No photo available yet
Agilent Technologies IDP 10 Dry Scroll Pump
Agilent Technologies IDP 10 Dry Scroll Pump — Liquid chromatography with Motor power 0.746 kW, Ultimate pressure 2 mbar.
The IDP-10 is a single-phase dry scroll vacuum pump with a peak pumping speed of 170 L/min and an ultimate pressure of 1.5 × 10⁻² Torr. Its 1 HP motor runs at a factory-set 1900 RPM (adjustable 1350–1900 RPM), and it handles up to 120 g/hr of water vapor with both gas ballast ports open. It is intended for oil-free evacuation of clean, non-corrosive gases in laboratory and analytical systems.
Full specifications
| Motor power | 1 HP |
| Ultimate pressure | 1.5 Torr |
| Coverslip type | No corrosive, explosive or particulate forming gases |
| Operating Speed | 1350...1900 RPM |
| Water Vapor Handling Per Pneurop 6602 | 120 g/hr |
| Peak pumping speed mfr | 170 L/min |
| Maximum recommended continuous inlet pressure mfr | 300 Torr |
| Maximum outlet pressure mfr | 115 psig |
| Exhaust connection mfr | 1 Thread |
| Gas ballast mfr | 1 |
| Motor thermal protection mfr | Automatic thermal protection protection |
| Factory setting operating speed mfr | 1900 RPM |
| Hour meter mfr | 1 hour |
| Conformance standards mfr | 61010 marked. |
| Installation mfr | 2 |
| Emissions and Immunity mfr | 61236 device |
| Safety mfr | 61010 Canada |
| Europe mfr | 10 m |
| Denmark mfr | 10 m |
| Switzerland mfr | 10 m |
| UK/Ireland mfr | 13 m |
| India mfr | 10 m |
| Israel mfr | 10 m |
| Japan mfr | 12 m |
| North America mfr | 15 m |
| IDP-10 Exchange Pump mfr | 3810 |
| IDP-10 Exchange with Inlet Valve mfr | 3810 |
| IDP-10 Full Pump Rebuild mfr | 3810 RP |
| IDP-10 Full Pump (with Inlet Valve) Rebuild mfr | 3810 RP |
| Description of the IDP-10 Dry Scroll Vacuum Pump mfr | 184 |
| Technical Specifications mfr | 185 |
| IDP-10 Dry Scroll Vacuum Pump Outline Drawing mfr | 187 |
| Electrical Connections mfr | 189 |
| Power Cord mfr | 189 |
| Grounding Instructions mfr | 189 |
| Extension Cords mfr | 190 |
| Mechanical Connections mfr | 191 |
| Isolation Valve mfr | 191 |
| Pump Inlet mfr | 191 |
| Pump Exhaust mfr | 191 |
| Gas Ballast mfr | 192 |
| Pump Operations mfr | 193 |
| Protection Features mfr | 194 |
| Troubleshooting mfr | 195 |
| Kits and Service Options mfr | 196 |
| Tip Seal Replacement mfr | 197 |
| Minimum clearance around equipment mfr | 60 cm |
| Maximum speed at 10 VDC mfr | 1900 RPM |
| Default speed (no input / below 1 V) mfr | 1800 tr/min |
| Pin 1 input impedance mfr | 10 Kohm |
| Recommended potentiometer mfr | 5 Kohm |
| Manual lifting threshold mfr | 20 kg |
| Weight | 24.7 kg |
| Mounting of the IDP-10 Scroll Vacuum Pump mfr | 193 |
| Packaged weight mfr | 28.1 kg |
| Inrush current | 191 ampere |
| Maximum inlet pressure at nominal voltage mfr | 1.0 atmosphere |
| Operating voltages mfr | 1 VAC |
| Fuse type mfr | 5 replaceable |
| Pin 7 reference voltage mfr | 10 Vcc |
| Inlet flange | 25 |
| Interface dimensions mfr | 297 mm |
| Cooling System | Air-cooled |
| Storage temperature | 20 °C |
| Altitude mfr | 2000 m |
| Ambient Relative Humidity mfr | 5 condensing |
| Noise Level (per ISO 11201) mfr | 53 dB |
| Vibration level at inlet (per ISO 10816-1) mfr | 1 mm/s |
Description
Construction and Series Positioning
The IDP-10 is a dry scroll pump available in two configurations: the standard model X3810-64000 and the X3810-64010 with an integral inlet valve. The inlet valve opens 30 seconds after start and closes immediately on power loss, providing system isolation. The pump weighs 24.7 kg and uses air cooling. It is not intended for corrosive, explosive, or particulate-forming gases.
Performance and Operating Behavior
Peak pumping speed is 170 L/min (10.2 m³/h), and ultimate pressure is 1.5 × 10⁻² Torr (2.0 × 10⁻² mbar). The maximum recommended continuous inlet pressure is 300 Torr (400 mbar), and the maximum inlet pressure at nominal voltage is 1.0 atm. The maximum outlet pressure is 115 psig. The operating speed is adjustable from 1350 to 1900 RPM; the factory setting is 1900 RPM, with an optional 1800 RPM setting. Under heavy load, overload control reduces motor speed to 1740 RPM after a time delay. Overtemperature protection shuts off the motor at approximately 85 °C while the fan continues running; the pump restarts automatically when cooled. Water vapor handling capacity is 120 g/hr with both gas ballast ports open. Noise level is 53 ± 2 dB(A) per ISO 11201.
Electrical, Control, and Installation
The pump operates on 1-phase 100–120 VAC 50/60 Hz (9 A typical) or 200–240 VAC 50/60 Hz (5 A typical). Inrush current can reach 50 A at 100–127 V, 60 Hz, or 49 A at 200 V, 50 Hz. A DSUB9 connector provides remote on/off (switch closure to ground or logic LO on pin 2) and remote speed control (0–10 VDC input; 1900 RPM at 10 V, 1350 RPM at 6.9 V; input impedance 10 kΩ). The slide switch must be in the 0 position for remote operation. Gas ballast port 1 is for moderate water vapor loads; port 2 is for high loads. For dry nitrogen gas ballast, use a 1/4 NPT connection at approximately 5 L/min. The pump requires at least 60 cm (24 in) of clear space on all sides for ventilation and access. Mounting holes accept four 1/4" or M6 through holes; rubber feet are secured by M3 hardware. The fuse (M5 × 20 mm, 10 A, Slo-Blo) is not user replaceable.
Integration & Operating Notes
The pump draws up to 50 A inrush at 100–127 V, 60 Hz (49 A at 200 V, 50 Hz). Select circuit breakers, wiring, and connectors rated for this starting current to avoid nuisance trips or voltage drop during startup.
For remote on/off via the DSUB9 connector, the front-panel slide switch must be set to 0. Use a shielded cable and ferrite bead to maintain EMC compliance. The remote speed input expects 0–10 VDC into 10 kΩ.
Operating continuously above 300 Torr (400 mbar) inlet pressure can overload the pump. The maximum inlet pressure at nominal voltage is 1.0 atm, but continuous operation should remain within the recommended limit to avoid overheating and shortened tip seal life.
The pump is air-cooled and requires at least 60 cm (24 in) of free space on all sides for ventilation and access. Insufficient clearance can cause overtemperature shutdown at approximately 85 °C and reduced service life.
The 10 A Slo-Blo fuse (M5 × 20 mm, Littelfuse 215010 or equivalent) is internal and must not be replaced by the user. If the pump fails to power on, contact Agilent service rather than attempting fuse replacement.
Exceeding 120 g/hr of water vapor can shorten tip seal life. Use gas ballast port 1 for moderate loads and port 2 for high loads; when both ports are open, install the sintered filter plug in port 2 to prevent debris ingress.
The base has four through holes for 1/4" or M6 bolts. Rubber feet are secured by M3 hardware and must be removed before rigid mounting. Ensure the mounting surface is level and can support the 24.7 kg weight.
Selection & Integration Q&A
Choose the X3810-64010 when the system requires automatic isolation on power loss: its inlet valve closes immediately on power loss and opens 30 seconds after start. The standard X3810-64000 lacks this valve and is suitable when external isolation is provided or not needed.
The IDP-10 operates on 1-phase 100–120 VAC 50/60 Hz (9 A typical) or 200–240 VAC 50/60 Hz (5 A typical). Circuit breakers and wiring must handle inrush currents up to 50 A at 100–127 V, 60 Hz. Use a grounded outlet and a power cord rated for the current; extension cords must be at least 16-gage and no longer than 25 ft (7.6 m).
Install the pump indoors at altitudes up to 2000 m, with ambient temperatures between 5 °C and a maximum not exceeding the pump's rating (refer to manual), and relative humidity 5–95% non-condensing. Leave at least 60 cm (24 in) clearance on all sides for air cooling and access. The pump is air-cooled and must not be enclosed without ventilation.
The base delivery does not include power cord, inlet and exhaust fittings, gas ballast filter plug, or remote-control cable. Tip seal replacement kits and service options are available separately. The fuse is not user replaceable; contact Agilent service for internal fuse replacement. Regular maintenance includes exterior cleaning and tip seal replacement per the manual.