BINDER VD 23
BINDER VD 23 — Drying ovens with Temperature 9…220 °C.
The BINDER VD 23 is a 24 L vacuum drying chamber with a 16 mm small-flange vacuum connection, operating from +9 to 220 °C with 0.1 K fluctuation and 1.0 K uniformity. As the smallest of the VD 23/56/115 series, it suits benchtop loads up to 50 kg (20 kg per rack) where space and sample volume are limited. It is used for gentle drying of non-flammable solvents under vacuum, with evacuation, ventilation and inert-gas operation managed by the RD4 controller.
BINDER VD Series — Model Comparison & Selection
Full specifications
| Temperature | 9...220 °C |
| Heating time to 100°C | 80 min |
| Interior volume | 24 L (VD023-230V) |
| Leak rate | 0.01 bar/h |
| Load per rack | 20 kg (VD023-230V) |
| Nominal power | 0.9 kW |
| Permitted load | 50 kg (VD023-230V) |
| Pressure fluctuation over time | 1 mbar |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 1.0 K |
| Rear wall clearance | 100 mm (VD023-230V) |
| Side wall clearance | 70 mm (VD023-230V) |
| Article Number mfr | 9630 |
| Pressure fluctuation over time at 400 mbar mfr | 70 mbar |
| Depth | 413 mm (VD023-230V) |
| Height | 698 mm (VD023-230V) |
| Width | 523 mm (VD023-230V) |
| Power frequency | 50 Hz |
| Rated voltage | 200...230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
| Vacuum connection with small flange mfr | 16 mm |
Description
Construction and series positioning
- The VD 23 is the 24 L member of the VD series (siblings VD 56 and VD 115 share the +9 to 220 °C range but larger interiors).
- Interior dimensions 413 mm depth × 698 mm height × 523 mm width (net, VD023-230V); chamber volume 24 L.
- Corrosion-resistant stainless-steel inner chamber; delivered with factory calibration and adjustment.
- Standard controller is the RD4 microprocessor program controller; the MB2 is optional. Both support temperature and pressure set-point entry, programs, alarms and data handling.
- Two electrical variants: VD023-230V (200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Art. 9630-0001) and VD023UL-120V (100...120 V 1~ 50/60 Hz, NEMA 5-15 plug, Art. 9630-0005).
Behavior under load and vacuum
- Temperature range +9 to 220 °C; temperature fluctuation 0.1 K and uniformity 1.0 K define the working tolerance for drying processes.
- Heating-up time to 100 °C is 80 min at 0.9 kW nominal power.
- Vacuum interface is a 16 mm small flange; leak rate 0.01 bar/h and pressure fluctuation over time 1 mbar (70 mbar at 400 mbar per extra spec) indicate the pressure stability achievable.
- Permitted load 50 kg total, 20 kg per rack. Use only the delivered expansion racks and do not mix aluminum and stainless-steel racks, as this invalidates calibration and temperature performance.
- The vacuum system supports evacuation, ventilation after drying (ambient air or inert gas), ventilation/breaking vacuum after power failure, and inert-gas operation.
Installation, control and operation
- Wall clearances: 100 mm at the rear and 70 mm at the sides (VD023-230V). IP protection 20 per EN 60529; protect from water ingress.
- Electrical supply: 200...230 V, 1-phase, 50/60 Hz, unit fuse 6.3 A. The 120 V UL variant is for North American supplies.
- The RD4 controller provides set-point entry, program start/stop, tolerance-range and alarm-delay settings, a data recorder (DL1) and USB data transfer; Ethernet and APT-COM 4 software enable networking.
- Temperature safety device class 2.0 per DIN 12880 (overtemperature protective device class 2 / thermal switch) is standard; alarm messages “Safety controller” and “Overtemperature” require cooling, acknowledgement and setpoint/safety-value checks.
- Annual maintenance is required at least once a year; electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended at initial startup and after maintenance or repair.
Integration & Operating Notes
The 230 V version (Art. 9630-0001) requires 200...230 V 1~ 50/60 Hz with a CEE 7/7 plug and 6.3 A unit fuse; the UL 120 V version (Art. 9630-0005) uses a NEMA 5-15 plug. Connecting the wrong variant risks damage and invalidates safety certification; check the type plate and supply voltage during commissioning.
The permitted load is 50 kg for the chamber and 20 kg per expansion rack. Exceeding these limits can deform racks or the chamber interior and compromise temperature uniformity. Distribute samples evenly and use only the delivered racks, keeping rack material consistent (no mixing of aluminum and stainless steel).
The VD 23 is intended for drying non-flammable solvents. Introducing flammable substances creates a fire or explosion risk because the chamber is not designed for them. Verify solvent compatibility with the stainless-steel interior; corrosive damage from certain ingredients is excluded from BINDER liability.
If the “Safety controller” alarm appears, the inner chamber temperature has reached the safety controller value. Set the safety controller value above the intended temperature setpoint and check for heating-system errors. After the alarm, let the chamber cool, acknowledge the alarm, and verify both values before restarting.
Maintenance must be performed at least once a year, and an annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Maintenance by non-authorized personnel voids the warranty. Replace the door gasket only when cold to avoid damaging it.
If the chamber is permanently turned off, troubleshooting includes checking the chamber fuse (6.3 A). A blown fuse may indicate an electrical fault; disconnect the power plug before any maintenance work and do not remove the rear panel. Only licensed electricians or BINDER-authorized experts should perform maintenance.
Selection & Integration Q&A
Choose the VD 23 for small loads and limited bench space: 24 L interior, 50 kg total load, 20 kg per rack. The VD 56 and VD 115 offer larger interiors for higher throughput but share the same +9 to 220 °C range and vacuum principle. Select the VD 23 when sample volume fits 24 L and portability or footprint matters.
The chamber has a 16 mm small-flange vacuum connection. A vacuum pump is not included; BINDER offers a pump module with pump stand and connection kit, and VD connection kits for VP 1 / VP 2 pumps. Ensure the pump is suitable for the 0.01 bar/h leak rate and 1 mbar pressure fluctuation specification.
Annual maintenance is required, plus electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. The base delivery includes the chamber and expansion racks; a vacuum pump, pump module, connection kits, APT-COM 4 software, analog outputs, Pt 100 object sensor, LED lighting and measuring access port are optional accessories. Door gaskets are consumables replaced when cold.