BINDER KBF 260
BINDER ·Made in Germany · In Active Production

BINDER KBF 260

BINDER KBF 260 — Climate chambers with Temperature 0…70 °C.

The BINDER KBF 260 is a 262 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, housed in a 260-size cabinet with a triangular instrument panel. It is the mid-series 230 V single-phase model (220...240 V, 1.4 kW, 16 A fuse) distinguished from smaller 130 and larger 470/1060/1600 siblings by its 40 kg-per-rack load capacity and 410 W max heat compensation. It is used for stable, reproducible temperature and humidity conditioning of samples in laboratory and industrial test processes.

Temperature
0…70 °C

BINDER KBF Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF 260 Current 0–70262404101.40.3
KBF 1060 0–701070406401.60.3
KBF 130 0–70127302701.60.2
KBF 1600 0–701610406401.70.4
KBF 470 0–70472405301.40.2
KBF 720 0–70727406401.60.3

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 262 L (KBF260-230V / KBF260UL-120V)
Load per rack 40 kg (KBF260-230V / KBF260UL-120V)
Maximum heat compensation 410 W
Nominal power 1.4 kW (KBF260-230V / KBF260UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 2 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 3 min
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (KBF260-230V / KBF260UL-120V)
Rated voltage 220...240 V
Unit fuse 16.0 A (KBF260-230V / KBF260UL-120V)
Operating Conditions, Environment & Protection
Humidity fluctuation 1.5 % RH
Humidity range 10...80 % RH

Description

Construction and series positioning

Behavior under load and in humidity mode

Installation, connection and control

Integration & Operating Notes

Electrical
Match site voltage and frequency to the ordered mains variant

The 230 V variant requires 220...240 V single-phase 50/60 Hz with a 16 A unit fuse and CEE 7/7 plug; the UL variant requires 120 V with NEMA 5-20. Connecting a chamber to a supply outside its type-plate rating risks damage and invalidates safe operation, so verify the type plate before installation.

Data Sheet Model KBF 260 en.pdf; Operating Manual KBF/KBF PRO (E7), Section 1.6 Type plate
Mounting
Maintain the specified wall distance using the wall spacer

From chamber size 260 onward a spacer for wall distance is required; without it, heat rejection from the rear of the chamber is impaired. The flexible tilt protection kit is not used on this size — it applies only to sizes 130 and 470.

Operating Manual KBF/KBF PRO (E7), Sections 4.1 and 4.2
Process Limits
Keep sample heat load within the 410 W heat compensation limit

The maximum heat compensation of 410 W defines how much heat the chamber can remove from operating samples. Exceeding it prevents the chamber from holding the set temperature, so calculate the total dissipated power of the load before designing a test.

Factory Specification Sheet
Process Limits
Water connections must be rated for at least 11 bar

The chamber's maximum operating pressure is 11 bar. Freshwater and wastewater fittings, tubing and valves connected to the rear panel must be rated for at least this pressure to avoid leakage or rupture when connected to a pressurized water main.

Operating Manual KBF/KBF PRO (E7), Section 1.6 Type plate
Maintenance
Plan annual maintenance and twice-yearly condenser cleaning

Regular maintenance is required at least once a year, and the steam humidifier must also be serviced at least annually. Condensers should be cleaned at least twice a year by a qualified technician, and more often in dusty ambient air. Maintenance by non-authorized personnel voids the warranty.

Operating Manual KBF/KBF PRO (E7), Section 23.2
Maintenance
Refrigeration work requires EN 13313:2010-certified personnel

The refrigeration system contains flammable R600A hydrocarbon refrigerant. Any maintenance on it must be performed by qualified personnel trained in accordance with EN 13313:2010 (e.g. refrigeration technician certified per Regulation (EC) No. 303/2008).

Operating Manual KBF/KBF PRO (E7), Sections 1.6 and 23.2
Maintenance
Replace the door gasket only when the chamber is cold

Replacing the door gasket while the chamber is warm may damage the new gasket. Allow the chamber to cool to ambient before attempting replacement, and include electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 in annual service.

Operating Manual KBF/KBF PRO (E7), Sections 23.1 and 23.2

Selection & Integration Q&A

When should I choose the KBF 260 rather than the KBF 130 or KBF 470?

Choose the 260 for 262 L interior volume and 40 kg per rack when the 130 is too small and the 470 is unnecessary. All share 0...70 °C; the 260 uses the triangular instrument panel of sizes 130/260, while 470 and larger use the signal triangle and adjustable control terminal.

What mounting and environmental limits apply to the KBF 260?

From size 260 a wall spacer is required to maintain the specified wall distance; the tilt protection kit does not apply to this size. The chamber is IP 20 per EN 60529 and must not be sprayed with water. Ambient dust increases condenser cleaning frequency beyond twice a year.

What maintenance and accessories are needed beyond the base delivery?

Annual maintenance and annual steam humidifier service are required, plus condenser cleaning at least twice a year; BINDER Service must reset the maintenance message after 8760 h or two years. Options not included: APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, Pt 100 object sensor, external freshwater/wastewater cans, BINDER Pure Aqua Service.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF 260 en
Datasheet ·728,2 КБ ·9 pp.
View PDF
PDF
KBF+KBF PRO (E7) 09 2025 en
Manual ·16,5 МБ ·171 pp.
View PDF