Agilent Technologies 2D-LC System
Agilent Technologies 2D-LC System — Liquid chromatography with Max pressure 130 megapascal.
The Agilent 2D-LC System is a two-dimensional liquid chromatography cluster built around a 1290 Binary Pump or 1290 Infinity II/III High-Speed Pump as the second-dimension pump, with 6-position/14-port multiple heart-cutting valves storing up to 12 cuts. It supports LC-LC, MHC, HiRes, LCxLC and ASM modes, with 4.6 mm second-dimension columns at typical 1 mL/min flow and 9 ppm carryover.
Agilent Technologies HPLC Systems Series — Model Comparison & Selection
| Model | Max pressure, megapascal | Carryover, % | Column inner diameter, mm | Maximum pressure, bar | Sample capacity | Typical flow rate, mL/min |
|---|---|---|---|---|---|---|
| 2D-LC System Current | 130 | 9×10⁻⁴ | 4.6 | 1300 | 6144 | 1 |
| Analytical HPLC and UHPLC Systems | — | 9×10⁻⁴ | 4.6 | 1300 | 6144 | 1 |
| GPC/SEC Systems, Macromolecular Analysis and Solution | — | 9×10⁻⁴ | 4.6 | 1300 | 6144 | 1 |
| Preparative HPLC Systems, InfinityLab LC Purification Solutions | — | 9×10⁻⁴ | 4.6 | 1300 | 6144 | 1 |
Full specifications
| Max pressure | 1300 bar |
| Carryover | 9 ppm |
| Gas Flow | 11 L/min |
| Ion source | 6 |
| Maximum pressure | 1300 bar |
| Minimum Dwell Time | 200 ms |
| Sample Capacity | 6144 samples |
| Typical Flow Rate | 1 mL/min |
| Productivity range mfr | Micrograms Micrograms |
| instrument downtime reduction mfr | 24 |
| ESI Source Parameter mfr | Similar to the High-end MS parameter parameter |
| SCAN mfr | 100...500 mode |
| Ion Mode mfr | Dual AJS ESI ESI |
| Ion Polarity mfr | Positive Positive |
| Storage Mode mfr | Both, Centroid preferred preferred |
| LCMS Stream mfr | MS MS |
| Acquisition Mode mfr | 1 |
| Instrument Parameters mfr | Source Parameters Parameters |
| Nebulizer mfr | 40 psig |
| SheathGasFlow mfr | 12 L/min |
| Scan Segment mfr | 1 |
| Fragmentor mfr | 120 |
| Skimmer1 mfr | 45 |
| OctopoleRFPeak mfr | 750 |
| Stop Time mfr | As pump/No limit limit |
| Ion polarity mfr | Positive Positive |
| ReferenceMasses mfr | Ref Mass Enabled Enabled |
| Column Inner Diameter | 4.6 mm |
| Peak Width | 0.06 min |
| Nozzle Voltage | 300 V |
| Vcap mfr | 3500 V |
| Gas Temp | 250 °C |
Description
Architecture and positioning in the series
- The 2D-LC System is a cluster, not a single module: the second dimension must be driven by a 1290 Binary Pump or any 1290 Infinity II/III High-Speed Pump (User Manual, Ch. 10, p. 312).
- Multiple heart-cutting uses 6-position/14-port selection valves with six loops each (Ch. 2, p. 18); Agilent multiple heart-cutting valves store up to 12 cuts at one time (Ch. 1, p. 10).
- Unlike the analytical, application-specific, bio and GPC/SEC siblings, this modification is defined by the valve/loop front end and the 2D pump requirement, not by a single separation chemistry.
- An optional valve allows switching between 1D and 2D separation without hardware reconfiguration (Ch. 1, p. 11); analytical fraction collection can be added (Ch. 1, p. 11).
Behavior in the different modes
- In heart-cutting (LC-LC), a started second-dimension analysis always finishes; a second first-dimension peak may be lost if sampled while the 2D run is still active (Ch. 2, p. 17). Second-dimension run time must therefore match the retention gap between 1D peaks.
- Multiple heart-cutting avoids that loss by parking cuts in the six loops while the second dimension is occupied (Ch. 2, p. 18).
- Active Solvent Modulation reduces first-dimension solvent effects, optimizing the second-dimension separation and increasing sensitivity, and is intended for MHC and high-resolution sampling across a wide range of 1D/2D solvents and gradients (Ch. 1, p. 10).
- Comprehensive 2D-LC (LCxLC) injects the total first-dimension eluent onto the second-dimension column, in contrast to the selective heart-cutting modes (Ch. 2).
Installation, control and operation
- Operation is under MassHunter Workstation; the 2D-LC instrument must be licensed in MassHunter, and the hardware license lives on a USB dongle installed in the 2D pump (Ch. 10, pp. 312-313).
- Agilent Lab Advisor 2.17 or higher is required for license handling, capillary configuration, instrument control, declustering, pump head leak test, pump leak rate test and system pressure test; the 2D-LC Capillary Leak Test needs 2.18 or higher (Ch. 10, pp. 312-324).
- The Capillaries Configuration Tool takes Length [mm] and Diameter [mm] for the generic sample loop, transfer capillary and ASM capillary, and calculates Volume [µL] automatically (Ch. 10, p. 314).
- Before a run the system can interactively and automatically flush both dimensions' flow paths and all sample loops (Ch. 1, p. 10).
Integration & Operating Notes
No other pump model may serve as the 2D pump in the cluster. If the existing system uses a 1260 pump in the second dimension, it must be replaced before the 2D-LC valve/loop front end can be commissioned, since the manual restricts the 2D pump to the 1290 Binary Pump or any 1290 Infinity II/III High-Speed Pump.
A started second-dimension analysis always runs to completion, so a second peak eluting from the first dimension can be lost if it is sampled while the 2D analysis is still running. Method development must match the second-dimension run time to the retention time between two first-dimension peaks, or move to multiple heart-cutting with loop storage.
Firmware files from different firmware sets must not be combined, and the power supply of the device and the PC must not be interrupted during a firmware update. Either action can corrupt the module firmware, so schedule updates on a stable supply and use one consistent firmware set across the cluster.
For 2D-LC instruments the checkout includes a pressure test of the 1D-LC part, a pressure test of the 2D-LC part, and a pump characteristic test that is Pump Ripple for a 1260 pump and Tuning for a 1290 pump. The correct test must be selected for the pump actually installed in each dimension.
Selection & Integration Q&A
Pick 2D-LC when one-dimensional resolution is insufficient and you need selective heart-cutting, multiple heart-cutting with up to 12 stored cuts, HiRes, LCxLC or ASM. The analytical and application-specific siblings are single-dimension systems; the bio sibling is the biocompatible variant of this same 2D concept.
The 2D pump must be a 1290 Binary Pump or any 1290 Infinity II/III High-Speed Pump. MassHunter Workstation must license the 2D-LC instrument, the license dongle sits in the 2D pump, and Lab Advisor 2.17+ (2.18+ for the capillary leak test) is needed for configuration and tests.
A started second-dimension run always completes, so a following 1D peak can be lost if sampled during it; match 2D run time to the 1D peak spacing or use multiple heart-cutting loops. Before a run, flush both dimensions' flow paths and all sample loops to reduce carryover.
The USB hardware dongle license, the optional 1D/2D switching valve, ASM capillaries and the analytical fraction collection hardware are options. Service work needs a 1/4-1/5 inch wrench and a blank nut, plus Lab Advisor 2.17/2.18; valve heads and Quick Change Valve Head parts are listed as maintenance parts.