Agilent Technologies 6890N GC
Agilent Technologies 6890N GC — Gas chromatography.
The Agilent 6890N GC Option 652 Oral Fluids Analyzer is a factory-configured two-dimensional heart-cutting GC/MS system built on the 6890N GC with Fast Oven, FID, split/splitless inlet, PCM, Dean's switch, JAS Cryo Trap and 5975N MSD. It targets derivatized THC in oral fluids using a 15 m DB-1ms primary and 15 m DB-17ms secondary column, with automated 3-minute post-run backflushing.
Full specifications
| Column temperature | 340 °C |
| Oven Ramp 1 Rate | 45.00 °C/min |
| Oven Ramp Rate 2 | 10.00 °C/min |
| Run Time | 9.72 min |
| Sample volume | 2.00 µL |
| Primary column film thickness mfr | 0.25 μm |
| Secondary column film thickness mfr | 0.25 μm |
| Oven initial time mfr | 0.50 min |
| Oven equilibration time mfr | 0.50 min |
| Oven post-run time mfr | 3.00 min |
| Syringe size mfr | 5.0 μL |
| Pre-injection solvent B washes mfr | 1 |
| Post-injection solvent A washes mfr | 10 |
| Post-injection solvent B washes mfr | 10 |
| Plunger speed mfr | Slow Slow |
| Post-run Column 1 pressure mfr | 1.0 psi |
| Post-run Column 2 pressure mfr | 65.0 psi |
| Solvent delay mfr | 5.00 min. |
| EM offset mfr | 200 |
| Maximum controlled inlet pressure program rate mfr | 99.99 psi/min |
| Required MS ion gauge pressure during post-run backflush mfr | 10 torr |
| THC checkout sample concentration mfr | 1 μg/mL |
| BSTFA stabilizer concentration mfr | 50 μg/mL |
| Oven post time mfr | 3.00 min |
| Post run time mfr | 3.00 min |
| Post run Column 1 pressure mfr | 1.0 psi |
| Post run Column 2 pressure mfr | 65.0 psi |
| Trap reheating offset after cut end mfr | 0.20 min |
| Example cut start time mfr | 5.56 min |
| Example cut end time mfr | 5.67 min |
| Brass sealing cap tightening mfr | 15 degrees |
| Minimum ion gauge pressure during post run backflushing mfr | 10 torr |
| Minimum void times through Column 1 for backflush mfr | 5 times |
| EMISSION mfr | 34.610 |
| ENERGY mfr | 69.922 |
| REPELLER mfr | 19.904 |
| IONFOCUS mfr | 84.463 |
| ENTRANCE_LE mfr | 22.000 |
| EMVOLTS mfr | 776.471 |
| AMUGAIN mfr | 1693.000 |
| AMUOFFSET mfr | 119.938 |
| ENTLENSOFFS mfr | 18.071 |
| MASSGAIN mfr | -694.000 |
| MASSOFFSET mfr | -36.000 |
| Column 1: (forward column) mfr | S/S inlet inlet |
| Column 2: (rearward column) mfr | Pass the column through the cryo trap, from bott |
| FID Restrictor tube: mfr | Dean’s switch plate bottom fitting fitting |
| Cryo trap air pressure mfr | 100 psig |
| Cryo trap air pressure minimum mfr | 40 psig |
| Primary column ID mfr | 0.25 mm |
| Secondary column ID mfr | 0.25 mm |
| FID restrictor tube ID mfr | 0.10 mm |
| FID restrictor tube OD mfr | 0.36 mm |
| MS ion gauge pressure during backflush mfr | 10 torr |
| Analysis portion end time mfr | 9.72 min |
| Solvent saving volume mfr | 1.0 μL |
| THC test sample concentration mfr | 1 μg/mL |
| Primary column inner diameter mfr | 0.25 mm |
| Primary column length mfr | 15 m |
| Secondary column inner diameter mfr | 0.25 mm |
| Secondary column length mfr | 15 m |
| FID restrictor tube inner diameter mfr | 0.10 mm |
| FID restrictor tube outer diameter mfr | 0.36 mm |
| FID restrictor tube length as supplied mfr | 1 m |
| FID restrictor tube installed length mfr | 31.2 cm |
| SIM ion mass 303.10 dwell mfr | 10 ms |
| SIM ion mass 303.20 dwell mfr | 10 ms |
| SIM ion mass 371.20 dwell mfr | 10 ms |
| SIM ion mass 374.20 dwell mfr | 10 ms |
| SIM ion mass 386.20 dwell mfr | 10 ms |
| SIM ion mass 389.20 dwell mfr | 10 ms |
| Resulting EM voltage mfr | 976.5 V |
| Dean's switch fitting nut tightening mfr | 15...20 degrees |
| Cryo trap coolant air supply pressure mfr | 100 psig |
| Viscosity delay mfr | 2 s |
| Cryo trap coolant air connection mfr | 1 inch |
| FILAMENT mfr | 1.000 |
| Cryo trap coolant air tube size mfr | 1 inch |
| Initial Oven Temperature | 150 °C |
| Ms Quadrupole Temperature | 150 °C |
| Source Temperature | 280 °C |
| Trap cooling rate mfr | 800 °C/min |
| Trap cooling start offset before cut start mfr | 2.00 min |
| Primary column (Column 1) part number mfr | 122 |
| Secondary column (Column 2) part number mfr | 122 |
| FID restrictor tube part number mfr | 160 |
| THC checkout sample part number mfr | 1540 |
| 5-μL syringe part number mfr | 5181 |
| 10-μL syringe part number mfr | 5181 |
| Brass sealing cap part number mfr | 2855 |
Description
Construction and positioning
- The Option 652 is a complete, factory-integrated analyzer: 6890N GC with Fast Oven option, FID, split/splitless capillary inlet, pneumatics control module (PCM), Dean's switch, JAS Cryo Trap with PCM, 7683B Automatic Liquid Sampler with tray, and 5975N MSD with MS ChemStation.
- The flow path is two-dimensional: Column 1 (DB-1ms, 15 m × 0.25 mm × 0.25 µm) separates the sample; the Dean's switch (Valve #1) either sends effluent to the FID (state 0, OFF) or diverts the heart cut through the cryo trap to Column 2 (DB-17ms, 15 m × 0.25 mm × 0.25 µm) and the MSD (state 1, ON).
- A FID restrictor tube (0.10 mm ID × 0.36 mm OD) balances the Dean's switch; its installed length must be 31.2 ±0.1 cm after trimming.
Behavior under load and in different modes
- Normal no-cut operation routes carrier to the FID only. The heart-cut mode traps the target in the cryo trap at 100 °C, then releases it to the MSD. Trap cooling from 300 °C to 100 °C at 800 °C/min takes 0.25 min and must start 2.00 min before the cut start time.
- The start-up eMethod OFA_THC_Base.emeth uses an initial oven of 150 °C (0.50 min hold), ramp 1 at 45.00 °C/min, ramp 2 at 10.00 °C/min, column temperature 340 °C, run time 9.72 min, and post-run at 300 °C for 3.00 min.
- Post-run backflushing reverses flow through Column 1: Column 1 pressure drops to 1.0 psi while Column 2 pressure rises to 65.0 psi for 3.00 min, providing at least 5 void times to flush heavy matrix components out the split vent.
- MS SIM parameters include solvent delay 5.00 min, EM offset 200 (resulting EM voltage 976.5 V), MS quad 150 °C (max 200 °C), MS source 280 °C (max 290 °C), and THC ions 303.10, 303.20, 371.20, 374.20, 386.20, 389.20 at 10 ms dwell.
Installation, connection and operation
- The cryo trap coolant air connection is a 1/4-inch plastic tube at the lower-right rear; supply must be oil- and moisture-free at 100 psig (689.5 kPa), with a minimum of 40 psig (276 kPa).
- The GC requires a stable AC power source capable of maintaining voltage at the full rated wattage; if the oven Ramp LED flashes 4 or more times during an analysis, actual oven temperature is no longer accurately tracking the controller.
- The ALS wash mode must be set manually through the GC keypad to Use A-A2, B-B2 and Solvent Saving to 1.0 µL; these are not included in the eMethod. Either a 5-µL syringe (P/N 5181-1273) or a 10-µL syringe (P/N 5181-1267) may be installed, and the installed volume must match the method configuration.
- Hydrogen must not be used as carrier gas due to tube breakage and explosion hazard inside the oven; helium is the carrier, hydrogen only for the FID flame.
Integration & Operating Notes
The JAS Cryo Trap requires coolant air at 100 psig (689.5 kPa) nominal, with an absolute minimum of 40 psig (276 kPa). Insufficient pressure prevents the trap from reaching its 100 °C trapping temperature in time, so the THC heart cut is not focused and the analysis fails. Air must be oil- and moisture-free.
Hydrogen must not be used as carrier gas with this product because of the possibility of tube breakage and explosion hazard inside the oven. Use helium as carrier; hydrogen is permitted only for the FID flame. This constraint applies to the whole analyzer configuration.
After trimming and end preparation, the FID restrictor tube must measure 31.2 ±0.1 cm installed. This length is critical for correct Dean's switch flow balance and cut performance; a tube outside tolerance will shift the split between FID and MSD paths.
Tighten the Dean's switch fitting nut by no more than 15 to 20 degrees after finger-tightening. Over-tightening can damage the fitting or ferrule. Use brass sealing cap P/N G2855-20590, finger-tight plus 15 degrees, to protect disconnected column and restrictor tubing ends.
If the oven Ramp LED flashes 4 or more times during an analysis, actual oven temperature is no longer accurately tracking the controller. This may indicate the AC power source capacity cannot keep up with oven heater demand; verify the supply can maintain voltage at the full rated wattage of the GC.
The Dean's switch valve is assigned as Valve #1; state 0 (OFF) routes flow to the FID, state 1 (ON) routes flow to the cryo trap and MSD. Cut times are entered as runtime valve events on Valve #1. Trap cooling must start 2.00 min before the cut start time so the trap is stable at 100 °C before the cut begins.
The ALS wash mode must be set to Use A-A2, B-B2 and Solvent Saving to 1.0 µL manually through the GC keypad; these parameters are not included in the eMethod. The installed syringe volume (5 µL P/N 5181-1273 or 10 µL P/N 5181-1267) must match the method configuration to avoid injection errors.
During post-run backflushing, Column 2 pressure is raised to 65.0 psi, dramatically increasing flow to the MSD. The MS vacuum system must maintain an ion gauge pressure of 10⁻³ torr or less; only turbopumping systems have sufficient capacity, so diffusion pump systems are eliminated.
Example THC cut times are 5.56 min start and 5.67 min end for a peak eluting at about 5–6 minutes, but cut times must be determined for each system and revisited after column aging, trimming, or replacement. The 3-minute post-run backflush provides at least 5 void times through Column 1 to remove high-boiling components.
Selection & Integration Q&A
Choose Option 652 when the method requires heart-cutting two-dimensional GC/MS of derivatized THC in oral fluids, with a factory-integrated Dean's switch, JAS Cryo Trap, FID and 5975N MSD. A standard 6890N without the Fast Oven, cryo trap and Dean's switch cannot perform the timed cut-and-trap sequence or the 3-minute backflush.
The MSD must be turbopumped: during post-run backflushing with Column 2 at 65.0 psi, the ion gauge must stay at 10⁻³ torr or less, which eliminates diffusion pump systems. The cryo trap needs oil- and moisture-free coolant air at 100 psig nominal (40 psig minimum) via a 1/4-inch plastic tube at the lower-right rear. Helium carrier is mandatory; hydrogen carrier is prohibited.
The GC needs a stable AC source able to hold voltage at the full rated wattage during the aggressive oven program (150 °C to 340 °C at 45.00 then 10.00 °C/min). If the oven Ramp LED flashes 4 or more times during an analysis, oven temperature is no longer tracking the controller, indicating insufficient supply capacity.
Plan for column and FID restrictor tube replacement (restrictor trimmed to 31.2 ±0.1 cm), a 5-µL (5181-1273) or 10-µL (5181-1267) syringe matching the method, brass sealing caps G2855-20590 for disconnected tubing, and THC checkout sample G1540-85010 (1 µg/mL derivatized THC in isoOctane with 50 µg/mL BSTFA, refrigerated up to one year). Re-determine cut times after any column change.