Agilent Technologies 240FS AA
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 240FS AA

Agilent Technologies 240FS AA — Atomic spectroscopy.

The Agilent 240FS AA is a Fast Sequential flame atomic absorption spectrometer with transverse Zeeman background correction and an 8-position lamp turret, measuring up to 10 elements per sample in about 2 minutes. It sits below the 280FS AA in the flame series and is distinct from the 240Z/280Z graphite-furnace Zeeman instruments. It is used for sequential multi-element trace metal determination in liquid samples by air/acetylene or nitrous-oxide/acetylene flame.

Agilent Technologies Flame Atomic Absorption Instruments Series — Model Comparison & Selection

Model SensitivityPrecision, mmAnalysis time 8 elements, minutesDilution error, %Sample consumption, mLSensitivity increase, %
240FS AA Current —195-11040
280FS AA —195-11040
55B AA 0.90.5————

Full specifications

Core Performance & Functional Specifications
Acetylene Purity 99.5 %
Analysis Time (8 Elements) 95 minutes
Dilution Error -1 %
Direct Viewing Distance From Flame 0.2 m
Exempt Distance From Flame 7.1 m
Maximum Flame Door Open Exposure 13 minutes
Precision 1 %RSD
Sample Consumption 10 mL
Sensitivity Increase 40 %
Viewing Distance Rg1 2.3 m
Viewing Distance Rg2 0.8 m
Lamps mfr four four
Background correction mfr Transverse Zeeman Zeeman
Dilution factor mfr 1
Number of determinations mfr 53
Reproducibility mfr 1
Elements measured per sample mfr 10 elements
Analysis time per sample mfr 2 minutes
Lamp positions mfr 8 positions
Default operating mode mfr Fast Sequential Sequential
Exhaust ventilation rate mfr 6 minute
Exhaust fan distance from flame mfr 3 meters
UV intensity emitted mfr 37.7
UV hazard classification mfr 3
Burner slot replacement limit, nitrous-oxide/acetylene mfr 0.47 mm
Burner slot replacement limit, air/acetylene mfr 0.54 mm
Acetylene maximum pressure mfr 105 kPa
Acetylene fuel supply pressure range mfr 65 kPa
Acetylene cylinder minimum pressure before replacement mfr 700 kPa
Acetylene maximum consumption rate mfr 1 hour
Solvent minimum specific gravity mfr 0.75
Acetylene fuel supply pressure mfr 65...100 kPa
Acetylene cylinder replacement pressure mfr 700 kPa
Burner slot replacement threshold mfr 0.54 mm
Minimum solvent specific gravity mfr 0.75
Schedule mfr 62
Cleaning mfr 63
Gas cylinders mfr 64
Preventing flashbacks mfr 65
Spare parts mfr 65
Ultraviolet Hazard mfr 3
Acetylene cylinder minimum pressure mfr 700 kPa
Acetylene maximum consumption mfr 1 hour
Zeeman magnetic field peak mfr 0.8 Tesla
Zeeman magnet exclusion distance mfr 300 mm
UV intensity mfr 37.7
Hollow cathode lamp capacity 240 series mfr Four lamp capacity capacity
Hollow cathode lamp capacity 280 series mfr Eight lamp capacity capacity
Dimensions, Weight & Mechanical Design
Burner slot width maximum, N2O/Acetylene Mark 7 mfr 0.46 mm
Burner slot width maximum, Air/Acetylene Mark 7 mfr 0.54 mm
Burner slot maximum width mfr 0.54 mm
Nitrous-oxide/acetylene burner slot maximum width mfr 0.47 mm
Air/acetylene burner slot maximum width mfr 0.54 mm
Mark 7 N2O/acetylene burner slot maximum width mfr 0.46 mm
Mark 7 air/acetylene burner slot maximum width mfr 0.54 mm
Electrical, Power & Drive Specifications
Non-Zeeman 240/280FS fuse rating mfr 4 mm
Zeeman circuit breaker rating mfr 15 breaker
Zeeman fuse T3.15 A rating mfr 3.15 mm
Zeeman fuse F315 mA rating mfr 315 mm
Zeeman fuse F1A rating mfr 1 mm
Zeeman fuse T6.3 A rating mfr 6.3
Zeeman fuse F800 mA rating mfr 800 mm
Standards, Safety & General Data
Standard concentration mfr 5 mg/L

Description

Construction and series positioning

Behavior and performance

Installation, connection and control

Integration & Operating Notes

Interlock
Flame door interlock blocks ignition and must never be bypassed

The flame door interlock prevents ignition if the door is opened during ignition; if the interlock is lost, contact Agilent immediately. Bypassing it is prohibited. With the flame door open, UV exposure must not exceed 13 minutes over an 8-hour period, and direct viewing at 0.2 m or less is Risk Group 3 (UV intensity 37.7 mW/m2).

240FS AA User Manual, Safety Practices and Hazards, p.13-14
Maintenance
Daily leak test and weekly O-ring inspection are mandatory

Perform leak tests at all joints and seals daily before use with soapy water or a proprietary solution, never with a naked flame. Weekly, check the O-rings inside the burner, spray chamber and bung and the two sealing the gas inlets to the spray chamber; damaged O-rings can cause gas leakage, fire or flashback and must be replaced immediately.

240FS AA User Manual, Safety Practices and Hazards, p.12; Maintenance p.62-63
Electrical
Verify operating voltage before connecting to mains

Do not connect the instrument to mains until the operating voltage is verified for the specific outlet. Use only the Agilent-supplied power cord, replacing it only with an equivalent specified in the AA Site Preparation Guide, to ensure electrical safety and proper grounding.

240FS AA User Manual, Safety Practices and Hazards, p.9

Selection & Integration Q&A

When should I choose the 240FS AA rather than the 280FS AA or a graphite-furnace 240Z/280Z model?

Choose the 240FS AA for sequential multi-element flame analysis (up to 10 elements per sample in about 2 minutes) with transverse Zeeman background correction. The 280FS AA is the higher sibling in the same flame Fast Sequential series; 240Z/280Z AA are graphite-furnace Zeeman instruments for trace levels, and Duo systems combine one flame and one furnace unit under one computer.

What gas supply and plumbing requirements must the lab meet before installing the 240FS AA?

Acetylene must be instrument grade, at least 99.5 % purity in acetone, supplied at 65–100 kPa and never above 105 kPa; replace cylinders below 700 kPa. Use left-hand threads for fuel and right-hand threads for support gases, and avoid copper or brass with more than 65 % copper. Only air, nitrous oxide and acetylene may be used as oxidant/fuel; never oxygen.

What ventilation and UV exposure limits apply to the 240FS AA installation?

The exhaust system must extract at least 6 m3/min (200 cfm) with the fan at least 3 m from the flame, venting to outside air via a collector hood and ducting, never within the building. With the flame door open, UV exposure is limited to 13 minutes per 8-hour period; direct viewing at 0.2 m or less is Risk Group 3 (37.7 mW/m2).

What maintenance and accessories are required beyond the base instrument delivery?

Daily leak tests and weekly O-ring checks (burner, spray chamber, bung, gas inlets) are mandatory; the maintenance schedule also covers exhaust, gas hoses, drain tube, burner, spray chamber, liquid trap, compressor filter and coolant. The SIPS 10/20 sample introduction pump systems, SPS 4 autosampler, USB-GPIB HS/HS+ communications and Tube-CAM are separate accessories; only Agilent-supplied or approved spare parts may be used.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent AA Spectrometer User's Guide
Manual ·68 pp.
View PDF