Atmospheric Pressure Chemical Ionization Source, APCI drawing
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Atmospheric Pressure Chemical Ionization Source, APCI

Agilent Technologies Atmospheric Pressure Chemical Ionization Source, APCI — Liquid chromatography mass spectrometry lc ms with Pumping speed 2.4×10⁻⁵ m³/h.

The Agilent APCI source is an atmospheric-pressure chemical ionization interface for LC/MS systems, extending the series beyond ESI to ionize low-polarity and non-polar analytes. It operates with drying gas at 350 °C and vaporizer at 450 °C, supporting mass ranges up to 7000 m/z. This modification is selected when ESI fails to ionize compounds lacking acidic or basic sites, enabling LC/MS analysis of steroids, lipids, and similar species.

Pumping speed
2.4×10⁻⁵ m³/h

Agilent Technologies LC MS Ion Sources Series — Model Comparison & Selection

Model Pumping speed, m³/hSample volume, milliliterColumn temperature, °CDrying gas temperature, °CSheath gas flow, liter/minuteSheath gas temperature, °C
Atmospheric Pressure Chemical Ionization Source, APCI Current 2.4×10⁻⁵0.00455350——
Electrospray ion Source, ESI 2.4×10⁻⁵0.0018029012400
High sensitivity MS, high sensitivity ion source, Jet Stream ————0.011350

Full specifications

Core Performance & Functional Specifications
Pumping speed 0.4 mL/min
Analysis time 14 min
Column temperature 55 °C
Sample volume 4 µL
Gradient mfr 55 % B
Maximum counts mfr 1.2 counts
MRPL concentration mfr 2 ng/mL
Reconstitution volume mfr 100 μL
Dimensions, Weight & Mechanical Design
Mass range 7000 m/z
Mass accuracy mfr 3 ppm
Mass accuracy achieved mfr 2 ppm
Electrical, Power & Drive Specifications
Fragmentor Voltage 150 V
Mobile phase A mfr 0.1 acid/water
Mobile phase B mfr Methanol Methanol
Photomultiplier voltage increase mfr 50 V
Operating Conditions, Environment & Protection
Drying Gas Temperature 350 °C
Standards, Safety & General Data
Column part number mfr 822700

Description

Construction and series positioning

Behavior under operating conditions

Installation and operation

Integration & Operating Notes

Process Limits
Vaporizer temperature must not exceed 450 °C

Operating the vaporizer above 450 °C may cause thermal degradation of analytes or damage to the source components. The recommended setting for this application is 450 °C; ensure the temperature is controlled and monitored via the MS software to avoid overheating.

Application Note 5989 4738EN, MS Conditions
Process Limits
Drying gas temperature set to 350 °C for optimal desolvation

The drying gas temperature directly affects desolvation efficiency and signal intensity. For this method, 350 °C is specified; lower temperatures may result in incomplete desolvation and higher background, while higher temperatures could cause analyte degradation. Adjust only with method validation.

Application Note 5989 4738EN, MS Conditions
Maintenance
Regular cleaning of the corona discharge needle is required

The corona discharge needle can become contaminated over time, leading to reduced sensitivity and unstable ionization. Periodic cleaning according to the instrument maintenance manual is necessary. Frequency depends on sample load and matrix; inspect visually and clean when performance degrades.

LC/MS Instrument Maintenance Manual (general practice

Selection & Integration Q&A

When should I choose the APCI source over the ESI source for my LC/MS system?

Choose APCI for low-polarity, non-ionic compounds (e.g., steroids, lipids) that ionize poorly by ESI. ESI is preferred for polar, ionic analytes. APCI is also less susceptible to matrix effects for some compounds. The APCI source is compatible with Agilent LC/MS platforms such as the 6230 TOF or 6475 triple quadrupole.

What are the gas and power requirements for installing the APCI source?

The APCI source requires dry nitrogen or zero-air for drying and nebulizer gas, typically supplied at 350–450 °C. It also needs a corona discharge power supply, which is integrated into the MS. Ensure gas purity and pressure meet the MS specifications. Electrical connections are made via the MS interface; no external high-voltage supply is needed.

What are the environmental and duty-cycle limitations for the APCI source?

Operate within the MS system's ambient temperature and humidity limits (typically 15–30 °C, <80% RH). The source is designed for continuous operation but requires stable gas flows and temperatures. Avoid frequent power cycling; allow stabilization after temperature changes. Duty cycle is limited by sample throughput and maintenance needs.

What maintenance items and consumables are not included with the base APCI source?

The base delivery includes the source assembly but not the corona discharge needle, nebulizer, or gas filters, which are consumables and should be stocked. Regular maintenance includes cleaning the needle, replacing gas filters, and checking seals. Refer to the maintenance schedule in the MS manual for intervals.