Peter Huber Kältemaschinenbau CC-300BX Brückenthermostat
Peter Huber Kältemaschinenbau CC-300BX Brückenthermostat — Thermostatic baths.
The CC-300BX is a bridge-mounted immersion circulator (Brückenthermostat) with a variable-speed pressure/suction pump and Pilot ONE controller, spanning 28–300 °C on its own, 20–300 °C with water cooling and -20–300 °C with a refrigerator. Its 3–3.5 kW heater and ±0.02 K stability distinguish it from the CC-200BX in the series. It serves external closed-loop temperature control of baths and connected equipment.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Brueckenthermostate Series — Model Comparison & Selection
| Model | Current consumption, A | Dimensions w d h | Max delivery flow rate, liter/minute | Temperature range, °C | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| CC-300BX Brückenthermostat Current | 16 | 345 x 190 x 392 | 25 | 28–300 | 13 | 5–40 |
| CC-200BX Brückenthermostat | 10 | 345 x 200 x 326 | 22 | 28–200 | 7 | 5–40 |
Full specifications
| Dimensions (w × D × H) | 345 x 190 x 392 mm |
| Maximum delivery flow rate | 25 l/min (pressure pump; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 28...300 °C with cooling unit: -20...300 °C · with water cooling: 20...300 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 3 - 3, 5 kW |
| Delivery capacity suction pump max. mfr | 18, 5 l/min (suction pump; 0, 4 bar) |
| Weight | 13 kg |
| Current consumption | 16 A |
| Fuse rating | 20 A |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Immersion circulator with telescopic bridge arm extending to 884 mm and 175 mm immersion depth; mounts across a bath opening rather than as a standalone bath.
- Pump connection M16x1 male; pressure and suction pump, variable speed, with pressure side up to 25 l/min at 0.7 bar and suction side up to 18.5 l/min at 0.4 bar.
- Pilot ONE controller with Ethernet, USB host/device and RS232; external Pt100 sensor input.
- Within the Brückenthermostate series the CC-300BX is the higher-capacity modification relative to the CC-200BX; the 3–3.5 kW heating capacity and 16 A current draw are the practical differentiators.
Thermal behavior and media
- The achievable range is set by the cooling method: 28–300 °C without auxiliary cooling, 20–300 °C with water cooling, -20–300 °C with a refrigerator. The cooling source is therefore a selection decision, not an accessory.
- Temperature stability is stated as ±0.02 K.
- Thermofluid kinematic viscosity must not exceed 50 mm²/s; this constrains fluid choice at the low end of the range where viscosity rises.
- Adjustable overtemperature protection per DIN 12876, safety class III/FL.
Installation, control and operation
- Supply 200–240 V, 1~ or 2~, 50/60 Hz, max. current 16 A, internal fuse 20 A; voltage and frequency each ±10% but not deviating in opposite directions.
- External branch circuit protection to UL 489 is required. Single/two-phase units ship with power cable and country-specific plug that must be specified at order.
- IP20 housing, Class A Group 1 to EN55011; ambient 5–40 °C.
- 4-year warranty subject to registration.
Integration & Operating Notes
The datasheet makes external branch circuit protection to UL 489 mandatory. Size it for the 16 A maximum current and the 20 A internal fuse, and confirm the site supply is 200–240 V at 50/60 Hz with voltage and frequency tolerances not running in opposite directions.
Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified at order time. Without this, the delivered cable may not mate with the site receptacle and commissioning is delayed.
The unit reaches 28 °C without auxiliary cooling, 20 °C with water cooling and -20 °C with a refrigerator. Select the cooling configuration against the lowest process temperature before specifying; the base unit alone cannot go below 28 °C.
The maximum permissible kinematic viscosity is 50 mm²/s. At the low end of the range viscosity rises, so the fluid must be chosen to stay within this limit at the coldest operating point, otherwise pump delivery and heat transfer degrade.
The telescopic arm extends to 884 mm and immersion depth is 175 mm. The bath opening and wall geometry must accommodate the bridge span and place the immersion section at the correct depth; the M16x1 male pump connection then dictates the mating hose or adapter.
The housing is IP20 and ambient must stay between 5 and 40 °C. Washdown areas, wet rooms and dusty environments are excluded; site planning must verify the room conditions before the unit is placed.
Overtemperature protection per DIN 12876 is adjustable and the unit carries safety class III/FL. The cut-out must be configured for the bath and medium in use so that a control fault cannot overheat the process.
The 4-year warranty is conditional on registration. Complete registration at commissioning so that warranty coverage is preserved for the equipment.
Selection & Integration Q&A
Pick the CC-300BX when the process needs the higher 3–3.5 kW heating capacity and the 16 A supply it implies, or the full -20–300 °C span with a refrigerator. The CC-200BX is the lower-capacity sibling; confirm its ratings against your load before downsizing.
A 200–240 V, 1~ or 2~, 50/60 Hz supply rated for 16 A with a UL 489 branch protective device, plus the country-specific plug specified at order. Hydraulically, M16x1 male pump connections require matching hoses or adapters, and the external loop must fit within 25 l/min at 0.7 bar pressure and 18.5 l/min at 0.4 bar suction.
Ambient must remain 5–40 °C, the housing is IP20 so water and dust ingress must be excluded, and the unit is Class A Group 1 to EN55011, intended for non-domestic environments. The telescopic bridge needs a bath opening within its 884 mm span and 175 mm immersion depth.
The country-specific plug must be specified at order, and the cooling source (water cooling or refrigerator) is a separate selection needed for operation below 28 °C. External Pt100 sensor, hoses/adapters for the M16x1 connection and the UL 489 protective device are also site-provided; warranty registration is required.