Peter Huber Kältemaschinenbau CC-300BX Brückenthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-300BX Brückenthermostat

Peter Huber Kältemaschinenbau CC-300BX Brückenthermostat — Thermostatic baths.

The CC-300BX is a bridge-mounted immersion circulator (Brückenthermostat) with a variable-speed pressure/suction pump and Pilot ONE controller, spanning 28–300 °C on its own, 20–300 °C with water cooling and -20–300 °C with a refrigerator. Its 3–3.5 kW heater and ±0.02 K stability distinguish it from the CC-200BX in the series. It serves external closed-loop temperature control of baths and connected equipment.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Brueckenthermostate Series — Model Comparison & Selection

Model Current consumption, ADimensions w d hMax delivery flow rate, liter/minuteTemperature range, °CWeight, kgAmbient temperature limit, °C
CC-300BX Brückenthermostat Current 16345 x 190 x 3922528–300135–40
CC-200BX Brückenthermostat 10345 x 200 x 3262228–20075–40

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 345 x 190 x 392 mm
Maximum delivery flow rate 25 l/min (pressure pump; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 28...300 °C
with cooling unit: -20...300 °C · with water cooling: 20...300 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 3 - 3, 5 kW
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump; 0, 4 bar)
Dimensions, Weight & Mechanical Design
Weight 13 kg
Electrical, Power & Drive Specifications
Current consumption 16 A
Fuse rating 20 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Provide a UL 489 branch protective device before commissioning

The datasheet makes external branch circuit protection to UL 489 mandatory. Size it for the 16 A maximum current and the 20 A internal fuse, and confirm the site supply is 200–240 V at 50/60 Hz with voltage and frequency tolerances not running in opposite directions.

CC-300BX Immersion Circulator — Technical Data Sheet, p.2
Electrical
Order the country-specific plug with the unit

Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified at order time. Without this, the delivered cable may not mate with the site receptacle and commissioning is delayed.

CC-300BX Immersion Circulator — Technical Data Sheet, p.2
Process Limits
Cooling method fixes the usable low-temperature end

The unit reaches 28 °C without auxiliary cooling, 20 °C with water cooling and -20 °C with a refrigerator. Select the cooling configuration against the lowest process temperature before specifying; the base unit alone cannot go below 28 °C.

CC-300BX Immersion Circulator — Technical Data Sheet, p.1
Cooling & Media
Keep thermofluid viscosity below 50 mm²/s

The maximum permissible kinematic viscosity is 50 mm²/s. At the low end of the range viscosity rises, so the fluid must be chosen to stay within this limit at the coldest operating point, otherwise pump delivery and heat transfer degrade.

CC-300BX Immersion Circulator — Technical Data Sheet, p.1
Mounting
Match bath opening to the telescopic bridge span

The telescopic arm extends to 884 mm and immersion depth is 175 mm. The bath opening and wall geometry must accommodate the bridge span and place the immersion section at the correct depth; the M16x1 male pump connection then dictates the mating hose or adapter.

CC-300BX Immersion Circulator — Technical Data Sheet, p.1
Mounting
Install only in a dry, dust-protected indoor location

The housing is IP20 and ambient must stay between 5 and 40 °C. Washdown areas, wet rooms and dusty environments are excluded; site planning must verify the room conditions before the unit is placed.

CC-300BX Immersion Circulator — Technical Data Sheet, p.1
Control & I/O
Set the adjustable overtemperature cut-out for the application

Overtemperature protection per DIN 12876 is adjustable and the unit carries safety class III/FL. The cut-out must be configured for the bath and medium in use so that a control fault cannot overheat the process.

CC-300BX Immersion Circulator — Technical Data Sheet, p.1
Maintenance
Register the unit to retain the 4-year warranty

The 4-year warranty is conditional on registration. Complete registration at commissioning so that warranty coverage is preserved for the equipment.

CC-300BX Immersion Circulator — Technical Data Sheet, p.1

Selection & Integration Q&A

When should I choose the CC-300BX rather than the CC-200BX in this series?

Pick the CC-300BX when the process needs the higher 3–3.5 kW heating capacity and the 16 A supply it implies, or the full -20–300 °C span with a refrigerator. The CC-200BX is the lower-capacity sibling; confirm its ratings against your load before downsizing.

What must I provide on the electrical and hydraulic side to integrate this circulator?

A 200–240 V, 1~ or 2~, 50/60 Hz supply rated for 16 A with a UL 489 branch protective device, plus the country-specific plug specified at order. Hydraulically, M16x1 male pump connections require matching hoses or adapters, and the external loop must fit within 25 l/min at 0.7 bar pressure and 18.5 l/min at 0.4 bar suction.

What site and duty conditions limit where this unit can be installed?

Ambient must remain 5–40 °C, the housing is IP20 so water and dust ingress must be excluded, and the unit is Class A Group 1 to EN55011, intended for non-domestic environments. The telescopic bridge needs a bath opening within its 884 mm span and 175 mm immersion depth.

What is not included in base delivery that I must plan for?

The country-specific plug must be specified at order, and the cooling source (water cooling or refrigerator) is a separate selection needed for operation below 28 °C. External Pt100 sensor, hoses/adapters for the M16x1 connection and the UL 489 protective device are also site-provided; warranty registration is required.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-300BX Immersion Circulator — Technical Data Sheet
Datasheet ·69,9 КБ ·2 pp.
View PDF