Agilent Technologies Bond Elut Carbon S & Carbon S/NH2

Agilent Technologies Bond Elut Carbon S & Carbon S/NH2 — Sample preparation with Pumping speed 0.024 m³/h.

Bond Elut Carbon S is a 150 mg SPE cartridge combining carbon and NH2 sorbents for post-extraction cleanup of PFAS from complex matrices. It is validated for 59 target compounds in soil and egg matrices, with average recovery of 99.2% and RSD of 8.5% across all compounds. The cartridge is used upstream of LC/MS/MS to remove matrix interferences while preserving PFAS analytes.

Pumping speed
0.024 m³/h

Agilent Technologies Bond Elut Series — Model Comparison & Selection

Model Pumping speed, m³/hSample volume, milliliterDrying gas temperature, °CAverage recovery all compounds, %Cartridge sorbent mass, mgCentrifugation speed, rpm
Bond Elut Carbon S & Carbon S/NH2 Current 0.0240.00523099.21501900
Bond Elut HLB | SPE Sorbent 1.8×10⁻⁵0.0024073.4...118500—
Bond Elut Lipid Extraction 3...5—30———
Bond Elut LMS | Large Molecule Size —0.005———2800

Full specifications

Core Performance & Functional Specifications
Pumping speed 0.400 mL/min
Average Recovery (all Compounds) 99.2 %
Centrifugation Speed 1900 rpm
Centrifugation Time 10 minutes
Column temperature 30 °C
Drying Gas Flow 4 L/min
Extraction Solvent Volume 10 mL
Nebulizer Gas Pressure 15 psi
Rsd (all Compounds) 8.5 %
Sample Amount 2 g
Sample volume 5 µL
Sheath Gas Flow 12 L/min
Target Spiking Solution Concentration 250 ng/mL
Number of target compounds mfr 59
Polarity mfr Negative Negative
Calibration levels mfr Six Six
Recovery range (high-level spikes) mfr 70...130
RSD (high-level spikes) mfr 30
Ionization polarity mfr Negative Negative
PFOS maximum level in eggs mfr 1.0 μg/kg
PFOA maximum level in eggs mfr 0.30 μg/kg
PFNA maximum level in eggs mfr 0.70 μg/kg
PFHxS maximum level in eggs mfr 0.30 μg/kg
Sum of four PFAS maximum level in eggs mfr 1.7 μg/kg
Limit of quantification for PFOS in eggs mfr 0.30 μg/kg
Limit of quantification for PFOA in eggs mfr 0.30 μg/kg
Limit of quantification for PFNA in eggs mfr 0.30 μg/kg
Limit of quantification for PFHxS in eggs mfr 0.30 μg/kg
PFPeA calibration range mfr 0.15...2.0 μg/kg
PFBS calibration range mfr 0.13...1.8 μg/kg
PFHxA calibration range mfr 0.15...2.0 μg/kg
HFPO-DA calibration range mfr 0.3...4.0 μg/kg
PFOA calibration range mfr 0.15...2.0 μg/kg
PFOS calibration range mfr 0.14...1.92 μg/kg
PFNA calibration range mfr 0.15...2.0 μg/kg
PFOSA calibration range mfr 0.3...4.0 μg/kg
PFHpS recovery mfr 95.5
PFOSA recovery mfr 109.0
PFDS recovery mfr 107.4
Recovery range mfr 70...130
RSD high-level spikes mfr 30
Target compounds mfr 59
Carbon S for PFAS cartridge, 250 mg, 6 mL mfr 5610
Polypropylene autosampler screw top vials, 2 mL, and caps mfr 5191
Centrifuge tubes and caps, 15 mL mfr 5610
InfinityLab PFC delay column, 4.6 × 30 mm mfr 5062
ZORBAX RRHD Eclipse Plus C18 column, 2.1 × 100 mm, 1.8 μm mfr 959758
Vac Elut SPS 24 manifold with collection rack for 10 × 75 mm test tubes mfr 12234003
Collection rack and funnel set for 12 or 15 mL conical tubes, for Vac Elut SPS 24 manifold mfr 12234027
diSAmPAP mfr < MRL MRL
DONA mfr < MRL MRL
EtFOSE mfr < MRL MRL
FBSA mfr < MRL MRL
FDSA mfr < MRL MRL
FHxSA mfr < MRL MRL
HFPO-DA mfr < MRL MRL
MeFOSE mfr < MRL MRL
N-EtFOSA mfr < MRL MRL
N-EtFOSAA mfr < MRL MRL
NFDHA mfr < MRL MRL
N-MeFOSA mfr < MRL MRL
N-MeFOSAA mfr < MRL MRL
PFBA mfr 4.5 ng/g
PFBS mfr < MRL MRL
PFDA mfr < MRL MRL
PFDoDA mfr < MRL MRL
PFDoS mfr < MRL MRL
PFDS mfr < MRL MRL
ammonia concentration in methanol mfr 1
tumble time mfr 1 hour
cartridge rinse volume mfr 5 mL
acetic acid neutralization volume mfr 50 μL
extract aliquot volume mfr 790 μL
purified water volume mfr 200 μL
Dimensions, Weight & Mechanical Design
Cartridge Sorbent Mass 150 mg
Operating Conditions, Environment & Protection
Drying Gas Temperature 230 °C
Sheath Gas Temperature 250 °C
Standards, Safety & General Data
isotope performance standard volume mfr 10 μL

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection, and operation

Integration & Operating Notes

Process Limits
Pumping speed must be controlled at 0.400 mL/min

Deviating from the validated 0.400 mL/min flow changes analyte residence time on the carbon sorbent and can shift recoveries outside the 70–130% acceptance range. Use a calibrated syringe or peristaltic pump and verify flow before each batch.

Application Note 5994-4770EN, Method Section
Process Limits
Do not exceed 2 g sample load per cartridge

The method is validated for 2 g sample amount; higher loads can overload the 150 mg sorbent bed, causing breakthrough of PFAS and matrix interference in the LC/MS/MS. For higher sample masses, split across multiple cartridges or use a larger sorbent mass.

Application Note 5994-4770EN, Sample Preparation
Maintenance
Centrifuge eluate at 1900 rpm for 10 minutes before analysis

After SPE, the eluate must be centrifuged at 1900 rpm for 10 minutes to remove residual particulates that would otherwise foul the LC column and nebulizer. Skipping this step risks clogging and carryover in the LC/MS/MS.

Application Note 5994-4770EN, Post-Extraction Cleanup
Control & I/O
LC/MS/MS interface settings are method-specific

The validated method uses drying gas 4 L/min at 230 °C, sheath gas 12 L/min at 250 °C, nebulizer 15 psi, nozzle voltage 0 V, column 30 °C, and negative polarity. These settings are optimized for PFAS; re-optimize if using a different instrument or source geometry.

Application Note 5994-7358EN, LC/MS/MS Parameters

Selection & Integration Q&A

When should I choose Bond Elut Carbon S instead of the standard Bond Elut Carbon or C18 for PFAS cleanup?

Choose Carbon S for PFAS in complex matrices (soil, eggs) where both carbon and NH2 cleanup are needed; the 'S' variant is validated for 59 PFAS with 99.2% average recovery. Standard Carbon lacks the S modification; C18 is for non-polar analytes and does not provide the same PFAS recovery.

What LC/MS/MS conditions and adjacent system parts are required for this cartridge method?

The method uses an Agilent 6475 triple quadrupole LC/MS with negative polarity, drying gas 4 L/min at 230 °C, sheath gas 12 L/min at 250 °C, nebulizer 15 psi, nozzle voltage 0 V, column 30 °C, and 5 µL injection. A centrifuge capable of 1900 rpm for 10 minutes is needed post-SPE.

What are the mounting and duty-cycle limits for using this cartridge in a batch workflow?

The cartridge is used on a standard SPE vacuum manifold; no electrical connection. Each cartridge is single-use; the validated method uses 2 g sample and 10 mL extraction solvent at 0.400 mL/min. Centrifugation at 1900 rpm for 10 minutes is required after elution. No duty-cycle limit is specified beyond single-use.

What maintenance and consumables are needed beyond the cartridge itself?

The base delivery is the cartridge only; the method requires LC/MS-grade solvents for the 10 mL extraction, a centrifuge for 1900 rpm/10 min, and an SPE manifold. No replacement parts are listed; the cartridge is single-use. LC column and MS source maintenance follow standard PFAS methods.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Per- and Polyfluoroalkyl Substances Analysis in Soil
Manual ·874,8 КБ ·11 pp.
View PDF