BINDER FD-S 56

BINDER FD-S 56 — Drying ovens with Temperature 10…250 °C.

The FD-S 56 is a forced-convection drying and heating oven with a 55 L stainless steel chamber, rated +10 to 250 °C and 1.1 kW on 230 V single-phase. Its 4 h⁻¹ air-change rate and 15 min heat-up to 150 °C distinguish it from natural-convection ED-S siblings. It serves solvent-free drying, thermal treatment and aging where uniform forced airflow matters.

Temperature
10…250 °C

BINDER FD-S Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior depth, mm
FD-S 56 Current 10–25045771562.5310
FD-S 115 10–25026162073.5350

Full specifications

Core Performance & Functional Specifications
Temperature 10...250 °C
Air change rate 4 x/h
Heating time to 150°C 15 min
Interior volume 55 L
Load per rack 15 kg (FDS056-230V)
Nominal power 1.1 kW
Permitted load 45 kg (FDS056-230V)
Temperature fluctuation 0.4 K
Temperature uniformity 2.6 K
Rear wall clearance 160 mm (FDS056-230V)
Side wall clearance 100 mm (FDS056-230V)
Article Number mfr 9090
Recovery time after door was opened for 30 s at 150°C mfr 5 min
Dimensions, Weight & Mechanical Design
Depth 577 mm (FDS056-230V)
Height 625 mm (FDS056-230V)
Interior depth 310 mm (FDS056-230V)
Interior height 440 mm (FDS056-230V)
Interior width 400 mm (FDS056-230V)
Width 614 mm (FDS056-230V)
Net weight of the unit (empty) mfr 35 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior

Installation and operation

Integration & Operating Notes

Electrical
Supply must stay within ±10 % of 230 V nominal

The chamber is rated 230 V 1~ 50 Hz with a 6.3 A fuse. Verify mains quality before installation; out-of-tolerance voltage can affect controller operation and heating performance.

Type plate, p. 8; Instruction Manual, chap. 4.3
Mounting
Respect 160 mm rear and 100 mm side clearances

Insufficient clearance or installation in unventilated recesses prevents heat dispersal and can cause overheating damage. Plan the bench layout around these minimum distances.

Instruction Manual, chap. 3.4, p. 17
Process Limits
No explosive or combustible charges permitted

The chamber has no explosion protection. Operating temperature must remain below the flash point or sublimation point of the charging material; verify material properties before loading.

Instruction Manual, chap. 1.9, p. 10-11
Process Limits
Avoid corrosive materials that attack stainless steel

Interior is 1.4016 (AISI 430). Acids and halides can corrode it; corrosion damage is excluded from liability. Use only harmless charging materials.

Instruction Manual, chap. 2, p. 14
Maintenance
Annual electrical safety test and maintenance required

Plan at least yearly maintenance and repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. Disconnect power before service; do not remove rear panel; replace door gasket only when cold.

Instruction Manual, chap. 12.1-12.2, p. 29-30
Control & I/O
Display resolution is 1 °C, not 0.1 °C

FD-S ovens show temperature in 1 °C steps; sub-degree readout is available only on BD-S incubators. Set and read set-points accordingly.

Instruction Manual, chap. 2, p. 14

Selection & Integration Q&A

When should I choose the FD-S 56 over the ED-S 56 or the larger FD-S 115?

Pick FD-S 56 for forced convection and 55 L capacity with 4 h⁻¹ air change. Choose ED-S 56 if natural convection suffices; choose FD-S 115 for larger loads. FD-S 56 suits uniform drying where airflow is needed.

What supply and fusing does the 230 V FD-S 56 require?

230 V 1~ 50 Hz, 1.1 kW, 6.3 A chamber fuse, CEE 7/7 plug, ±10 % voltage tolerance. Provide a grounded single-phase circuit with 6.3 A protection per chamber.

What maintenance and accessories are required beyond base delivery?

Annual maintenance and electrical safety test per EN 50678/VDE 0701 and EN 50699/VDE 0702. Door gasket is a wear part (replace cold). Racks and exhaust connection are optional accessories; base delivery excludes them.

Dimensional Drawings & Technical Documentation

Dimensional drawings

FD-S 56 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation