Agilent Technologies Ion Sources for Gas Chromatography/Mass Spectrometry
Agilent Technologies Ion Sources for Gas Chromatography/Mass Spectrometry — Gas chromatography mass spectrometry gc ms.
The Agilent JetClean ion source system (G7077) is a hydrogen‑plasma cleaning accessory for Agilent GC/MS ion sources, operating in Acquire & Clean or Clean Only mode. It supports source temperatures of 150–350 °C, emission currents of 10–35 µA, and hydrogen flows of 0.13–0.49 mL/min (Acquire & Clean) or 0.13–3.52 mL/min (Clean Only). It is used to remove contamination in situ, extending source maintenance intervals and preserving attogram‑level detection limits.
Agilent Technologies GC MS Ion Sources Inlets Series — Model Comparison & Selection
| Model | Detection limit | Clean duration lower setpoint, min | Clean duration upper setpoint, min | Emission current lower setpoint, microampere | Emission current upper setpoint, microampere | Hydrogen flow acquire clean mode lower setpoint, mL/min |
|---|---|---|---|---|---|---|
| Ion Sources for Gas Chromatography/Mass Spectrometry Current | attogram | 1 | 120 | 10 | 35 | 0.13 |
| Maintenance free, JetClean ion source for GC/MS instrument | — | 1 | 120 | 10 | 35 | 0.13 |
Full specifications
| Detection limit | attogram |
| Clean Duration, Lower Setpoint | 1 min |
| Clean Duration, Upper Setpoint | 120 min |
| Hydrogen Flow (acquire & Clean Mode), Lower Setpoint | 0.13 mL/min |
| Hydrogen Flow (acquire & Clean Mode), Upper Setpoint | 0.49 mL/min |
| Hydrogen Flow (clean Only Mode), Lower Setpoint | 0.13 mL/min |
| Hydrogen Flow (clean Only Mode), Upper Setpoint | 3.52 mL/min |
| Ionization source mfr | low energy capable electron ionization ionizatio |
| Hydrogen flow step size mfr | 0.069 sccm |
| Mass analyzer type mfr | single quadrupole quadrupole |
| Emission Current, Lower Setpoint | 10 µA |
| Emission Current, Upper Setpoint | 35 µA |
| Source Temperature, Lower Setpoint | 150 °C |
| Source Temperature, Upper Setpoint | 350 °C |
Description
Construction and series positioning
- The JetClean system is a hydrogen‑plasma cleaning module that mounts to the ion source of Agilent MS systems (5975/5977 series, 7000E/7010C).
- It is not a standalone mass analyzer; it is an accessory that integrates with the ion source of the host GC/MS. The sibling models listed (5977C GC/MSD, 7000E, 7010D, 7250 GC/Q‑TOF) are host instruments, not variants of the JetClean itself.
- The system provides two operating modes: Acquire & Clean (cleaning during acquisition) and Clean Only (dedicated cleaning).
Behavior under load and in different modes
- In Acquire & Clean mode, hydrogen flow is limited to 0.13–0.49 mL/min, allowing cleaning while the MS continues to acquire data.
- In Clean Only mode, hydrogen flow can be raised to 0.13–3.52 mL/min for more aggressive cleaning when the MS is not acquiring.
- Cleaning duration is programmable from 1 to 120 min. Emission current is set between 10 and 35 µA, and source temperature between 150 and 350 °C.
- The system is intended for low‑energy‑capable electron ionization sources; the host mass analyzer may be a triple quadrupole (as in the 7000E/7010D) or other Agilent MS types.
Installation, connection, control and operation
- The JetClean system requires a hydrogen supply with a flow step size of 0.069 sccm (as listed in the factory specifications).
- Configuration and parameter setpoints are managed through the host MS software (MassHunter suite is referenced in the brochure).
- The operating manual (G7077‑90052 and G7077‑92052) covers configuration, hydrogen supply plumbing, and troubleshooting. No ordering codes or part numbers for the JetClean itself are printed in the digests.
- The system is designed for in‑situ cleaning, reducing the need to vent the MS and manually clean the source.
Integration & Operating Notes
In Acquire & Clean mode the hydrogen flow must stay within 0.13–0.49 mL/min; in Clean Only mode it may be raised to 0.13–3.52 mL/min. Exceeding the Acquire & Clean limit while the MS is acquiring can perturb the ion source and affect data quality. The flow step size is 0.069 sccm, so setpoints should be chosen as multiples of this step.
The emission current setpoint range is 10–35 µA. Operating outside this range may not sustain a stable plasma or may damage the source. The current must be set in conjunction with the source temperature (150–350 °C) to achieve effective cleaning without overheating the ion source.
The JetClean system uses hydrogen, which is flammable. The operating manual covers hydrogen supply plumbing and troubleshooting. Before initial use and after any plumbing change, verify all connections for leaks. Use only hydrogen‑compatible tubing and fittings. The manual does not specify a maintenance interval for the JetClean itself, but regular checks of the hydrogen supply are implied.
Selection & Integration Q&A
The JetClean is an accessory for in‑situ source cleaning, not a standalone GC/MS. Choose it when you already have a compatible Agilent MS (5975/5977, 7000E/7010C) and want to extend source maintenance intervals. The 5977C, 7000E, 7010D, and 7250 are complete instruments; select those when you need a full GC/MS system, not just cleaning capability.
The JetClean requires a hydrogen supply with a flow step size of 0.069 sccm and must be plumbed to the ion source region. The operating manual (G7077‑90052) covers hydrogen supply plumbing and configuration. The system is controlled through the host MS software (MassHunter). Ensure the hydrogen supply is regulated and leak‑tested before connection.
The base delivery includes the JetClean module and documentation, but hydrogen supply plumbing, tubing, and fittings are not included and must be sourced separately. The manual covers troubleshooting and hydrogen supply plumbing but does not specify a regular maintenance schedule for the JetClean itself. Periodic leak checks of the hydrogen supply are recommended.