Agilent Technologies 5300 Fragment Analyzer

Agilent Technologies 5300 Fragment Analyzer — Automated electrophoresis.

The 5300 Fragment Analyzer is a 48-capillary automated electrophoresis system for sizing, quantification and integrity assessment of IVT mRNA drug substances and LNP-encapsulated mRNA drug products. It resolves fragments differing by ~8 bp at 500 bp and sizes DNA from 100 to 6,000 bp, with sensitivity down to 5 pg/µL. It serves as the CQA workhorse between the lower-throughput 5200 and the ultra-high-throughput 5400 in the Fragment Analyzer series.

Agilent Technologies Fragment Analyzer Systems Series — Model Comparison & Selection

Model Measurement rangePlate capacityAverage sizing of 2 200 nt ivt mrna, ntLower marker size of hs rna kit, ntMaximum ivt mrna sample size analyzed, ntSample volume per well, microliter
5300 Fragment Analyzer Current ——2370156000—
5200 Fragment Analyzer ——23701560002
Fragment Analyzer Capillary Arrays —12———2
RNA QC, DNA QC, Fragment Analyzer —12———2
Ultra-high-throughput fragment analysis 5400 Fragment Analyzer 0.5nine plates2370156000—

Full specifications

Core Performance & Functional Specifications
Sensitivity 5 pg/µL
Average Sizing Of 2,200 Nt Ivt Mrna 2370 nt
Lower Marker Size Of Hs Rna Kit 15 nt
Maximum Ivt Mrna Sample Size Analyzed 6000 nt
Resolution 5 milligram
Sample types analyzed mfr IVT mRNA drug substances and LNP-encapsulated IV
Additional sample mfr 1
Instrument mfr 5300 system
LNP disruption mfr varying amounts of surfactant surfactant
Surfactant levels tested mfr 20
Sample concentration range mfr 1 ng/μL
Sizing range mfr 100...5,000 bp
Extended method mfr Method BE BE
Dilution ratios tested mfr 1
Sizing accuracy mfr 15
Quantitative range, smears mfr 100...5,000 pg/μL
Quantitative range, fragments mfr 5...500 pg/μL
Quantitative accuracy mfr 25
Quantitative range mfr 10...1,000 pg/μL
DNA sizing range mfr 100 bp
Separation resolution mfr 8 bp
DNA sizing precision mfr 2 % CV
DNA sizing accuracy mfr 5 better
DNA fragment concentration range mfr 0.005 DNA
IVT mRNA drug product sizes analyzed mfr 2,200 nt
Surfactant concentrations tested for LNP disruption mfr 20
Sample concentration range assessed mfr 1 ng/μL
Dilution ratios assessed with extended method mfr 1
DNA sizing precision1 mfr 2 % CV
DNA sizing accuracy1 mfr 5 better
Intercalating Dye volume (48 samples) mfr 2.5 μL
Gel volume (48 samples) mfr 25 mL
Intercalating Dye volume (96 samples) mfr 4.0 μL
Gel volume (96 samples) mfr 40 mL
Intercalating Dye volume (144 samples) mfr 5.5 μL
Gel volume (144 samples) mfr 55 mL
Intercalating Dye volume (192 samples) mfr 7.0 μL
Gel volume (192 samples) mfr 70 mL
Intercalating Dye volume (240 samples) mfr 8.5 μL
Gel volume (240 samples) mfr 85 mL
Inlet Buffer fill volume per well mfr 1.0 mL/well
Capillary Storage Solution fill volume per well mfr 100 μL/well
Marker plate fill volume per well mfr 30 μL/well
MDK DNA Ladder volume mfr 24 μL
disruption time mfr 20 minutes
sizing accuracy specification mfr 15 % error
sizing precision specification mfr 5 % CV
HS NGS kit sizing range mfr 100...6,000 bp
HS D5000 ScreenTape sizing range mfr 100...5,000 bp
HS D5000 limit of detection mfr 5 pg/uL
SPRI bead cleanup ratio mfr 0.8 x
spike-in fragment concentration mfr 50 pg/μL
library concentration mfr 500 pg/μL
Dimensions, Weight & Mechanical Design
Peak Width 40 seconds
Sample sizes (nucleotide length) mfr 2,200 nt
minimum peak height mfr 10 RFU
Standards, Safety & General Data
standard sizing upper limit mfr 6000 nt

Description

Construction and series positioning

Behavior across assays and media

Installation, connection and operation

Integration & Operating Notes

Process Limits
Dilute DNA samples into the 0.005–2 ng/µL input window

Samples outside the supported input concentration range must be diluted before loading; the stated sizing precision (2% CV) and accuracy (±5%) are only valid for standards initially prepared in 1x TE buffer, so buffer mismatch will degrade results even when concentration is in range.

Agilent CRISPR Discovery Gel Kit 5300 Fragment Analyzer System 48-Capillary Array Quick Guide, Kit Specifications, page 1
Control & I/O
Match the software method to the installed array variant

The CRP-910-(33 or 55) – CRISPR Discovery method must be selected from the software dropdown to match the 33 or 55 array variant; running the wrong method for the installed array will produce invalid sizing and should be treated as a run-setup interlock before queuing samples.

Agilent CRISPR Discovery Gel Kit 5300 Fragment Analyzer System 48-Capillary Array Quick Guide, Fragment Analyzer software operating procedure, page 4
Maintenance
Scale gel/dye mixture to the actual sample count

Gel/dye mixture volumes are tabulated from 48 samples (2.5 µL dye + 25 mL gel) to 240 samples (8.5 µL dye + 85 mL gel); preparing the wrong scale wastes kit reagents or starves the array, and one well per separation is typically reserved for the ladder.

Agilent CRISPR Discovery Gel Kit 5300 Fragment Analyzer System 48-Capillary Array Quick Guide, Gel preparation, page 2

Selection & Integration Q&A

When should I choose the 5300 over the 5200 or 5400 Fragment Analyzer?

Pick the 5300 when your throughput fits a 48-capillary array and you need IVT mRNA/LNP CQA plus DNA sizing to 6,000 bp at 8 bp resolution near 500 bp. The 5200 suits lower throughput; the 5400 is the ultra-high-throughput variant for larger batch campaigns.

What consumables and adjacent parts must be in place before a 5300 run?

You need the 48-capillary array, the assay kit (e.g., CRISPR Discovery Gel Kit DNF-810 with gel, intercalating dye, inlet buffer, conditioning solution, markers, ladder, dilution buffer, mineral oil), a marker plate with mineral oil overlay, and the matching software method for the 33 or 55 array.

What environmental and duty-cycle limits apply to the 5300 in routine operation?

Inlet buffer must be replaced daily and capillary storage solution every 2–4 weeks; marker wells need a mineral oil overlay to prevent evaporation. Gel/dye mixture must be prepared fresh and scaled to sample count, so plan runs around daily buffer refresh rather than multi-day unattended campaigns.

What maintenance schedule and non-included accessories should I budget for?

Budget daily inlet buffer replacement (1.0 mL/well, rows A–D), capillary storage solution refresh every 2–4 weeks (100 µL/well, tray 3), and periodic capillary array replacement. The kit lists one item sold separately; the MDK DNA Ladder is used undiluted at 24 µL in well D12 or H12.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Not for use in diagnostic procedures
Datasheet ·4 pp.
View PDF
PDF
CQA Assessment of LNP-mRNA Drug Products
Brochure ·381,9 КБ ·7 pp.
View PDF
PDF
Detection of Adapter Dimers in NGS
Brochure ·1,5 МБ ·8 pp.
View PDF