Agilent Technologies VHS-250 Diffusion Pump
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies VHS-250 Diffusion Pump

Agilent Technologies VHS-250 Diffusion Pump — Vacuum technologies with Motor power 1.45 kW, Pumping speed 4320 m³/h.

The VHS-250 is a 4-stage, self-aligning stainless-steel jet diffusion pump with a standard cold cap, rated for 3700 L/s air and 4600 L/s helium. It is the largest of the VHS-4/VHS-6/VHS-250 series, requiring a 17 cfm backing pump and 0.25 gpm cooling water at 15 psi. It serves high-gas-load vacuum processes where a 10-minute warm-up and 0.65 Torr maximum forepressure are acceptable.

Motor power
1.45 kW
Pumping speed
4320 m³/h

Agilent Technologies Small Diffusion Pumps Series — Model Comparison & Selection

Model Motor power, kWPumping speed, m³/hCooldown time, minutesCooling water pressure, psiDiffusion pump pumping speed, L/sWarm up time, minutes
VHS-250 Diffusion Pump Current 1.45432010155010
VHS-4 Diffusion Pump 1.45432010155010
VHS-6 Diffusion Pump 1.45432010155010

Full specifications

Core Performance & Functional Specifications
Motor power 1450 Watts
Pumping speed 1200 l/s
Cooldown Time 10 minutes
Cooldown Time (using Quick Cool Coil) 10 minutes
Diffusion Pump Pumping Speed 50 L/s
Extended Cold Cap 120 V
Heater Power 1450 watts
Warm-up Time 10 minutes
Vacuum gauge measurement range mfr 6 pressure
Dry scroll pump typical size mfr 3...30 h
Turbomolecular pump pumping speed mfr 70...3 L/s
Optimum Operating Range mfr 9 mbars
Maximum Forepressure mfr 0.65 Torr
Backstreaming Rate at Pump Inlet mfr 5 minute
Backing Pump Size mfr -4 throughput
Recommended Jet Assembly mfr 4 steel
Maximum Throughput mfr 1.2 l/s
Optimum Operating Range (Torr) mfr 1
Maximum Throughput (Torr-l/s) mfr 1.2 Torr
Poor system pressure mfr System dirty. dirty.
Poor ultimate pressure mfr Water in Quick Cool Coil. Coil.
Slow pumpdown (prolonged cycle after checking gas load conditions) mfr Malfunctioning pump assembly Improperly located
Electrical Requirements mfr 120 phase
Pressure drop across coils mfr -4 psi
Foreline Baffle mfr Stacked half moon moon
Water Connections mfr 1 FPT
Installation mfr 2
Heater circuit resistance (ohms) — 120 V mfr 9.4
Heater circuit resistance (ohms) — 208 V mfr 28.3
Heater circuit resistance (ohms) — 240 V mfr 37.7
Clearance (for heater removal) mfr 6 minimum
Materials — Body, Flanges, Foreline, Baffle mfr stainless steel steel
Materials — Jet Assembly mfr stainless steel steel
Materials — Cold Cap mfr nickel-plated copper copper
Rough pump pressure before startup mfr 0.5 Torr
Heater clamp bolt torque mfr 250 pounds
Dimensions, Weight & Mechanical Design
Shipping Weight 55 pounds
Interfaces, Connections & Control
Turbomolecular pump controller protection class mfr 54
Inlet Flange, Nominal Size mfr 250 F
Foreline Flange, Nominal Size mfr 1 ASA
Materials, Media & Process Compatibility
Fluid Charge mfr -4 cc
VHS-4 oil charge mfr 300 cc
VHS-6 and VHS-250 oil charge mfr 500 cc
VHS-4 fluid charge mfr 300 cc
VHS-6 and VHS-250 fluid charge mfr 500 cc
Operating Conditions, Environment & Protection
Cooling Water Pressure 15 psi
Cooling Water Requirements mfr 60 °F
Cooling Water Flow mfr 0.25 gpm
Thermal Switches mfr 300 °F
Environmental mfr 113 °F
Altitude mfr 6562
Materials — Body Cooling Coils mfr copper copper
Materials — Quick Cooling Coil mfr stainless steel steel
VHS-4 cooling water flow mfr 0.15 gpm
VHS-6 and VHS-250 cooling water flow mfr 25 gpm
Standards, Safety & General Data
Standard cold cap mfr 240

Description

Construction and series positioning

Operating behavior and limits

Installation and operation specifics

Integration & Operating Notes

Process Limits
Keep foreline below 0.65 Torr to sustain pumping

The maximum forepressure (no load) is 0.65 Torr (0.87 mbar). If the backing pump cannot hold the foreline below this value, the jet assembly cannot maintain its compression ratio and the pump will stall or backstream. Size the backing line and pump accordingly.

User Manual, Table 1-1, page 19
Cooling & Media
Provide 0.25 gpm cooling water at 15 psi and 60–80 °F

Cooling water flow for the VHS-250 is 0.25 gpm at 15 psi, with a maximum inlet temperature of 60/80 °F (15–26 °C). Insufficient flow or excessive inlet temperature will overheat the pump and may trip the 300 °F thermal switches, forcing a manual reset after cooldown.

User Manual, Table 1-1, page 20
Electrical
Safety circuits must hold the contactor closed

The pump will not start if safety circuits and/or protective devices prevent the contactor from staying closed. Verify that flow switches, interlocks, and thermostat contacts are closed and that all utilities are present before troubleshooting the heater circuit.

User Manual, Table 4-1, page 51
Maintenance
Manual-reset thermal switches require cooldown before restart

Thermal switches are manual reset at 300 °F (149 °C). After a trip, allow the pump to cool, correct the cause (usually cooling water flow or temperature), and manually reset the switch; automatic restart is not provided.

User Manual, Table 1-1, page 20
Process Limits
Avoid prolonged operation above 10⁻³ Torr

Prolonged operation at high throughput at pressure above 10⁻³ Torr (1.33×10⁻³ mbar) can cause high chamber contamination. Use the pump within its optimum operating range and avoid sustained high-pressure operation to protect process cleanliness.

User Manual, Table 4-1, page 51

Selection & Integration Q&A

When should I choose the VHS-250 instead of the VHS-4 or VHS-6?

Choose the VHS-250 for the highest gas load in the VHS series: 3700 L/s air and 4600 L/s helium with standard cold cap, versus 1200 L/s for the VHS-4. It requires a 17 cfm backing pump and 0.25 gpm cooling water; the VHS-4 needs only 10 cfm. Select the VHS-4 for smaller chambers and lower throughput.

What backing pump and foreline conditions does the VHS-250 require?

Use a backing pump of at least 17 cfm for maximum throughput and ensure the foreline pressure stays below the 0.65 Torr maximum forepressure (no load). The backing line must be sized to handle the gas load without exceeding this limit, otherwise pumping action is lost.

What maintenance and service items are required, and what is not included in base delivery?

The manual covers cold cap and jet assembly removal/installation, heater replacement, and pump fluid charge; these are service actions, not included accessories. Leak checks should target inlet and foreline connections, drain and fill plugs, compression fittings, and threaded connections. Order replacement parts per the manual's parts list; cables and controllers are not listed as included.