BINDER KBF PRO 1600
BINDER KBF PRO 1600 — Climate chambers with Temperature -20…100 °C.
The KBF PRO 1600 is a 1,610 L constant climate chamber with a -20…100 °C range and 10…98 % RH, the largest chamber in the KBF PRO series. It is distinguished from smaller siblings by its 40 kg per-rack load capacity and 1,000 W max heat compensation, enabling high-throughput testing of bulky or heat-dissipating samples. It serves as a programmable test platform for reproducible temperature/humidity profiles in laboratory and industrial qualification.
BINDER KBF PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW | Temperature uniformity, K |
|---|---|---|---|---|---|---|
| KBF PRO 1600 Current | -20–100 | 1610 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 1060 | -20–100 | 1070 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 130 | -20–100 | 127 | 30 | 450 | 1.9 | 0.2 |
| KBF PRO 260 | -20–100 | 262 | 40 | 800 | 1.6 | 0.2 |
| KBF PRO 470 | -20–100 | 472 | 40 | 800 | 1.6 | 0.3 |
| KBF PRO 720 | -20–100 | 727 | 40 | 1000 | 1.8 | 0.3 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 1610 L |
| Load per rack | 40 kg |
| Maximum heat compensation | 1000 W |
| Nominal power | 1.9 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 3 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 7 min |
| Net weight of the unit (empty) mfr | 450 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity fluctuation | 1 % RH |
| Humidity range | 10...98 % RH |
Description
Construction and series positioning
- The KBF PRO 1600 (Art. No. 9020-0444) is the largest volume in the KBF PRO series, which also includes 130, 260, 470, 720 and 1060 L models. All share the -20…100 °C temperature range, but the 1600 adds a 1,610 L interior and a 40 kg per-rack load rating.
- The chamber uses a steam humidifier and a refrigeration system with R600A refrigerant; the humidifying and wastewater systems differ by serial number (up to/from 20250000010225), so spare parts must be selected accordingly.
- A rear connection panel provides water connections; the control terminal is ergonomically adjustable on larger sizes.
Behavior under load and in different media
- Temperature fluctuation is 0.1 K and uniformity 0.3 K, supporting stable profiles. Humidity fluctuation is 1 % RH with a sensor tolerance of ±2 % RH.
- Recovery after a 30 s door opening is 3 min at 25 °C/60 % RH and 7 min at 40 °C/75 % RH, indicating rapid return to setpoint after access.
- The 1,000 W max heat compensation allows testing of samples that dissipate heat, but the load per rack must not exceed 40 kg.
- The humidity sensor must not be exposed to substances above the tolerated concentrations listed in the manual (e.g., ammonia 5500 ppm permanent load, formaldehyde 2400 ppm).
Installation, control and operation
- The MB2 controller supports fixed-value, time-program and week-program modes, with chart-recorder display and alarm functions. Temperature safety devices include over-temperature protection class 1 and safety controller class 2/3.3 per DIN 12880.
- Electrical connection requires 208…240 V, 1~, 50 Hz, with a 16 A unit fuse; nominal power is 1.9 kW. The chamber is IP 20 and must not be sprayed with water.
- A wall spacer is required from size 260 upward, and the chamber must be connected to freshwater and wastewater lines. Maintenance includes annual service, condenser cleaning twice a year, and a maintenance reminder after 8760 h or two years.
Integration & Operating Notes
The manual specifies a wall spacer from chamber size 260 onward. For the 1600, maintain the prescribed distance to walls and adjacent equipment to prevent recirculation of exhaust air and ensure uniform temperature distribution. Failure to do so may degrade uniformity and overload the refrigeration system.
The humidity sensor has defined maximum permanent load concentrations for substances such as ammonia (5500 ppm), formaldehyde (2400 ppm) and hydrochloric acid (300 ppm). Exceeding these limits can permanently damage the sensor and invalidate humidity readings. Review the resistance table in the manual before testing materials that emit vapors.
Regular maintenance must be performed at least once a year, including the steam humidifier. Condensers should be cleaned at least twice a year by a qualified technician; with dusty ambient air, clean more frequently by suction. The maintenance reminder appears after 8760 operating hours or two years and requires reset by BINDER Service.
The chamber requires freshwater and wastewater connections. The maximum operating pressure is 11 bar, so all supply fittings and hoses must be rated accordingly. The humidifying and wastewater systems differ by serial number; identify the chamber serial number to select the correct connection method and spare parts.
Selection & Integration Q&A
Choose the 1600 when you need the 1,610 L interior volume and a 40 kg per-rack load capacity for large or heavy test batches. The 470 and 1060 share the same temperature range but offer less volume and lower load ratings, so they are suitable for smaller sample sets.
Perform maintenance at least annually, including the steam humidifier. Clean condensers at least twice a year (more often in dusty environments). The controller displays a maintenance reminder after 8760 operating hours or two years, which must be reset by BINDER Service.
The base delivery does not include the wall spacer (required from size 260), freshwater/wastewater cans, APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, or flexible Pt 100 sensor. These must be ordered separately as options or accessories.