Agilent Technologies TriScroll 300, Two-Stage, Dry, Inverter-Driven Pump
Agilent Technologies TriScroll 300, Two-Stage, Dry, Inverter-Driven Pump — Vacuum technologies with Motor power 0.5 kW, Noise level 50 dB, Temperature 20 °C.
The TriScroll 300 Inverter is a two-stage dry scroll pump driven by an integrated single-phase-to-three-phase inverter, reaching 1.3 mbar ultimate pressure at a peak pumping speed of 250 l/min (8.9 cfm). Its 35–65 Hz speed range, factory-set at 62 Hz, distinguishes it from fixed-speed siblings and allows pressure-linked speed control. It serves as a clean, oil-free roughing stage for mass spectrometry and other contamination-sensitive vacuum systems.
Agilent Technologies TriScroll Series — Model Comparison & Selection
| Model | Motor power, kW | Storage temperature, °C | Noise level, dB | Ultimate pressure, mbar | Operating speed, rpm | Shipping weight |
|---|---|---|---|---|---|---|
| TriScroll 300, Two-Stage, Dry, Inverter-Driven Pump Current | 0.5 | -20...60 | 50 | 1.3 | 334.2 | 33.6 |
| TriScroll 600, Two-Stage, Dry, Inverter-Driven Pump | 1 | -6.67 | 50 | 1.3 | 1450 | 41 |
Full specifications
| Motor power | 0.67 HP |
| Noise level | 50 dB |
| Temperature | 20 °C |
| Ultimate pressure | 1.3 mbar |
| Coverslip type | Clean air. No toxic, corrosive, explosive or particulate forming gases |
| Inlet Pressure | 1.0 atmosphere |
| Operating Speed | 35 Hz |
| Rotational speed | 35 Hz |
| Peak pumping speed mfr | 250 |
| Exhaust connection mfr | 1 Thread |
| Gas ballast connection mfr | 16 |
| Power requirements mfr | 200 amps |
| Maximum continuous inlet pressure mfr | 100 Torr |
| Maximum outlet pressure mfr | 1.1 atmosphere |
| Gas ballast mfr | 1 Thread |
| Motor thermal protection mfr | Automatic Automatic |
| Installation and storage mfr | 2 storage |
| Peak pumping speed (TS300inv) mfr | 250 hr |
| Peak pumping speed (TS600inv) mfr | 500 h |
| Ultimate pressure (TS600inv) mfr | 9.3 mbar |
| Europe mfr | 10 m |
| Denmark mfr | 10 m |
| Switzerland mfr | 10 m |
| UK/Ireland mfr | 13 m |
| India mfr | 10 m |
| Israel mfr | 10 m |
| North America mfr | 10 m |
| Ultimate pressure (Torr) mfr | 1.0 Torr |
| Major Maintenance Kit mfr | 300 pumps. |
| Maintenance Tool Kit mfr | 300 pumps. |
| Tip Seal Tool Kit mfr | 300 pumps. |
| Replacement Tip Seal Set mfr | 300 pumps. |
| Tip Seal Replacement Manual mfr | 300 models |
| Pump Module Replacement Manual mfr | 300 models |
| Major Maintenance Manual mfr | 300 models |
| Advance Exchange TriScroll 300 Pump Module Only mfr | 0300 SC |
| Service/Rebuild TriScroll 300 Inverter Pump mfr | 0300 KMA |
| Fax: 1 781 860 9252 mfr | please visit our website for individual office i |
| Gas ballast flow rate mfr | 5 lpm |
| Vibration reduction mfr | 86 |
| HEPA filter efficiency mfr | 99.97 |
| HEPA particle size mfr | 0.3 microns |
| Shipping Weight | 33.6 kg |
| Weight | 26 kg |
| Weight – pump only mfr | 26 kg |
| Weight (TS600inv) mfr | 31 kg |
| Bearing purge pressure mfr | 2 less |
| Bearing purge pressure limit mfr | 5 psig |
| Circuit Breaker Capacity | 10 A |
| Frequency | 65 Hz |
| Input voltage | 200 V |
| Motor Line Current | 5 A |
| Operating voltage mfr | 50 |
| Main fuse mfr | 10 V |
| Normal Operation maximum power mfr | 750 W |
| Inlet connection | 25 |
| Gas ballast port mfr | 1 NPT |
| Interface dimensions mfr | 1 |
| Cooling System | Air-cooled |
| Noise level | 50 dB |
| Storage temperature | -20...60 °C |
| Pollution degree mfr | 2 |
| Altitude mfr | 2000 m |
| Protection level mfr | 20 |
| Noise level (per ISO 11201) mfr | 55 dB |
| Vibration level at inlet (per ISO 10816-1) mfr | Variable with frequency frequency |
| Noise level per ISO 11201, variable w/frequency mfr | 55 dBA |
Description
Construction and series positioning
- Two-stage dry scroll element on a single shaft, direct-coupled to a three-phase induction motor fed by an on-board inverter; no oil in the swept volume.
- Within the Triscroll series, this model sits between the 15 m3/hr and 30 m3/hr entries and below the TriScroll 600 (1.0 kW, same 1.3 mbar ultimate). Its distinguishing feature is the inverter, which replaces fixed-speed line operation with a 35–65 Hz adjustable range.
- Pump-only weight 26 kg; shipping weight 33.6 kg. Air-cooled, indoor installation only.
Behavior under load and in different modes
- Ultimate pressure 1.0 × 10⁻² Torr (1.3 × 10⁻² mbar); the manual notes up to one hour with the inlet sealed may be needed to reach it.
- Maximum continuous inlet pressure 100 Torr; maximum outlet pressure 1.1 atmosphere. Do not run continuously above the inlet limit.
- Speed is set by frequency: 35 Hz minimum to 65 Hz maximum, factory 62 Hz (1800 RPM). If the pump needs more than 750 W to hold the set frequency, the inverter automatically reduces output frequency (autotuning) — lower speed under high load is normal protection, not a fault.
- Noise scales with frequency, 55–68 dB(A) per ISO 11201; the exhaust silencer kit and lower-speed operation are the documented mitigations.
Installation, connection and control
- Supply: 200–240 VAC single-phase, 50–60 Hz ±10% (100–120 VAC only with inverter P/N S4897001). Motor line current 5 A at 200 VAC, 4.2 A at 240 VAC; main fuse Type H 10 A, 250 V; circuit breaker capacity 10 A.
- The P1 interlock connector must be fitted with pins 4–5 and 8–9 linked, or the pump will not start.
- Remote speed setting via 0–10 V analog input: >9 V = full speed (62 Hz), <1 V = minimum (35 Hz). J2 serial port (RS232 or RS485, max 9600 baud) exposes operating variables through T-Plus software.
- Automatic gas ballast with sintered filter handles condensate; a bearing purge port accepts dry nitrogen when pumping high water vapor.
Integration & Operating Notes
The supplied mating connector must be plugged into P1 with pins 4–5 and pins 8–9 connected together. Without these links the inverter stays disabled, so wiring the interlock into a system E-stop or door circuit must preserve continuity during normal run.
The standard inverter accepts 200–240 VAC single-phase, 50–60 Hz ±10%; 100–120 VAC operation requires the separate inverter P/N S4897001. Size the branch circuit for 5 A at 200 VAC and fit a 10 A circuit breaker; extension cords must be at least 16-gauge and no longer than 25 ft (7.6 m) to avoid voltage drop and overheating.
Continuous operation above 100 Torr inlet risks overheating and damage; maximum outlet pressure is 1.1 atmosphere. The inverter's 750 W power ceiling also means the pump will self-reduce frequency under heavy gas load rather than exceed its power budget, so plan roughing cycles around the 100 Torr ceiling.
When pumping high water vapor, feed dry nitrogen to the bearing purge port at roughly 5 lpm with pressure ≤2 psig and never above 5 psig. The purge kit PTSPURGEKIT is a separate accessory; the standard gas ballast uses a sintered filter, and blocking the gas ballast port is not recommended.
Ambient operating range is +5 °C to +40 °C; storage −20 °C to +60 °C. The pump is air-cooled, so fan ducts must not be blocked and the unit must sit on a stable surface indoors. For vibration-sensitive instruments, the vibration isolation kit replaces the standard feet and typically cuts transmitted vibration by 86% on this model.
After replacing tip seals, the pump may need up to 24 hours of running before it reaches full rotational speed and stable base pressure — do not judge the service as failed before that window. Bearing grease replenishment and tip seal replacement are the scheduled interventions, triggered when base pressure rises.
Selection & Integration Q&A
Pick the 300 Inverter when 250 l/min (8.9 cfm) at 1.3 mbar ultimate is enough and you want variable speed (35–65 Hz) for noise or pressure-linked control. The TriScroll 600 gives 1.0 kW for higher throughput at the same ultimate; fixed-speed variants suit simple, always-on roughing without inverter control.
A 200–240 VAC single-phase supply (50–60 Hz ±10%) on a 10 A breaker, the P1 mating connector with pins 4–5 and 8–9 linked, and optionally a 0–10 V analog signal for remote speed (>9 V full, <1 V minimum). J2 gives RS232/RS485 monitoring up to 9600 baud via T-Plus software.
A fast-acting normally closed isolation valve must open at least 350 ms after pump startup and close within 250 ms of shutoff to prevent debris backstreaming; VPI kits are available. For particle-sensitive processes, the HEPA inlet filter is rated 99.97% for particles >0.3 microns.
Base delivery excludes the exhaust silencer kit, vibration isolation kit, HEPA inlet filter, bearing purge kit PTSPURGEKIT, isolation valve kits, CT-100 speed control kit and the 100–120 VAC inverter. Scheduled service is bearing grease replenishment and tip seal replacement when base pressure rises, followed by up to 24 h run-in.