Agilent Technologies 6890N GC

Agilent Technologies 6890N GC — Gas chromatography.

The Agilent 6890N GC Option 652 Oral Fluids Analyzer is a factory-configured two-dimensional heart-cutting GC/MS system built on the 6890N GC with Fast Oven, FID, split/splitless inlet, PCM, Dean's switch, JAS Cryo Trap and 5975N MSD. It targets derivatized THC in oral fluids using a 15 m DB-1ms primary and 15 m DB-17ms secondary column, with automated 3-minute post-run backflushing.

Full specifications

Core Performance & Functional Specifications
Column temperature 340 °C
Oven Ramp 1 Rate 45.00 °C/min
Oven Ramp Rate 2 10.00 °C/min
Run Time 9.72 min
Sample volume 2.00 µL
Primary column film thickness mfr 0.25 μm
Secondary column film thickness mfr 0.25 μm
Oven initial time mfr 0.50 min
Oven equilibration time mfr 0.50 min
Oven post-run time mfr 3.00 min
Syringe size mfr 5.0 μL
Pre-injection solvent B washes mfr 1
Post-injection solvent A washes mfr 10
Post-injection solvent B washes mfr 10
Plunger speed mfr Slow Slow
Post-run Column 1 pressure mfr 1.0 psi
Post-run Column 2 pressure mfr 65.0 psi
Solvent delay mfr 5.00 min.
EM offset mfr 200
Maximum controlled inlet pressure program rate mfr 99.99 psi/min
Required MS ion gauge pressure during post-run backflush mfr 10 torr
THC checkout sample concentration mfr 1 μg/mL
BSTFA stabilizer concentration mfr 50 μg/mL
Oven post time mfr 3.00 min
Post run time mfr 3.00 min
Post run Column 1 pressure mfr 1.0 psi
Post run Column 2 pressure mfr 65.0 psi
Trap reheating offset after cut end mfr 0.20 min
Example cut start time mfr 5.56 min
Example cut end time mfr 5.67 min
Brass sealing cap tightening mfr 15 degrees
Minimum ion gauge pressure during post run backflushing mfr 10 torr
Minimum void times through Column 1 for backflush mfr 5 times
EMISSION mfr 34.610
ENERGY mfr 69.922
REPELLER mfr 19.904
IONFOCUS mfr 84.463
ENTRANCE_LE mfr 22.000
EMVOLTS mfr 776.471
AMUGAIN mfr 1693.000
AMUOFFSET mfr 119.938
ENTLENSOFFS mfr 18.071
MASSGAIN mfr -694.000
MASSOFFSET mfr -36.000
Column 1: (forward column) mfr S/S inlet inlet
Column 2: (rearward column) mfr Pass the column through the cryo trap, from bott
FID Restrictor tube: mfr Dean’s switch plate bottom fitting fitting
Cryo trap air pressure mfr 100 psig
Cryo trap air pressure minimum mfr 40 psig
Primary column ID mfr 0.25 mm
Secondary column ID mfr 0.25 mm
FID restrictor tube ID mfr 0.10 mm
FID restrictor tube OD mfr 0.36 mm
MS ion gauge pressure during backflush mfr 10 torr
Analysis portion end time mfr 9.72 min
Solvent saving volume mfr 1.0 μL
THC test sample concentration mfr 1 μg/mL
Dimensions, Weight & Mechanical Design
Primary column inner diameter mfr 0.25 mm
Primary column length mfr 15 m
Secondary column inner diameter mfr 0.25 mm
Secondary column length mfr 15 m
FID restrictor tube inner diameter mfr 0.10 mm
FID restrictor tube outer diameter mfr 0.36 mm
FID restrictor tube length as supplied mfr 1 m
FID restrictor tube installed length mfr 31.2 cm
SIM ion mass 303.10 dwell mfr 10 ms
SIM ion mass 303.20 dwell mfr 10 ms
SIM ion mass 371.20 dwell mfr 10 ms
SIM ion mass 374.20 dwell mfr 10 ms
SIM ion mass 386.20 dwell mfr 10 ms
SIM ion mass 389.20 dwell mfr 10 ms
Electrical, Power & Drive Specifications
Resulting EM voltage mfr 976.5 V
Interfaces, Connections & Control
Dean's switch fitting nut tightening mfr 15...20 degrees
Materials, Media & Process Compatibility
Cryo trap coolant air supply pressure mfr 100 psig
Viscosity delay mfr 2 s
Cryo trap coolant air connection mfr 1 inch
FILAMENT mfr 1.000
Cryo trap coolant air tube size mfr 1 inch
Operating Conditions, Environment & Protection
Initial Oven Temperature 150 °C
Ms Quadrupole Temperature 150 °C
Source Temperature 280 °C
Trap cooling rate mfr 800 °C/min
Trap cooling start offset before cut start mfr 2.00 min
Standards, Safety & General Data
Primary column (Column 1) part number mfr 122
Secondary column (Column 2) part number mfr 122
FID restrictor tube part number mfr 160
THC checkout sample part number mfr 1540
5-μL syringe part number mfr 5181
10-μL syringe part number mfr 5181
Brass sealing cap part number mfr 2855

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection and operation

Integration & Operating Notes

Process Limits
Cryo trap air supply below 40 psig prevents trapping

The JAS Cryo Trap requires coolant air at 100 psig (689.5 kPa) nominal, with an absolute minimum of 40 psig (276 kPa). Insufficient pressure prevents the trap from reaching its 100 °C trapping temperature in time, so the THC heart cut is not focused and the analysis fails. Air must be oil- and moisture-free.

Installation and Operation, page 20
Process Limits
Hydrogen carrier is prohibited in this oven

Hydrogen must not be used as carrier gas with this product because of the possibility of tube breakage and explosion hazard inside the oven. Use helium as carrier; hydrogen is permitted only for the FID flame. This constraint applies to the whole analyzer configuration.

Installation and Operation, page 18
Mounting
FID restrictor tube length must be 31.2 ±0.1 cm

After trimming and end preparation, the FID restrictor tube must measure 31.2 ±0.1 cm installed. This length is critical for correct Dean's switch flow balance and cut performance; a tube outside tolerance will shift the split between FID and MSD paths.

Installation and Operation, Table 1, page 60
Mounting
Dean's switch nut: 15–20° past finger-tight only

Tighten the Dean's switch fitting nut by no more than 15 to 20 degrees after finger-tightening. Over-tightening can damage the fitting or ferrule. Use brass sealing cap P/N G2855-20590, finger-tight plus 15 degrees, to protect disconnected column and restrictor tubing ends.

Installation and Operation, pages 63, 65
Electrical
Oven Ramp LED flashing 4+ times signals power shortfall

If the oven Ramp LED flashes 4 or more times during an analysis, actual oven temperature is no longer accurately tracking the controller. This may indicate the AC power source capacity cannot keep up with oven heater demand; verify the supply can maintain voltage at the full rated wattage of the GC.

Installation and Operation, pages 18, 45
Control & I/O
Valve #1 state 0 = FID, state 1 = trap/MSD

The Dean's switch valve is assigned as Valve #1; state 0 (OFF) routes flow to the FID, state 1 (ON) routes flow to the cryo trap and MSD. Cut times are entered as runtime valve events on Valve #1. Trap cooling must start 2.00 min before the cut start time so the trap is stable at 100 °C before the cut begins.

Installation and Operation, pages 36–39
Control & I/O
ALS wash and solvent-saving set manually on keypad

The ALS wash mode must be set to Use A-A2, B-B2 and Solvent Saving to 1.0 µL manually through the GC keypad; these parameters are not included in the eMethod. The installed syringe volume (5 µL P/N 5181-1273 or 10 µL P/N 5181-1267) must match the method configuration to avoid injection errors.

Installation and Operation, pages 21, 27
Process Limits
MS vacuum must hold ≤10⁻³ torr during backflush

During post-run backflushing, Column 2 pressure is raised to 65.0 psi, dramatically increasing flow to the MSD. The MS vacuum system must maintain an ion gauge pressure of 10⁻³ torr or less; only turbopumping systems have sufficient capacity, so diffusion pump systems are eliminated.

Installation and Operation, page 58
Maintenance
Re-determine cut times after column aging or trimming

Example THC cut times are 5.56 min start and 5.67 min end for a peak eluting at about 5–6 minutes, but cut times must be determined for each system and revisited after column aging, trimming, or replacement. The 3-minute post-run backflush provides at least 5 void times through Column 1 to remove high-boiling components.

Installation and Operation, pages 32–33, 52, 71

Selection & Integration Q&A

When should I choose the Option 652 Oral Fluids Analyzer rather than a standard 6890N GC configuration?

Choose Option 652 when the method requires heart-cutting two-dimensional GC/MS of derivatized THC in oral fluids, with a factory-integrated Dean's switch, JAS Cryo Trap, FID and 5975N MSD. A standard 6890N without the Fast Oven, cryo trap and Dean's switch cannot perform the timed cut-and-trap sequence or the 3-minute backflush.

What vacuum pump and gas supply does the adjacent MSD and cryo trap require?

The MSD must be turbopumped: during post-run backflushing with Column 2 at 65.0 psi, the ion gauge must stay at 10⁻³ torr or less, which eliminates diffusion pump systems. The cryo trap needs oil- and moisture-free coolant air at 100 psig nominal (40 psig minimum) via a 1/4-inch plastic tube at the lower-right rear. Helium carrier is mandatory; hydrogen carrier is prohibited.

What power and oven-tracking conditions must the installation site meet?

The GC needs a stable AC source able to hold voltage at the full rated wattage during the aggressive oven program (150 °C to 340 °C at 45.00 then 10.00 °C/min). If the oven Ramp LED flashes 4 or more times during an analysis, oven temperature is no longer tracking the controller, indicating insufficient supply capacity.

What maintenance and consumables are required beyond the base delivery?

Plan for column and FID restrictor tube replacement (restrictor trimmed to 31.2 ±0.1 cm), a 5-µL (5181-1273) or 10-µL (5181-1267) syringe matching the method, brass sealing caps G2855-20590 for disconnected tubing, and THC checkout sample G1540-85010 (1 µg/mL derivatized THC in isoOctane with 50 µg/mL BSTFA, refrigerated up to one year). Re-determine cut times after any column change.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Agilent 6890N Oral Fluids Analyzer
Manual ·78 pp.