6560 Ion Mobility QTOF LC/MS, Collision Cross Section drawing
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 6560 Ion Mobility QTOF LC/MS, Collision Cross Section

Agilent Technologies 6560 Ion Mobility QTOF LC/MS, Collision Cross Section — Liquid chromatography mass spectrometry lc ms with Pumping speed 1.6×10⁻⁸ m³/h.

The Agilent 6560 ion mobility LC/Q-TOF is a hybrid quadrupole time-of-flight mass spectrometer with an 80 cm drift tube for ion mobility separation, configured here with a nanospray source for low-flow proteomics. It covers 100–1,700 m/z in MS mode and 50–1,700 m/z in ion mobility mode, distinguishing it from the 6530 Q-TOF and 6230 TOF siblings by adding gas-phase mobility dimension. Primary use is structural and phosphopeptide characterization in LC/MS workflows.

Pumping speed
1.6×10⁻⁸ m³/h

Agilent Technologies Quadrupole Time Of Flight LC MS Series — Model Comparison & Selection

Model Pumping speed, m³/hMass rangeAcquisition rate, HzAutosampler temperature, °CColumn temperature, °CSample volume, milliliter
6560 Ion Mobility QTOF LC/MS, Collision Cross Section Current 1.6×10⁻⁸100...1,70010445—
6545XT AdvanceBio UHPLC QTOF, biopharmaceutical analysis 3×10⁻⁵100...10,0001.04605×10⁻⁴
Electron capture dissociation for QTOF LCMS of proteins 0.024—15600.01

Full specifications

Core Performance & Functional Specifications
Pumping speed 266 nL/min
Acquisition Rate 10 spectra/s
Column temperature 45 °C
Instrument mfr 6560 TOF
Ionization source mfr Agilent nanospray source source
Sample mfr BSA BSA
Sample concentration mfr 20 mg/mL
Needle dimensions mfr 360 length
Monomer m/z (charge 1) mfr 66435 m/z
Dimer m/z (charge 1) mfr 132869 m/z
Trimer m/z (charge 1) mfr 199303 m/z
Tetramer m/z (charge 1) mfr 265737 m/z
Collision energy slope mfr 2
Collision energy offset mfr 3
Max precursors per cycle mfr 3
Threshold for MS/MS mfr 1000 counts
Isotope model mfr Peptides Peptides
PhosphoMix dilution concentration mfr 6.66 pmol/μL
Injection volume, digested α-casein mfr 11 μL
Injection volume, digested β-casein mfr 10 μL
Injection volume, phosphopeptide enrichment eluate mfr 5 μL
Injection volume, PhosphoMixes mfr 2 μL
Dimensions, Weight & Mechanical Design
Mass range 100...1,700 m/z
Drift Tube Length 80 cm
Ion mobility mode mass range mfr 100...1,700 m/z
Isolation width mfr 4 m/z
Electrical, Power & Drive Specifications
Mobile phase 1 mfr 150 acetate
Mobile phase 2 mfr 0.1 acid
High pressure funnel voltage mfr 200 peak
Trapping funnel voltage mfr 200 peak
High pressure funnel RF mfr 100 V
Trap funnel RF mfr 100 V
Drift tube exit voltage mfr 250 V
Operating Conditions, Environment & Protection
Autosampler Temperature 4 °C

Description

Architecture and series positioning

Behavior and workflow facts

Installation and operation

Integration & Operating Notes

Process Limits
Ion mobility mode narrows the usable mass range

When the drift-tube separation is engaged the documented mass range becomes 50–1,700 m/z rather than the 100–1,700 m/z of standard MS acquisition, so method mass windows must be set against the mode actually used.

Factory Specification Sheet; Application Note 5994-1617EN
Electrical
Funnel and drift-tube voltages are method-set, not fixed

Documented values include high-pressure funnel RF 100 V, trap funnel RF 100 V, drift tube exit 250 V, and 200 V funnel voltages in the BSA method; these are tuned per method and must be carried into the acquisition method rather than assumed as defaults.

Application Note 5994-1554EN
Control & I/O
Auto-MS/MS gating depends on threshold and precursor cap

The documented phosphorylation method triggers MS/MS only above 1000 counts, allows at most 3 precursors per cycle, and uses a 4 m/z isolation width with collision energy slope 3.1 and offset 3.6; changing any of these alters which species are fragmented.

Application Note 5994-1617EN
Maintenance
Nanospray source and Nanodapter are separate orderable items

The nanospray source (G1992A) and Infinity UHPLC Nanodapter (G1988A) are distinct accessories cited in the workflows; they are not part of the base instrument delivery and must be specified separately when configuring a nanospray system.

Application Notes 5994-1554EN and 5994-1617EN

Selection & Integration Q&A

When should I choose the 6560 ion mobility LC/Q-TOF over the 6530 Q-TOF or 6230 TOF siblings?

Pick the 6560 when gas-phase ion mobility separation is required, e.g. resolving BSA multimers or phosphopeptide isomers; its 80 cm drift tube adds a dimension the 6530 Q-TOF lacks. The 6230 TOF covers 800–8,000 m/z for large-mass work, and the 6475/6495 triple quads suit targeted quantitation instead.

What adjacent LC and sample-prep hardware does the documented nanospray workflow require?

The phosphorylation workflow uses an Infinity II 1290 binary pump, Infinity UHPLC Nanodapter (G1988A), nanospray source (G1992A), and AssayMAP Bravo for sample prep, with 150 mM ammonium acetate and 0.1% formic acid mobile phases at 266 nL/min.

Which accessories are not included in the base instrument and must be ordered separately?

The nanospray ion source (G1992A) and Infinity UHPLC Nanodapter (G1988A) are cited as separate accessories in the application workflows, along with the AssayMAP Bravo sample-prep system; none are part of the base 6560 delivery and must be specified in the configuration.

Dimensional Drawings & Technical Documentation

Dimensional drawings

6560 Ion Mobility QTOF LC/MS, Collision Cross Section — drawing 1 6560 Ion Mobility QTOF LC/MS, Collision Cross Section — drawing 2 6560 Ion Mobility QTOF LC/MS, Collision Cross Section — drawing 3
performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation