BINDER KBF PRO 260
BINDER KBF PRO 260 — Climate chambers with Temperature -20…100 °C.
The KBF PRO 260 is a 262 L constant climate chamber with a temperature range of -20 to 100 °C and humidity control from 10 to 98 % RH, distinguishing it from smaller or larger siblings by its mid-range volume and 40 kg per rack load capacity. It is designed for precise temperature and humidity testing, offering uniformity of 0.2 K and fluctuation of 0.1 K, with a maximum heat compensation of 800 W to handle specimen loads.
BINDER KBF PRO Series — Model Comparison & Selection
| Model | Temperature, °C | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW | Temperature uniformity, K |
|---|---|---|---|---|---|---|
| KBF PRO 260 Current | -20–100 | 262 | 40 | 800 | 1.6 | 0.2 |
| KBF PRO 1060 | -20–100 | 1070 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 130 | -20–100 | 127 | 30 | 450 | 1.9 | 0.2 |
| KBF PRO 1600 | -20–100 | 1610 | 40 | 1000 | 1.9 | 0.3 |
| KBF PRO 470 | -20–100 | 472 | 40 | 800 | 1.6 | 0.3 |
| KBF PRO 720 | -20–100 | 727 | 40 | 1000 | 1.8 | 0.3 |
Full specifications
| Temperature | -20...100 °C |
| Interior volume | 262 L |
| Load per rack | 40 kg |
| Maximum heat compensation | 800 W |
| Nominal power | 1.6 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 2 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 3 min |
| Net weight of the unit (empty) mfr | 197 kg |
| Power frequency | 50 Hz |
| Rated voltage | 208...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity fluctuation | 1 % RH |
| Humidity range | 10...98 % RH |
Description
Construction and Series Positioning
The KBF PRO 260 (article number 9020-0440) is a constant climate chamber from the KBF PRO series, which includes sizes 130, 260, 470, 720, 1060, and 1600. This model offers a 262 L interior volume, positioning it between the smaller KBF PRO 130 and the larger KBF PRO 470. It shares the same temperature range (-20 to 100 °C) as its siblings but provides a specific balance of chamber size and load capacity, with a maximum load of 40 kg per rack. The chamber features a triangular instrument panel for sizes 130 and 260, and a rear connection panel for water connections. The unit has a net weight of 197 kg and requires a wall spacer from size 260 onward to ensure proper ventilation.
Performance and Behavior
Under load, the chamber maintains tight temperature control with a uniformity of 0.2 K and fluctuation of 0.1 K. Humidity control spans 10 to 98 % RH with a fluctuation of 1 % RH. The maximum heat compensation of 800 W allows the chamber to handle specimens that generate heat, ensuring stable conditions. Recovery times after door openings are rapid: 2 minutes at 25 °C and 60 % RH, and 3 minutes at 40 °C and 75 % RH, minimizing downtime during testing. The refrigeration system uses R600A refrigerant (charge -0.095 kg), and the chamber has a maximum operating pressure of 11 bar for water connections.
Installation, Control, and Operation
The chamber requires a 208...240 V, 1-phase, 50 Hz power supply with a nominal power of 1.6 kW and a 16 A unit fuse. It is equipped with the MB2 chamber controller, which supports fixed value and program modes, including time and week programs, and offers a chart recorder display. The controller includes temperature safety devices per DIN 12880:2007 class 2/3.3, with over temperature protection and undertemperature protection. For humidity control, the chamber requires a freshwater connection and wastewater drainage; the humidifying system design differs by serial number (up to or from 20250000010225). The chamber has IP 20 protection, so it must not be exposed to water spray. Regular maintenance is essential: annual service, condenser cleaning twice a year, and a maintenance reminder after 8760 operating hours or two years.
Integration & Operating Notes
The chamber requires a spacer for wall distance from size 260 onward to ensure adequate airflow and prevent overheating. Failure to maintain the specified clearance can lead to reduced cooling efficiency and potential damage to the refrigeration system.
The chamber is rated for 208...240 V, 1~, 50 Hz, with a nominal power of 1.6 kW and a 16 A fuse. Connecting to a different voltage or frequency can damage the electrical components and void warranty. Ensure overcurrent protection is correctly sized.
The humidity sensor tolerates specific concentrations of substances such as ammonia, acetone, ethanol, and formaldehyde. Exceeding the listed permanent load concentrations can impair sensor accuracy and lead to false readings. Refer to the resistance table for maximum workplace threshold values.
Regular maintenance must be performed at least once a year, including steam humidifier service. Condensers should be cleaned at least twice a year, or more frequently in dusty environments. The maintenance reminder appears after 8760 operating hours or two years and must be reset by BINDER Service.
The chamber is equipped with a temperature safety device class 2/3.3 per DIN 12880. The safety controller can be set as a temperature limiter or temperature protection, with adjustable limit and offset values. Ensure the correct mode is selected for the specific test requirements to prevent overheating or freezing of samples.
Selection & Integration Q&A
Choose the 260 for a 262 L volume and 40 kg per rack load, balancing space and capacity. The 130 is smaller (130 L), while the 470 offers more space but may be oversized for smaller batches. All share the same temperature range.
The chamber requires a freshwater connection and wastewater drainage. Ensure the water pressure does not exceed 11 bar and that connections are rated accordingly. The humidifying system design varies by serial number, so identify the serial number for correct installation.
Annual maintenance by authorized personnel, condenser cleaning twice a year, and door gasket replacement when cold. Accessories like APT-COM 4 software, analog outputs, and Pt 100 sensor are optional. A maintenance reminder appears after 8760 hours or two years.