Agilent Technologies 1260 Infinity III Bio-Inert LC System

Agilent Technologies 1260 Infinity III Bio-Inert LC System — Liquid chromatography with Flow rate 0.01 liter/minute, Max pressure 60 megapascal, Pumping speed 6×10⁻⁴ m³/h.

The 1260 Infinity III Bio-inert LC System is a metal-free, titanium-based HPLC system rated to 600 bar with a pH range of 1–13 and 2 M salt tolerance. It combines a Bio-inert pump (10 mL/min at 200 bar), multisampler (carryover <9 ppm), and thermostatted column compartment (10 °C below ambient to 85 °C). It is intended for SEC, ion-exchange, and peptide mapping of biotherapeutics where sample contact with metal must be avoided.

Flow rate
0.01 liter/minute
Max pressure
60 megapascal
Pumping speed
6×10⁻⁴ m³/h

Agilent Technologies Bio LC Solutions Series — Model Comparison & Selection

Model Flow rate, liter/minuteMass rangeMax pressure, megapascalPumping speed, m³/hWavelength rangeSample volume, milliliter
1260 Infinity III Bio-Inert LC System Current 0.01—606×10⁻⁴190...950—
1260 Infinity Amino Acid Analysis System ——————
1290 Infinity III Bio LC | Biocompatible UHPLC system —300...3,200———0.005

Full specifications

Core Performance & Functional Specifications
Flow rate 10 mL/min
Max pressure 600 bar
Pumping speed 10 mL/min
Wavelength range 190...950 nm
Carryover 9 ppm
Data Rate 74 Hz
Internal Volume 1.0 µL
Maximum Data Acquisition Rate 120 Hz
Maximum Individual Bottle Volume 2 L
Maximum System Pressure 600 bar
Salt Tolerance 2 M
Sample Capacity 6144 samples
Solvent Cabinet Maximum Volume 8 L
Maximum system pressure (metal-free sample flow path) mfr 600 bar
Optional injection volume range mfr 5 mL
Double mfr 7116 A/B
Single mfr 7114 A/B
pH short term mfr 14
Interfaces, Connections & Control
Maximum torque for UHP-FF fitting mfr 0.8 Nm
Capillary jut-out before fitting installation mfr 5 mm
Capillary protrusion when sliding fitting on capillary mfr 5 mm
Materials, Media & Process Compatibility
Pressure stability of stainless-steel cladded PEEK capillaries mfr 600 bar
Operating Conditions, Environment & Protection
Column Compartment Temperature 10...85 °C
Standards, Safety & General Data
Standard injection volume mfr 20 μL

Description

Construction and series positioning

Behavior under load and media

Detection and control

Integration & Operating Notes

Process Limits
Never use DMF as mobile phase with PVDF leak sensor

DMF corrodes the PVDF leak sensor, leading to false leak detection or sensor failure. If DMF-containing methods are required, consult Agilent for alternative sensor material or additional protective measures; regular inspection of the leak sensor for corrosion is mandatory.

System Manual, Leak Sensor, page 42

Selection & Integration Q&A

When should I choose the 1260 Infinity III Bio-inert LC System over the 1290 Infinity III Bio LC or the 1260 Infinity III Prime Bio LC?

Choose the 1260 Bio-inert for methods requiring up to 600 bar and full metal-free flow path with pH 1–13 and 2 M salt tolerance. The 1290 Bio LC offers higher pressure (up to 1300 bar) for UHPLC; the 1260 Prime Bio LC is a lower-cost biocompatible option but may not have the same inertness. Select based on pressure needs and sample sensitivity to metal.

What are the requirements for connecting the Bio-inert pump to the column compartment and detector?

Use only bio-inert capillaries and fittings; do not mix with bio/biocompatible parts. UHP-FF fittings require 5 mm capillary protrusion and max 0.8 Nm torque. The Bio-inert ZDV union (5067-4741) must be installed in the prescribed sequence. Ensure the flow path is metal-free throughout, using stainless-steel-clad PEEK capillaries rated to 600 bar.

What maintenance kits and accessories are recommended, and what is not included in the base delivery?

The 1260 Infinity II Max Uptime Kit Bio-inert (G5654-68707) includes capillaries, ZDV unions, fittings, solvent inlet filters, and tools. Valve kits (e.g., G5631A, G5632A, G5639A, G4235A) are separate. The HPLC System Tool Kit is included. Not included: columns, mobile phases, sample vials, and the InfinityLab Level Sensing (G7175A) unless ordered. Regular maintenance includes leak sensor inspection and capillary/fitting replacement.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
1260 Infinity III Bio-Inert LC System Manual
Manual ·76 pp.