Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies FRG-700/702

Agilent Technologies FRG-700/702 — Vacuum technologies with Max pressure 1 megapascal.

The FRG-700/702 is a combined Pirani and inverted-magnetron cold-cathode gauge covering 5×10⁻⁹ to 1000 mbar with a single logarithmic 0.6 V/decade output. The FRG-700 uses FPM seals for elastomer-sealed KF/CF installations; the FRG-702 is the all-metal, 150 °C bakeable variant. It serves as the wide-range pressure sensor in rough-to-UHV pumping systems.

Max pressure
1 megapascal

Agilent Technologies Active Gauges Series — Model Comparison & Selection

Model Measurement rangeOperating temperature, °CStorage temperature, °CInternal volume, microliterOperationResponse time, ms
FRG-700/702 Current 1.82...8.6540202910
PCG 752 510204.71030

Full specifications

Core Performance & Functional Specifications
Max pressure 10 bar
Measurement range 1.82...8.6 V
Diffusion Pump Pumping Speed 50 L/s
Electrical connection 28
Internal Volume 20 cm³
Operation 29
Response Time 10 ms
Vacuum connection stainless steel
Vacuum gauge measurement range mfr 6 pressure
Dry scroll pump typical size mfr 3...30 h
Turbomolecular pump pumping speed mfr 70...3 L/s
Measuring range (air, N2) mfr -9...1000 mbar
Accuracy (N2) mfr 30
Reproducibility mfr 5
Error signal mfr 0.5 supply
Output impedance mfr 2
Minimum loaded impedance mfr 10 proof
Gauge identification mfr 85 common
Adjustment <HV> potentiometer mfr 10 mbar
Adjustment <ATM> potentiometer mfr at atmospheric pressure pressure
Sensor cable conductors mfr 8 shielding
Grounding concept: supply common-signal common mfr 10
Materials exposed to vacuum: Vacuum connection mfr stainless steel steel
Materials exposed to vacuum: Measuring chamber mfr stainless steel steel
Materials exposed to vacuum: Feedthrough mfr ceramic ceramic
Materials exposed to vacuum: Internal seals mfr Ag, Cu, soft solder solder
Materials exposed to vacuum: Anode mfr Mo Mo
Materials exposed to vacuum: Ignition aid mfr stainless steel steel
Materials exposed to vacuum: Pirani measuring tube mfr Ni, Au Au
FRG700KF25 mfr 25 KF
FRG702KF25 mfr 25 KF
FRG700CF35 mfr 40 C
FRG702KF40 mfr 40 KF
FRG702CF35 mfr 40 C
p > 10-6 mbar mfr 10 ms
p = 10-8 mbar mfr 1000 ms
p > 10-2 mbar mfr Pirani-only mode mode
p < 10-2 mbar mfr Cold cathode ignited ignited
Lamp mfr 6
Fuse1) mfr 1 AT
Adjustment mfr 10 mbar
Sensor cable mfr 68 poles
Vacuum connection-signal common mfr see paragraoh Electrical Connection Connection
Supply common-signal common mfr 10
Measuring chamber mfr stainless steel steel
Feedthrough mfr ceramic ceramic
Internal seals mfr 75
Anode mfr Ag, Cu, soft solder solder
Ignition aid mfr Mo Mo
Pirani measuring tube mfr stainless steel steel
Product Description mfr 16
Technical Specifications mfr 17
FRG-700 and FRG-702 Outline mfr 20
Installation mfr 24
Removing the Magnet Unit (only for Gauges with CF Flanges) mfr 27
Use With Another Control Device mfr 28
Adjusting the Gauge mfr 32
Deinstallation mfr 34
Cleaning FRG-700 mfr 35
Replacing the Pirani element mfr 36
Cleaning FRG-700, Replacing Parts mfr 36
Disassembling FRG-700 mfr 37
Reassembling FRG-700 mfr 38
Cleaning FRG-702 mfr 39
Cleaning FRG-702, Replacing Parts mfr 40
Disassembling FRG-702 mfr 41
Reassembling FRG-702 mfr 42
Troubleshooting mfr 44
Spare Parts mfr 45
p < 1x10-2 mbar mfr Cold cathode ignited ignited
About this manual mfr 6
Validity mfr 6
Definitions and terminology mfr 7
Definition of Caution, Warning and Note mfr 7
Warning Symbols mfr 8
Safety mfr 10
Proper use mfr 10
Improper use mfr 11
Safety guideline for FRG-700 and FRG-702 mfr 12
Disposal mfr 13
Service mfr 14
Response time p = 10-8 mbar mfr 1000 ms
Dimensions, Weight & Mechanical Design
Weight 700 g
Sensor cable line length mfr 50 m
Weight FRG702KF25 mfr 730 g
Weight FRG700CF35 mfr 980 g
Weight FRG702KF40 mfr 750 g
Weight FRG702CF35 mfr 1010 g
Line length mfr 8 m
Mounting orientation mfr any any
Electrical, Power & Drive Specifications
Power consumption 2 W
Voltage vs. pressure mfr 0.6 decade
Supply voltage at the gauge mfr 15.0...30.0 VDC
Fuse mfr 1 AT
Voltage at the supply unit with maximum line length mfr 16.0...30.0 VDC
Operating voltage mfr 3.3 kV
Operating current mfr 500 μA
Grounding concept voltage differential with respect to safety mfr 50 V
Grounding concept voltage differential with respect to accuracy mfr 10 V
Interfaces, Connections & Control
Turbomolecular pump controller protection class mfr 54
Output signal measuring range mfr 1.82...8.6 V
Sensor cable connector mfr 68 poles
Use With an Agilent Controller mfr 28
Materials, Media & Process Compatibility
Gas type dependence mfr 21
Pirani filament mfr Ni, Au Au
Operating Conditions, Environment & Protection
Bakeout Temperature 150 °C
Operating temperature 5 °C
Protection rating 40 IP
Storage temperature 40 °C
Relative humidity mfr 80 °C

Description

Construction and series positioning

Behavior across the range

Installation, control and operation

Integration & Operating Notes

Electrical
Use SELV supply and fuse the 15–30 VDC line at ≤1 AT

The gauge draws ≤2 W but the cold-cathode circuit runs at up to 3.3 kV / 500 µA internally, so the feed must be a SELV source with ≤1 Vpp ripple and a ≤1 AT fuse. At the full 50 m cable length, voltage drop means the supply unit must be set to 16.0–30.0 VDC, not the 15.0 V minimum at the gauge.

User Manual, Technical Specifications, page 18
Mounting
KF flanges need a conductive metallic clamping ring for grounding

With KF connections, protective grounding per EN 61010 is only achieved through a conductive metallic clamping ring; elastomer-sealed KF joints also cannot withstand more than 2.5 bar. CF flange variants meet the grounding requirement inherently and are the safer choice where overpressure or fault currents are possible.

User Manual, Vacuum Connection, page 24
Process Limits
Avoid sustained operation between 10⁻⁴ and 10⁻² mbar

Continuous running in this decade contaminates the cold-cathode electrode and shortens uptime; below 1×10⁻⁶ mbar the gauge can operate more than a year without cleaning. Mount at a low partial pressure of pollutants and use geometric protection (screens, elbows) if the process crosses this band regularly.

User Manual, Operation, page 31
Process Limits
Do not start the gauge below 3×10⁻⁹ mbar

At pressures below 3×10⁻⁹ mbar the electronics cannot distinguish whether the cold-cathode discharge has ignited, so the reading and status output are unreliable. Sequence the start-up so pressure is slightly above this threshold, or leave the gauge permanently powered so ignition delay drops below 1 s.

User Manual, Operation, page 30; Troubleshooting, page 44
Maintenance
Isolate the gauge before helium leak testing

Helium entering the gauge during a tightness test can cause electric arcing that destroys the electronics. Put the product out of operation and remove the electronics unit before any helium leak test; after sensor reassembly, verify anode-pin centering (≤0.5 mm eccentricity) and leak-test to <10⁻⁹ mbar l/s.

User Manual, Vacuum Connection, page 25; Reassembling FRG-700, page 38
Maintenance
Pirani circuit is field-adjustable; cold cathode is not

After sensor exchange, extreme temperature exposure, aging or contamination, only the Pirani side can be readjusted via the <HV> and <ATM> potentiometers; the cold-cathode circuit is factory-calibrated. Adjustment has negligible effect between roughly 10⁻² and 10² mbar, so calibrate at the appropriate set points.

User Manual, Adjusting the Gauge, page 32

Selection & Integration Q&A

When should I choose the FRG-702 over the FRG-700, and when is a sibling like the FRG-720/730 the better fit?

Pick FRG-702 (all-metal, copper-sealed Pirani element) when the flange must be baked to 150 °C or operating temperature exceeds +55 °C; FRG-700 with FPM seals is limited to +5…+55 °C. Choose FRG-720/730 if you need Bayard-Alpert Pirani behavior at the high-vacuum end; CDG-500 for high-precision diaphragm measurement.

What supply, cabling and controller does the gauge need when it is not run from an Agilent AGC-100 or Turbo AG rack?

Provide a SELV 15.0–30.0 VDC supply (16.0–30.0 VDC at the supply unit for 50 m cable), ≤1 Vpp ripple, ≤1 AT fuse. Use an 8×0.14 mm² shielded cable with FCC68 8-pole female connector; for third-party controllers, build the sensor cable to the manual wiring diagram and read the signal between pins 3 and 5.

What maintenance and spare parts should be planned, and what is not in the base delivery?

Plan cleaning when operating in 10⁻⁴…10⁻² mbar; below 10⁻⁶ mbar over a year is possible without cleaning. Order kits by number: FRG700MAINT/FRG700REPR/FRG700IGN for FRG-700, FRG702MAINT/FRG702REPR for FRG-702, plus FRG700MEAS25/40 or FRG702MEAS25/40 measuring systems. Sensor cable, controller and clamping rings are not included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
FRG-700 and FRG-702 Vacuum Gauge User Manual
Manual ·51 pp.
View PDF