BINDER KBF 720
BINDER KBF 720 — Climate chambers with Temperature 0…70 °C.
The BINDER KBF 720 is a 727 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, housed in a 1,285 mm wide cabinet. It is the mid-large standard model in the KBF series, distinguished from smaller siblings by its 727 L interior and from the KBF PRO line by its narrower humidity range. It is used for reproducible climatic testing of bulk samples and components under stable temperature and humidity conditions.
BINDER KBF Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 727 L (KBF720-230V / KBF720UL-120V) |
| Load per rack | 40 kg (KBF720-230V / KBF720UL-120V) |
| Maximum heat compensation | 640 W |
| Nominal power | 1.6 kW (KBF720-230V / KBF720UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 5 min |
| Depth | 900 mm (KBF720-230V / KBF720UL-120V) |
| Height | 1951 mm (KBF720-230V / KBF720UL-120V) |
| Width | 1285 mm (KBF720-230V / KBF720UL-120V) |
| Power frequency | 50/60 Hz (KBF720-230V / KBF720UL-120V) |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A (KBF720-230V / KBF720UL-120V) |
| Humidity fluctuation | 1.5 % RH |
| Humidity range | 10...80 % RH |
Description
Construction and series positioning
- The KBF 720 (Art. No. 9020-0481, designation 720-230) is a floor-standing constant climate chamber with a 727 L interior volume and net dimensions of 1,285 mm W × 1,951 mm H × 900 mm D. It uses a 220...240 V, 1~, 50/60 Hz supply with a 16 A unit fuse and 1.6 kW nominal power.
- Within the KBF series, the 720 sits between the 470 and 1060 models; all share the 0...70 °C base temperature range. The KBF PRO variants (130–1600) offer an enlarged temperature and humidity range, so the 720 is the standard-range choice when the application stays within 0...70 °C and 10...80 % RH.
- The chamber is equipped with a temperature safety device per DIN 12880:2007 class 2/3.3 and an over-temperature protective device class 1. The MB2 controller provides fixed-value, time-program and week-program operating modes.
Behavior under load and in different media
- Temperature fluctuation is 0.1 K and temperature uniformity 0.3 K; humidity fluctuation is 1.5 % RH. These figures apply to the controlled working volume and should be used when defining sensor placement and sample loading.
- Recovery after a 30 s door opening is 2 min at 25 °C/60 % RH and 5 min at 40 °C/75 % RH, so short access events have limited impact on steady-state conditions.
- Maximum heat compensation is 640 W; loads with high heat dissipation must be checked against this limit. Maximum load per rack is 40 kg.
- The humidity sensor has a tolerated error of ±2 % RH. It must not be exposed to substances above the tolerated concentrations listed in the manual (e.g. NH3 20 ppm TLV / 5500 ppm permanent load; HCl 2 ppm TLV / 300 ppm permanent load).
Installation, connection and control
- From size 260 upward a wall spacer is required; the 720 therefore needs the specified rear wall distance. The rear connection panel carries the water connections for humidification and wastewater/condensate handling.
- Humidification requires a freshwater supply; the humidifying system and wastewater/condensate handling differ by serial number (up to 20250000010225 and from 20250000010226), so the correct manual section and spare parts must be selected by serial number.
- The MB2 controller supports Ethernet and USB data transfer, alarm and notification functions, a chart-recorder display, and user management with password-protected authorization levels. Options include APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock and a flexible Pt 100 object temperature sensor.
Integration & Operating Notes
Because the 720 is larger than size 260, the wall spacer must be fitted and the specified distance to the wall kept. Blocking the rear connection panel or reducing this gap impairs heat rejection and access to the water connections, and can push the refrigeration system outside its intended operating conditions.
The KBF720-230V is rated 220...240 V, 1~, 50/60 Hz with 1.6 kW nominal power and a 16 A unit fuse. Connecting it to a 120 V supply or to a circuit with insufficient overcurrent protection will prevent startup or cause nuisance tripping. The separate KBF720UL-120V variant uses a NEMA 5-20 plug and must not be confused with this model.
With humidity active the working range narrows to 10...70 °C; the full 0...70 °C range applies only without humidity control. Set-points outside these limits are not achievable, and condensation risk increases near the lower humidity temperature boundary, so process recipes must be written inside the combined envelope.
The chamber can compensate a maximum of 640 W of heat from the load. Exceeding this with powered test items or exothermic samples will cause the actual temperature to drift above set-point and invalidate uniformity figures. Calculate the total dissipation of all samples per rack (max 40 kg per rack) before loading.
The humidity sensor tolerates only limited concentrations of substances such as ammonia, hydrochloric acid, hydrogen sulphide and ozone. Permanent exposure above the listed values (e.g. NH3 above 5500 ppm, HCl above 300 ppm) will degrade the sensor and shift readings beyond the ±2 % RH tolerance, so outgassing samples must be evaluated against the manual's resistance table first.
The steam humidifier needs a freshwater connection and a wastewater/condensate path at the rear panel. The connection method depends on the serial number (up to or from 20250000010225), so the installation must be planned against the actual unit's serial number. Water quality affects humidifier maintenance intervals and steam output.
The chamber has an over-temperature protective device class 1 and a safety controller class 2/3.3 that can be set as temperature limiter or temperature protection, plus undertemperature protection class 3.2. Limit and offset values must be set for the process before time or week programs are started, otherwise an alarm state will halt the program and require manual reset.
Regular maintenance is required at least once a year, including the steam humidifier, and the condensers should be cleaned at least twice a year (more often in dusty ambient air). Refrigeration work must be done by personnel certified to EN 13313:2010. After 8760 operating hours or two years the controller shows a maintenance message every two weeks until BINDER Service resets it.
A test before initial startup and before resuming operation after maintenance or repair is required, and annual repeat inspection is recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702. This should be built into the site's commissioning checklist and annual service schedule rather than treated as optional.
Selection & Integration Q&A
Pick the KBF 720 when you need 727 L of usable volume and your process stays within 0...70 °C and 10...80 % RH. Choose the KBF 260 only for smaller batches; choose the KBF PRO 720 if you need the enlarged temperature and humidity range beyond the standard KBF envelope.
It requires 220...240 V, 1~, 50/60 Hz with a 16 A fuse and draws 1.6 kW nominal. A 120 V site needs the separate KBF720UL-120V variant (NEMA 5-20 plug), not this model; the two must not be interchanged.
The wall spacer must be fitted and the specified rear distance kept, since the 720 is above size 260. The rear panel must remain accessible for water connections. IP 20 means no water spraying; condenser cleaning frequency rises with dust in the ambient air.
Plan annual service including the steam humidifier, twice-yearly condenser cleaning, and annual electrical safety testing per EN 50678/50699. Options not included: APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, Pt 100 object sensor, and external freshwater/wastewater cans.