 kPa (170 psi); Multimode 827 kPa (120 psi); PTV 827 kPa (120 psi). p.23 ↗ | Gas supply pressure must be set according to inlet type. | GC inlets |
| Carrier minimum pressure is 137.9 kPa (20 psi) above maximum pressure used in method. p.23 ↗ | Ensure supply pressure exceeds method maximum by at least 20 psi. | GC inlets |