Agilent Technologies 1290 Infinity II GPC/SEC system
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 1290 Infinity II GPC/SEC system

Agilent Technologies 1290 Infinity II GPC/SEC system — Liquid chromatography with Flow rate 3×10⁻⁴ liter/minute, Pumping speed 3.6×10⁻⁶ m³/h.

The 1290 Infinity II GPC/SEC System is a high-performance gel permeation chromatography system configured for isocratic THF operation with a 1290 Infinity II pump and RID detection. It supports flow rates from 0.06 mL/min (micro-GPC) to 1.0 mL/min and column temperatures up to 35 °C, distinguishing it from lower-pressure 1260 Infinity II siblings. It is used for molecular-weight distribution analysis of polymers and resins.

Flow rate
3×10⁻⁴ liter/minute
Pumping speed
3.6×10⁻⁶ m³/h

Agilent Technologies GPC SEC Solutions Series — Model Comparison & Selection

Model Pumping speed, m³/hSample volume, milliliterColumn particle size, µmColumn temperature, °CColumn inner diameter, mmColumn length, m
1290 Infinity II GPC/SEC system Current 3.6×10⁻⁶0.023354.60.25
1260 Infinity II Bio-SEC System 0.024—1.8—2.10.05
1260 Infinity II Bio-SEC system | Biological polymer analysis 4.8×10⁻⁵0.012.730——
1260 Infinity II Multidetector GPC/SEC System —5...500—220——

Full specifications

Core Performance & Functional Specifications
Flow rate 0.3 mL/min
Pumping speed 0.06 mL/min
Column Particle Size 3 µm
Column temperature 35 °C
Data Rate 9.25 Hz
Filter Membrane Diameter 47 mm
Filter Membrane Pore Size 0.2 μm
Ptfe Frits Per Pm Kit 5 pk
Retention Time 13.805 min
Sample volume 20 µL
Solvent Bottle Volume 1 L
Solvent Filtration Assembly Flask Volume 1 L
Solvent Filtration Assembly Funnel Volume 250 mL
Solvent Inlet Filter Pore Size 10 μm
Well-plate Closing Mat Quantity 50 pk
Well-plate Volume 0.5 mL
Signal polarity mfr Positive Positive
Quick Connect heat exchanger flow rate mfr 2.5 mL/min
Capillary dimensions mfr 0.12 mm
Preventive maintenance cost reduction mfr 30
Pump seals per PM kit mfr 4
Seal cap assembly sets per PM kit mfr 2
Gold seals for outlet valve per PM kit mfr 2
Seal washes per PM kit mfr 4
Solvent bottles per convenience kit mfr 3 amber
Mp mfr 1314 g/mol
Mn mfr 1048 g/mol
Mw mfr 1265 g/mol
Solvent consumption per sample mfr 30
Flow Rates mfr 0.06 mL/min
Stop Time mfr 22 minutes
RID mfr 35 positive
Connection capillary id mfr 75 μm
Micro GPC flow rate mfr 60 μL/min
Dimensions, Weight & Mechanical Design
Column Inner Diameter 4.6 mm
Column Length 250 mm
Connection capillary length mfr 105 mm
Interfaces, Connections & Control
Quick Turn Fitting Pressure Limit 800 bar
Solvent bottle thread mfr 45
Standards, Safety & General Data
Calibration standard Mp mfr 47,190 g/mol
Standard RID cell volume mfr 8 μL

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection, and operation

Integration & Operating Notes

Process Limits
Do not exceed 800 bar at quick-turn fittings

Quick-turn fittings are rated to 800 bar; operating above this pressure risks fitting failure and leakage. The 1290 Infinity II pump can generate high pressures, so method development must respect this limit, especially with narrow-bore columns.

Factory Specification Sheet
Process Limits
Heat exchanger flow limited to 2.5 mL/min

The Quick Connect heat exchanger is specified for 2.5 mL/min. Higher flow rates may reduce thermal equilibration efficiency and cause retention-time drift, so methods above this flow should be validated for column temperature stability.

Factory Specification Sheet
Maintenance
PM kit includes specific wear parts; plan replacement intervals

The preventive maintenance kit contains 2 pump seals, 2 seal cap assembly sets, 2 gold seals for the outlet valve, and 4 seal washes. These consumables should be replaced at the manufacturer-recommended interval to avoid leaks and pressure fluctuations; the PM program is stated to reduce maintenance cost by 30%.

Flyer LC Supplies Recommendations 5994-5755EN
Process Media
Use 10 µm inlet filter and 0.2 µm solvent filtration

Solvent inlet filters have 10 µm pores, and solvent filtration assemblies use 47 mm, 0.2 µm membranes. Particulates in the mobile phase can damage the pump and column; filtering THF and using the inlet filter are essential for reliable GPC/SEC results.

Flyer LC Supplies Recommendations 5994-5755EN

Selection & Integration Q&A

When should I choose the 1290 Infinity II GPC/SEC System instead of a 1260 Infinity II LC System?

Choose the 1290 Infinity II GPC/SEC System for high-performance GPC/SEC with flow rates down to 0.06 mL/min and pressures up to 800 bar, especially for micro-GPC with 2.1 mm columns. The 1260 Infinity II LC System is a lower-pressure analytical HPLC platform not optimized for GPC/SEC.

What detector and column configurations are compatible with this system for GPC/SEC?

The application note describes use with either the 1290 Infinity II Micro-RID (G7162A) or the 1260 Infinity II RID (G7162B). Columns tested include a single MesoPore 2.1 × 250 mm (PL1913-5325) at 0.06 mL/min and two PLgel 3 µm MiniMixE 4.6 × 250 mm (PL1510-5300) at 1.0 mL/min.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Improved GPC with the Agilent 1290 Infinity II LC System