BINDER KBF 720
BINDER ·Made in Germany · In Active Production

BINDER KBF 720

BINDER KBF 720 — Climate chambers with Temperature 0…70 °C.

The BINDER KBF 720 is a 727 L constant climate chamber with a 0...70 °C temperature range and 10...80 % RH humidity control, housed in a 1,285 mm wide cabinet. It is the mid-large standard model in the KBF series, distinguished from smaller siblings by its 727 L interior and from the KBF PRO line by its narrower humidity range. It is used for reproducible climatic testing of bulk samples and components under stable temperature and humidity conditions.

Temperature
0…70 °C

BINDER KBF Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF 720 Current 0–70727406401.60.3
KBF 1060 0–701070406401.60.3
KBF 130 0–70127302701.60.2
KBF 1600 0–701610406401.70.4
KBF 260 0–70262404101.40.3
KBF 470 0–70472405301.40.2

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 727 L (KBF720-230V / KBF720UL-120V)
Load per rack 40 kg (KBF720-230V / KBF720UL-120V)
Maximum heat compensation 640 W
Nominal power 1.6 kW (KBF720-230V / KBF720UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 2 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 5 min
Dimensions, Weight & Mechanical Design
Depth 900 mm (KBF720-230V / KBF720UL-120V)
Height 1951 mm (KBF720-230V / KBF720UL-120V)
Width 1285 mm (KBF720-230V / KBF720UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (KBF720-230V / KBF720UL-120V)
Rated voltage 220...240 V
Unit fuse 16.0 A (KBF720-230V / KBF720UL-120V)
Operating Conditions, Environment & Protection
Humidity fluctuation 1.5 % RH
Humidity range 10...80 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection and control

Integration & Operating Notes

Mounting
Maintain rear wall distance with the supplied spacer

Because the 720 is larger than size 260, the wall spacer must be fitted and the specified distance to the wall kept. Blocking the rear connection panel or reducing this gap impairs heat rejection and access to the water connections, and can push the refrigeration system outside its intended operating conditions.

Operating Manual KBF/KBF PRO (E7), Section 4.1 Installation
Electrical
Verify site supply matches the 230 V variant before connection

The KBF720-230V is rated 220...240 V, 1~, 50/60 Hz with 1.6 kW nominal power and a 16 A unit fuse. Connecting it to a 120 V supply or to a circuit with insufficient overcurrent protection will prevent startup or cause nuisance tripping. The separate KBF720UL-120V variant uses a NEMA 5-20 plug and must not be confused with this model.

Data Sheet Model KBF 720; Operating Manual Section 1.6 Type Plate
Process Limits
Humidity control is limited to 10...80 % RH and 10...70 °C

With humidity active the working range narrows to 10...70 °C; the full 0...70 °C range applies only without humidity control. Set-points outside these limits are not achievable, and condensation risk increases near the lower humidity temperature boundary, so process recipes must be written inside the combined envelope.

Factory Specification Sheet; Operating Manual Section 18 Humidification/Dehumidification
Process Limits
Keep sample heat load below 640 W heat compensation

The chamber can compensate a maximum of 640 W of heat from the load. Exceeding this with powered test items or exothermic samples will cause the actual temperature to drift above set-point and invalidate uniformity figures. Calculate the total dissipation of all samples per rack (max 40 kg per rack) before loading.

Factory Specification Sheet
Process Limits
Do not expose the humidity sensor to aggressive media

The humidity sensor tolerates only limited concentrations of substances such as ammonia, hydrochloric acid, hydrogen sulphide and ozone. Permanent exposure above the listed values (e.g. NH3 above 5500 ppm, HCl above 300 ppm) will degrade the sensor and shift readings beyond the ±2 % RH tolerance, so outgassing samples must be evaluated against the manual's resistance table first.

Operating Manual Section 1.14 Humidity Sensor Resistance Table
Process Media
Humidification requires a freshwater supply and drain

The steam humidifier needs a freshwater connection and a wastewater/condensate path at the rear panel. The connection method depends on the serial number (up to or from 20250000010225), so the installation must be planned against the actual unit's serial number. Water quality affects humidifier maintenance intervals and steam output.

Operating Manual Sections 4.4 and 18.1
Control & I/O
Configure the temperature safety device before running programs

The chamber has an over-temperature protective device class 1 and a safety controller class 2/3.3 that can be set as temperature limiter or temperature protection, plus undertemperature protection class 3.2. Limit and offset values must be set for the process before time or week programs are started, otherwise an alarm state will halt the program and require manual reset.

Operating Manual Section 14 Temperature Safety Devices
Maintenance
Plan annual service and twice-yearly condenser cleaning

Regular maintenance is required at least once a year, including the steam humidifier, and the condensers should be cleaned at least twice a year (more often in dusty ambient air). Refrigeration work must be done by personnel certified to EN 13313:2010. After 8760 operating hours or two years the controller shows a maintenance message every two weeks until BINDER Service resets it.

Operating Manual Section 23 Maintenance and Service
Maintenance
Include electrical safety testing at commissioning and annually

A test before initial startup and before resuming operation after maintenance or repair is required, and annual repeat inspection is recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702. This should be built into the site's commissioning checklist and annual service schedule rather than treated as optional.

Operating Manual Section 23.1 General Information

Selection & Integration Q&A

When should I choose the KBF 720 rather than a smaller KBF 260 or the KBF PRO 720?

Pick the KBF 720 when you need 727 L of usable volume and your process stays within 0...70 °C and 10...80 % RH. Choose the KBF 260 only for smaller batches; choose the KBF PRO 720 if you need the enlarged temperature and humidity range beyond the standard KBF envelope.

What supply and protection does the KBF720-230V need, and can it run on a 120 V site?

It requires 220...240 V, 1~, 50/60 Hz with a 16 A fuse and draws 1.6 kW nominal. A 120 V site needs the separate KBF720UL-120V variant (NEMA 5-20 plug), not this model; the two must not be interchanged.

What mounting clearances and ambient conditions must be respected for the KBF 720?

The wall spacer must be fitted and the specified rear distance kept, since the 720 is above size 260. The rear panel must remain accessible for water connections. IP 20 means no water spraying; condenser cleaning frequency rises with dust in the ambient air.

What maintenance and accessories are needed beyond the base chamber delivery?

Plan annual service including the steam humidifier, twice-yearly condenser cleaning, and annual electrical safety testing per EN 50678/50699. Options not included: APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, Pt 100 object sensor, and external freshwater/wastewater cans.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF 720 en
Datasheet ·690,7 КБ ·9 pp.
View PDF
PDF
KBF+KBF PRO (E7) 09 2025 en
Manual ·16,5 МБ ·171 pp.
View PDF