BINDER KBF PRO 1060
BINDER KBF PRO 1060 — Climate chambers with Temperature -20…100 °C.
The KBF PRO 1060 is a 1,070 L constant climate chamber with a temperature range of -20 to 100 °C and humidity control from 10 to 98 % RH, extending the standard KBF series into a wider humidity envelope. It is distinguished within the KBF PRO series by its large interior volume and 40 kg per-rack load capacity, making it suited for high-throughput stability and climatic testing of bulky or densely packed samples.
BINDER KBF PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW | Temperature uniformity, K |
|---|---|---|---|---|---|---|
| KBF PRO 1060 Current | -20–100 | 1070 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 130 | -20–100 | 127 | 30 | 450 | 1.9 | 0.2 |
| KBF PRO 1600 | -20–100 | 1610 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 260 | -20–100 | 262 | 40 | 800 | 1.6 | 0.2 |
| KBF PRO 470 | -20–100 | 472 | 40 | 800 | 1.6 | 0.3 |
| KBF PRO 720 | -20–100 | 727 | 40 | 1000 | 1.8 | 0.3 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 1070 L |
| Load per rack | 40 kg |
| Maximum heat compensation | 1000 W |
| Nominal power | 1.9 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 6 min |
| Net weight of the unit (empty) mfr | 360 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity fluctuation | 1 % RH |
| Humidity range | 10...98 % RH |
Description
Construction and Series Positioning
- The KBF PRO 1060 is the largest volume model in the KBF PRO series, sharing the -20…100 °C temperature range with siblings KBF PRO 130, 260, 470 and 1600 but offering 1,070 L of interior volume.
- Unlike the standard KBF series, the PRO variant extends the humidity range to 10…98 % RH and permits humidity control across 10…90 °C, enabling combined temperature-humidity profiles beyond the standard KBF envelope.
- The chamber is equipped with a signal triangle for sizes 470, 720, 1060 and 1600, and an ergonomically adjustable control terminal for larger sizes, as described in the operating manual.
Behavior Under Load and in Different Media
- Temperature fluctuation is 0.1 K and temperature uniformity is 0.3 K, supporting stable conditions for sensitive samples.
- Humidity fluctuation is 1 % RH; the humidity sensor has a tolerated error of ±2 % r.h., which must be considered when interpreting readings.
- Recovery after a 30 s door opening is 2 min at 25 °C and 60 % RH, and 6 min at 40 °C and 75 % RH, indicating rapid restoration of set conditions after access.
- The chamber can compensate up to 1,000 W of heat load, allowing operation with active samples or equipment inside the workspace.
Installation, Connection and Control
- The chamber requires a 208…240 V, 1-phase, 50 Hz supply with a 16 A unit fuse; nominal power is 1.9 kW.
- A wall spacer is required from size 260 onward, so the KBF PRO 1060 must maintain the specified wall distance for proper operation.
- The MB2 chamber controller provides fixed-value, time-program and week-program modes, chart-recorder display, alarm functions and user management with password protection.
- Humidity control requires freshwater and wastewater connections; the humidifying and wastewater handling design differs by serial number (up to 20250000010225 and from 20250000010226), so the correct connection method must be identified from the serial number.
- Temperature safety devices include an over temperature protective device class 1 and an overtemperature safety controller class 2/3.3 (selectable as temperature limiter or temperature protection), plus undertemperature protection class 3.2.
Integration & Operating Notes
The KBF PRO 1060 requires a wall spacer to maintain the specified distance from adjacent surfaces. Insufficient clearance can impair heat dissipation and humidity control, leading to reduced performance or chamber faults.
The chamber is rated for 208…240 V, single-phase, 50 Hz with a 16 A unit fuse and 1.9 kW nominal power. Connecting to a different voltage or frequency may damage the unit and void warranty; ensure overcurrent protection is correctly sized.
Although the chamber operates from -20 to 100 °C, humidity control is only active between 10 and 90 °C and within 10…98 % RH. Operating humidity outside these limits may not be possible or may compromise accuracy.
The chamber can compensate a maximum heat load of 1,000 W from samples or equipment. Exceeding this value may prevent the chamber from maintaining set temperature, especially at low setpoints, and can lead to extended recovery times or alarms.
The humidity sensor must not be exposed to substances above the listed tolerated concentrations, such as ammonia (20 ppm TLV, 5500 ppm permanent load), acetone (500 ppm TLV, 3300 ppm permanent load), or formaldehyde (0.3 ppm TLV, 2400 ppm permanent load). Exceeding these limits can damage the sensor and invalidate humidity readings.
Regular maintenance must be performed at least once a year, including steam humidifier maintenance. Condensers should be cleaned at least twice a year by a qualified technician; in dusty environments, clean more frequently by suction. Warranty may be void if maintenance is not conducted by authorized personnel.
After 8760 operating hours or two years, a maintenance message appears and repeats every two weeks until reset by BINDER Service. Schedule the service reset before this interval to avoid recurring reminders and to maintain warranty compliance.
The refrigeration system contains flammable R600A refrigerant. Any maintenance on the refrigeration system must only be conducted by qualified personnel trained according to EN 13313:2010 (e.g., refrigeration technician with certified expert knowledge per Regulation (EC) No. 303/2008).
An annual repeat inspection and a test prior to initial startup and prior to resuming operation after maintenance or repair are required, recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702. Include these tests in commissioning and maintenance schedules.
The humidifying system and wastewater/condensate handling differ up to serial number 20250000010225 and from serial number 20250000010226. Before connecting freshwater and wastewater, confirm the chamber serial number to select the correct connection method and spare parts.
Selection & Integration Q&A
Choose the 1060 for high-throughput testing of bulky or densely packed samples requiring up to 1,070 L and 40 kg per rack. The 260 suits smaller batches; the 1600 offers more volume but may exceed space or utility capacity.
The chamber requires freshwater and wastewater connections for humidity control. The connection method depends on serial number (up to or from 20250000010225/20250000010226). Ensure supply pressure does not exceed 11 bar and use appropriate tubing.
Base delivery excludes APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, flexible Pt 100 sensor, external freshwater/wastewater cans, and BINDER Pure Aqua Service. Annual maintenance and condenser cleaning are also not included.