BINDER KBF 130
BINDER KBF 130 — Climate chambers with Temperature 0…70 °C.
The BINDER KBF 130 is a 127 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, using a steam humidifier and R600A refrigeration circuit. As the smallest standard KBF variant (Art. 9020-0478, 230 V), it is distinguished from larger siblings by its compact 130-size cabinet and tilt-protection kit, while sharing the MB2 program controller and DIN 12880 class 2/3.3 safety architecture. It is intended for reproducible conditioning and stability testing of samples at defined temperature and humidity in laboratory and quality-control environments.
BINDER KBF Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 127 L (KBF130-230V / KBF130UL-120V) |
| Load per rack | 30 kg (KBF130-230V / KBF130UL-120V) |
| Maximum heat compensation | 270 W |
| Nominal power | 1.6 kW (KBF130-230V / KBF130UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 3 min |
| Power frequency | 50/60 Hz (KBF130-230V / KBF130UL-120V) |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A (KBF130-230V / KBF130UL-120V) |
| Humidity fluctuation | 1.5 % RH |
| Humidity range | 10...80 % RH |
Description
Construction and series positioning
- The KBF 130 is the 130-size member of the KBF constant climate chamber series; larger siblings (KBF 260, 470, 1060, 1600) share the same 0...70 °C temperature range but differ in interior volume and, from size 260 upward, require a wall spacer.
- This model uses the triangular instrument panel layout specific to sizes 130 and 260, with the MB2 chamber controller providing program, fixed-value, timer and week-program operating modes.
- The 130-size cabinet is equipped with a flexible tilt-protection kit (Section 4.2); a condensate collection pan applies to units up to serial number 20250000010225 (Section 4.4).
- Humidity control is provided by a steam humidifier; the humidifying system and wastewater/condensate handling changed at serial number 20250000010226, so the serial number determines the correct spare-parts and connection scheme.
Behavior under load and in different media
- Temperature fluctuation is 0.1 K and temperature uniformity 0.2 K, with humidity fluctuation of 1.5 % RH; the humidity sensor itself has a tolerated error of ±2 % RH, which must be considered when interpreting readings.
- Recovery after a 30 s door opening is 2 min at 25 °C / 60 % RH and 3 min at 40 °C / 75 % RH, indicating the control dynamics available for repeated-access duty.
- Maximum heat compensation is 270 W, and the load per rack is limited to 30 kg; the chamber is not intended for high thermal-load samples beyond this compensation capacity.
- The humidity sensor must not be exposed to substances above the tolerated concentrations listed in Section 1.14 (e.g. NH3 5500 ppm, HCl 300 ppm, H2S 350 ppm permanent load); operation with aggressive media outside these limits will degrade the sensor.
Installation, connection and control
- Electrical supply is single-phase 220...240 V, 50/60 Hz, nominal power 1.6 kW, unit fuse 16.0 A; the 230 V variant uses a CEE 7/7 plug, while the UL variant (9020-0494) uses 120 V with NEMA 5-20.
- The rear connection panel carries the water connections for humidification; maximum operating pressure is 11 bar, and the freshwater/wastewater arrangement depends on the serial-number generation.
- The MB2 controller supports Ethernet and USB data transfer, chart-recorder display, alarm and notification functions, and user management with password-protected authorization levels.
- Temperature safety is provided by an over-temperature protective device class 1 plus an over-temperature safety controller class 2/3.3 (selectable as temperature limiter or temperature protection), with undertemperature protection class 3.2.
- The chamber is rated IP 20 (EN 60529), so it must not be sprayed with water; cleaning and decontamination procedures are described in the manual.
Integration & Operating Notes
The KBF 130-230V draws 1.6 kW nominal on a 1~ 220...240 V, 50/60 Hz supply and requires a 16.0 A unit fuse. Connecting to a 120 V circuit or a supply outside this band will prevent correct operation; the UL variant (9020-0494) is the separate 120 V model with NEMA 5-20 plug.
Although the chamber temperature range is 0...70 °C, humidity control is specified only for 10...70 °C. Operating the humidifier below 10 °C is outside the validated humidity range and may produce condensation or control instability; plan humidity tests within the 10...70 °C window.
Each rack is limited to 30 kg load, and the chamber can compensate a maximum of 270 W of heat input from samples. Exceeding either limit compromises temperature uniformity (0.2 K) and may prevent the chamber from holding set-point, especially at low temperatures where refrigeration capacity is already engaged.
Regular maintenance must be performed at least once a year, including the steam humidifier; condensers should be cleaned at least twice a year (more often in dusty ambient air). Refrigeration work requires personnel certified per EN 13313:2010. After 8760 operating hours or two years a maintenance message appears every two weeks until reset by BINDER Service.
The chamber's maximum operating pressure is 11 bar. Freshwater supply and wastewater connections must be rated accordingly and the pressure regulated if the site water main exceeds this value. The humidifier's maintenance interval also depends on the available water quality and steam output, so water hardness should be checked at commissioning.
Selection & Integration Q&A
Pick the KBF 130 for 127 L interior volume and loads within 30 kg per rack and 270 W heat compensation. Larger siblings (260, 470, 1060, 1600) share the 0...70 °C range but add volume and, from size 260, require a wall spacer; the 130 uses the compact triangular panel and tilt-protection kit.
The 230 V variant needs a 1~ 220...240 V, 50/60 Hz supply with 16 A fuse and CEE 7/7 plug (1.6 kW nominal). Water connections must withstand 11 bar; the humidifier needs freshwater and wastewater routing, with the connection scheme depending on whether the serial number is up to or above 20250000010225.
The 130-size unit requires the flexible tilt-protection kit and is rated IP 20, so it must not be sprayed with water. Unlike sizes 260 and larger, no wall spacer is specified for the 130. Ambient dust increases condenser cleaning frequency beyond the standard twice-yearly interval.
Plan annual maintenance including the steam humidifier, twice-yearly condenser cleaning, and electrical safety testing per EN 50678/EN 50699. The maintenance message appears at 8760 h or two years and requires BINDER Service reset. Optional accessories include APT-COM 4, analog outputs, zero-voltage alarm contacts, Pt 100 object sensor, and external freshwater/wastewater cans.