BINDER KBF PRO 130
BINDER ·Made in Germany · In Active Production

BINDER KBF PRO 130

BINDER KBF PRO 130 — Climate chambers with Temperature -20…100 °C.

The KBF PRO 130 is a 127 L constant climate chamber with a -20…100 °C temperature range and 10…98 % RH humidity range, distinguished within the KBF PRO series by its compact 130-size cabinet and triangular instrument panel. It is intended for precise, reproducible climatic testing of small to medium samples, with 0.1 K temperature fluctuation and 0.2 K uniformity.

Temperature
-20…100 °C

BINDER KBF PRO Series — Model Comparison & Selection

Model Temperature, °CInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kWTemperature uniformity, K
KBF PRO 130 Current -20–100127304501.90.2
KBF PRO 1060 -20–10010704010001.90.3
KBF PRO 1600 -20–10016104010001.90.3
KBF PRO 260 -20–100262408001.60.2
KBF PRO 470 -20–100472408001.60.3
KBF PRO 720 -20–1007274010001.80.3

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 127 L
Load per rack 30 kg
Maximum heat compensation 450 W
Nominal power 1.9 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 2 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 3 min
Dimensions, Weight & Mechanical Design
Net weight of the unit (empty) mfr 142 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1
Operating Conditions, Environment & Protection
Humidity fluctuation 1 % RH
Humidity range 10...98 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection and control

Integration & Operating Notes

Mounting
Install tilt protection kit on the 130-size cabinet

The flexible tilt protection kit is specified for chamber sizes 130 and 470. It must be mounted to prevent tipping during operation or relocation. Unlike larger sizes from 260 upward, no wall spacer is required for the 130, so the chamber can be placed closer to walls, but the tilt protection still applies.

Operating Manual KBF/KBF PRO (E7) 09-2025, Section 4.2, p.33
Electrical
Verify site voltage and fuse rating before connection

The chamber is rated for 208…240 V, 50 Hz, 1.9 kW, with a 16 A unit fuse. Connecting to a supply outside this range or with insufficient overcurrent protection can damage the unit or cause unsafe operation. Confirm the site voltage and that the circuit breaker or fuse matches the 16 A rating.

Data Sheet Model KBF PRO 130 en.pdf; Operating Manual KBF/KBF PRO (E7) 09-2025, Section 1.6 Type plate
Process Limits
Humidity control limited to 10…90 °C chamber temperature

Although the chamber temperature range is -20…100 °C, humidity control is only permitted between 10 and 90 °C. Operating humidity outside this window may lead to condensation, inaccurate control or equipment stress. Plan test profiles so that humidification phases stay within this temperature band.

Data Sheet Model KBF PRO 130 en.pdf, extra specification 'Temperature range with humidity'
Process Limits
Do not exceed 450 W sample heat load

The maximum heat compensation is 450 W. Samples dissipating more than this will exceed the chamber's ability to maintain setpoint, potentially causing temperature deviation or safety device activation. Calculate the total heat output of all samples per test and stay well below this limit.

Data Sheet Model KBF PRO 130 en.pdf, specification 'max-heat-compensation'
Maintenance
Plan annual maintenance and condenser cleaning twice yearly

Regular maintenance is required at least once a year, including the steam humidifier. Condensers should be cleaned at least twice a year by a qualified technician; in dusty environments, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and repeats every two weeks until reset by BINDER Service.

Operating Manual KBF/KBF PRO (E7) 09-2025, Section 23.2, p.133-134
Process Media
Humidity sensor tolerates only limited aggressive substances

The humidity sensor must not be exposed to substances above the listed tolerated concentrations. For example, ammonia permanent load is limited to 5500 ppm, hydrochloric acid to 300 ppm, and ozone to 0.5 ppm. Exceeding these limits can degrade sensor accuracy or damage it. Review sample outgassing before testing.

Operating Manual KBF/KBF PRO (E7) 09-2025, Section 1.14, p.20
Process Media
Freshwater and wastewater connections depend on serial number

The humidifying system and wastewater/condensate handling differ between chambers up to serial number 20250000010225 and from 20250000010226. Identify the serial number before ordering spare parts or planning plumbing. The maximum operating pressure for water connections is 11 bar.

Operating Manual KBF/KBF PRO (E7) 09-2025, Sections 18.1.1–18.1.5, p.117-118; Section 1.6 Type plate

Selection & Integration Q&A

When should I choose the KBF PRO 130 over the larger KBF PRO 260 or 470?

Choose the 130 for small samples or limited bench space: it offers 127 L and 30 kg per rack, with the same -20…100 °C range as larger models. Step up to 260 or 470 only when you need more volume or rack capacity; note that from size 260 a wall spacer is required, while the 130 uses a tilt protection kit instead.

What maintenance is required and what accessories are not included in the base delivery?

Annual maintenance is mandatory, including the steam humidifier, and condensers should be cleaned twice a year. The base delivery does not include APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, flexible Pt 100 sensor, external freshwater/wastewater cans, or BINDER Pure Aqua Service; these must be ordered separately.

Can I place the KBF PRO 130 directly against a wall, and what about tilt protection?

Yes, the 130 does not require the wall spacer that is mandatory from size 260 upward. However, the flexible tilt protection kit must be installed on the 130 (and 470) to prevent tipping. Ensure the floor is level and capable of supporting the 142 kg empty weight.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF PRO 130 en
Datasheet ·625,8 КБ ·8 pp.
View PDF
PDF
KBF+KBF PRO (E7) 09 2025 en
Manual ·16,5 МБ ·171 pp.
View PDF