The Xevo TQ-GC is a tandem quadrupole mass spectrometer with dual high-resolution quadrupole analyzers (MS1/MS2), a curved T-Wave collision cell, and an off-axis photomultiplier detector. It covers a mass range of 2–1250 m/z and supports EI (standard) and CI (optional) ionization, with acquisition rates up to 500 MRM/sec. This model is intended for targeted quantitation and confirmation in GC-MS/MS workflows, offering multiple acquisition modes including MRM, SIR, and simultaneous full scan/MRM (RADAR).

Direct drop-in replacements

Shimadzu
MALDI 8020
100%
match
⚖️ Compare
Shimadzu
Maldimini 1
66%
match
⚖️ Compare
Shimadzu
Nexera Mikros
3%
match
⚖️ Compare
Shimadzu
Axima Assurance
0%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Type tandem quadrupole
Configuration benchtop
Detector types Low noise, off axis, long life photomultiplier detector
Ion source EI standard, CI optional
Scan speed 20000 Da/s (MS scanning mode)
EI filaments mfr Dual EI Yttrium-coated long-lasting filaments
Electron energy mfr Range 10–200 eV
Pump down time mfr 15 minutes
Source temperature mfr 250 °C
Transfer line temperature mfr 350 °C
Collision cell mfr Innovative curved T-Wave
Vacuum system mfr One split-flow air-cooled vacuum turbomolecular
Regulatory approvals/marks mfr CE and NRTL
Software mfr MassLynx version 4.2; TargetLynx XS Application
Acquisition modes mfr Full scan; MS Product ion scan; Precursor ion sc
Acquisition rate mfr up to 500 MRM/sec
Ion source (alt var) mfr electron impact (EI) or chemical ionization (CI)
Dimensions, Weight & Mechanical Design
Mass range 2–1250 m/z
Height 59.3 cm
Length 1030 mm (40.6 inches) cm
Weight < 135 kg (< 297.6 lbs) kg
Width 97.0 cm
Depth mfr 91.3 cm (36.0 in)
Mass analyzer mfr Two high resolution, high stability quadrupole a
Width (instrument crate) (alt var) mfr 770 mm (30.3 inches) cm
Height (instrument crate) (alt var) mfr 880 mm (34.6 inches) cm

Description

Construction and Architecture

Integration & Operating Notes

Electrical
Requires dedicated power circuits for instrument and vacuum pumps

The system includes a split-flow turbomolecular pump and a backing pump, both of which have specific power requirements. Provide dedicated circuits with proper grounding and surge protection. Refer to the site preparation guide for voltage, current, and frequency ratings; do not share circuits with high-noise equipment.

Site Preparation Guide (typical

Selection & Integration Q&A

When should I choose the Xevo TQ-GC over a single-quadrupole GC-MS system?

Choose the Xevo TQ-GC when you need high-selectivity MRM quantitation at trace levels, confirmation via product ion scans, or simultaneous full-scan and MRM (RADAR). A single quad is adequate for simple full-scan screening but lacks the sensitivity and specificity of tandem MS for complex matrices.

What are the requirements for the GC system that connects to the Xevo TQ-GC?

The GC must provide a suitable column effluent interface to the transfer line, and its oven must be capable of reaching the transfer line temperature (up to 350 °C). The GC should have a compatible autosampler and be controllable via MassLynx or an independent controller; ensure synchronization of start signals.

What maintenance items and accessories are not included in the base delivery?

The base system includes the instrument, MassLynx 4.2 with TargetLynx XS, and standard EI source. Not included: CI source (optional), additional filaments, source cleaning kits, backing pump oil, GC column and consumables, and a dedicated PC if not specified. Plan for a spare filament and regular source cleaning supplies.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Waters Xevo TQ-GC Site Preparation Guide
Manual ·533,5 КБ ·28 pp.
View PDF
PDF
Xevo TQ-GC Overview and Maintenance Guide
Brochure ·6,9 МБ ·137 pp.
View PDF