Agilent Technologies 1290 Infinity II Bio LC System
Agilent Technologies 1290 Infinity II Bio LC System — Liquid chromatography with Pumping speed 3×10⁻⁴ m³/h.
The 1290 Infinity II Bio LC System is a biocompatible UHPLC platform built from stackable modules — Bio pump, Bio/Hybrid Multisampler, Multicolumn Thermostat with Bio heat exchanger, and Bio flow-cell detectors. It operates to 1300 bar at analytical flow rates up to 5 mL/min, with column thermostating from 20 °C below ambient to 110 °C across up to eight columns. It serves high-resolution bio-separations where inert flow-path parts and high-pressure performance are both required.
Full specifications
| Pumping speed | 5 mL/min |
| Analytical Flow Rate | 5 mL/min |
| Baseline Drift | 1 AU/h |
| Data Rate | 240 Hz |
| Heat Exchanger Internal Volume | 1.0 µL |
| Internal Volume | 1.0 µL |
| Maximum Individual Bottle Volume | 2 L |
| Maximum pressure | 1300 bar |
| Maximum Sampling Rate | 120 Hz |
| Maximum System Pressure | 1300 bar |
| Sample Capacity | 6144 samples |
| Solvent Cabinet Maximum Volume | 8 L |
| Valve Pressure Rating | 1300 bar |
| Maximum number of columns mfr | 8 columns |
| Number of columns per MCT mfr | 8 columns |
| Double mfr | 7116 A/B |
| Single mfr | 7114 A/B |
| Detector noise mfr | 1.5 μAU |
| Column thermostat cooling mfr | 20 ambient |
| Column thermostat heating mfr | 110 °C |
Description
Architecture and positioning
- Modular stack: Bio Flexible Pump (G7131A) or Bio High-Speed Pump (G7132A), Bio Multisampler (G7137A) or Hybrid Multisampler (G7137B), Multicolumn Thermostat G7116B with Bio Heat Exchanger, and Diode Array Detector FS (G7117A), DAD (G7117B), VWD (G7114B) or Fluorescence Detector (G7123B) with Bio flow cell.
- All modules are stackable; prescribed configurations (InfinityLab Flex Bench, Single Stack up to 4 modules, Benchtop, Two Stacks) set the flow path to minimize delay volume.
- Biocompatible modules must use bio/biocompatible parts only and must not be mixed with bio-inert parts.
Behavior under load and media
- System pressure ceiling is 1300 bar; the Multisampler injects at up to 1300 bar and holds up to 6144 samples.
- Analytical flow rate up to 5 mL/min; the Bio heat exchanger is offered at 1.0 µL (G7116-60091), 1.6 µL (G7116-60071) and 3.0 µL (G7116-60081) internal volumes for all flow rates — choose the lowest volume compatible with the flow to limit dispersion.
- Column thermostat: cooling to 20 °C below ambient, heating to 110 °C, up to eight columns (eight per MCT).
- Detector choice governs data rate: DAD FS G7117A up to 120 Hz; DAD G7117B and VWD G7114B up to 240 Hz. VWD noise <±1.5 µAU, baseline drift <1·10⁻⁴ AU/h.
- Valve pressure ratings differ: G5641A 2-position/10-port bio valve 1300 bar; G4235A bio-inert selector valve head 210 bar.
Installation, connection and control
- Module installation, specifications and maintenance are referenced to the individual module manuals; network integration is covered in each module's LAN Configuration chapter.
- LAN configuration is executed from the module with direct LAN connection to the controller software, usually the detector with the highest data rate.
- Rear DIP switches: all DOWN = default IP 192.168.254.11; switch 4 UP = DHCP; switch 5 UP = STORED address. MAC-based addressing only in DHCP configuration.
- Quick Start sequence: switch on detector, fill and place solvent bottles, purge (or prime after installation), select sampler tray format, add column and enter column information, select column position, set detector per method.
- Run the InfinityLab LC Performance Check after complete stack installation to confirm each module performs and is connected correctly.
Integration & Operating Notes
The system and Multisampler inject at up to 1300 bar, so every flow-path component must carry that rating. Note the G4235A bio-inert selector valve head is limited to 210 bar and must not be used in high-pressure methods; use the G5641A 2-position/10-port bio valve for 1300 bar service.
Dimethylformamide corrodes the PVDF leak sensor and must not be used as mobile phase. Inspect the sensor regularly for corrosion and keep metal parts off the leak panel and sensor, since a triggered sensor stops the running system immediately.
Install drain connectors on all modules of the stack except the lowest; the bottom module's drain outlet must be routed via waste tubing directly to a suitable waste container. Drain connectors can be retrofitted without powering off the stack and do not affect a running analysis.
All switches DOWN gives default IP 192.168.254.11; switch 4 UP selects DHCP; switch 5 UP selects the STORED address. MAC-address use is only possible in DHCP configuration. Configure the addressing mode before connecting the module to the controller network.
After complete installation of the module stack, run the InfinityLab LC Performance Check to confirm that each module performs and is connected correctly. This verifies the assembled system before routine use rather than relying on individual module self-tests.
Waste must drain directly into the waste containers with no restriction at bends or joints. The recommended container is 5043-1221 (6 L waste can with Stay Safe cap GL45, 4 ports); with a custom waste solution, tube ends must not immerse in the liquid to prevent back-siphon.
Selection & Integration Q&A
Choose the Bio configuration when the flow path must be biocompatible: use bio/biocompatible parts only and never mix them with bio-inert parts. It retains 1300 bar and 5 mL/min UHPLC performance with Bio flow cells and Bio heat exchangers, so bio-separations do not sacrifice pressure capability.
For fast separations select the DAD G7117B or VWD G7114B, both supporting up to 240 Hz; the DAD FS G7117A is limited to 120 Hz. Pair with the lowest-volume Bio heat exchanger compatible with the flow — 1.0 µL (G7116-60091), 1.6 µL (G7116-60071) or 3.0 µL (G7116-60081) — to minimize dispersion.
Plan for the recommended waste can 5043-1221 (6 L, Stay Safe cap GL45, 4 ports), the correct Bio capillaries per configuration, and the selected Bio heat exchanger and valve kit, which are listed as additional parts. Schedule the InfinityLab LC Performance Check after installation and before routine use.