Peter Huber Kältemaschinenbau CC-308B Wärme-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-308B Wärme-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-308B Wärme-Badumwälzthermostat — Thermostatic baths.

The CC-308B is a 7.6-litre stainless-steel bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated to 300 °C. Its operating range depends on the cooling method: 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator. It serves as a temperature-control source for external circuits and immersed samples in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 300C Series — Model Comparison & Selection

Model Current consumption, ADimensions w d hFuse rating, AWeight, kgAmbient temperature limit, °CController
CC-308B Wärme-Badumwälzthermostat Current 14242 x 404 x 39216185–40Pilot ONE
CC-304B Wärme-Badumwälzthermostat 14210 x 335 x 39216145–40Pilot ONE
CC-315B Wärme-Badumwälzthermostat 16335 x 382 x 43320225–40Pilot ONE

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 242 x 404 x 392 mm
Maximum delivery flow rate 25 l/min (pressure pump; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 28...300 °C
with cooling unit: -20...300 °C · with water cooling: 20...300 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 2, 2 - 3 kW
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump; 0, 4 bar)
Bath opening WxD/BD mfr 130 x 110 / 155 mm
Bath volume mfr 7, 6 l
Filling volume min. mfr 5, 8 l (minimum)
Dimensions, Weight & Mechanical Design
Weight 18 kg
Electrical, Power & Drive Specifications
Current consumption 14 A (max.)
Fuse rating 16 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behaviour across media and modes

Installation, control and operation

Integration & Operating Notes

Electrical
Branch circuit must include UL 489 protection and a 16 A fuse

The datasheet requires external branch circuit protection according to UL 489 and specifies a 16 A fuse with a maximum current of 14 A. The supply must be 200–240 V, 1~ or 2~, 50/60 Hz; voltage and frequency tolerances of ±10% apply only if they do not act in opposite directions.

CC-308B Technical Data Sheet, Specifications and installation notes
Electrical
Country-specific plug must be ordered with the unit

Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified at ordering. Without this specification the unit may arrive without a usable mains connection, delaying commissioning.

CC-308B Technical Data Sheet, installation notes
Process Limits
Cooling method sets the lower temperature limit

The operating range is 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator. The required process temperature therefore dictates whether water cooling or a refrigerator must be provided; the bath cannot go below 28 °C unaided.

CC-308B Technical Data Sheet, Specifications
Process Limits
Bath fluid viscosity must not exceed 50 mm²/s

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s. At low temperatures viscosity rises, so the fluid must be selected to stay within this limit across the whole operating range, otherwise pump delivery and heat transfer degrade.

CC-308B Technical Data Sheet, Specifications
Process Limits
Minimum filling volume is 5.8 l of the 7.6 l bath

The bath volume is 7.6 l and the minimum filling capacity is 5.8 l; with a displacement rack the usable bath capacity drops to 5.2 l. The fluid charge must be set accordingly, and sample or displacement inserts reduce the available volume.

CC-308B Technical Data Sheet, Specifications
Mounting
Install only in a dry location between 5 °C and 40 °C ambient

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature. IP20 offers no water protection, so the installation site must be free of splashing water and excessive dust, and the ambient limits must be respected for reliable operation.

CC-308B Technical Data Sheet, installation notes
Process Limits
Set adjustable overtemperature cut-out to suit the bath fluid

The unit has adjustable overtemperature protection according to DIN 12876 with safety classification III/FL. The cut-out must be set for the fluid and process so that it trips before the fluid's upper temperature limit is reached, protecting both the sample and the equipment.

CC-308B Technical Data Sheet, Specifications
Process Limits
External circuit must suit M16x1 male pump connections

The pump connections are M16x1 male threads. Hoses and adapters for the external circuit must match this thread; the pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, so the external loop pressure drop must be checked against these figures.

CC-308B Technical Data Sheet, Specifications

Selection & Integration Q&A

When should I choose the CC-308B rather than the CC-304B or CC-315B?

The CC-308B has a 7.6-litre bath (5.8 l minimum fill). Choose it when your sample or external loop needs that volume; the CC-304B and CC-315B are the smaller and larger siblings in the same series. All share the 300 °C upper limit and Pilot ONE control.

What external cooling and electrical supply must be provided for the CC-308B?

Without cooling the bath is limited to 28–300 °C; water cooling extends it to 20–300 °C and a refrigerator to -20–300 °C. The supply is 200–240 V, 1~ or 2~, 50/60 Hz, max. 14 A, with a 16 A fuse and UL 489 branch protection.

What are the ambient and installation limits for the CC-308B?

The unit is IP20 and must be operated between 5 °C and 40 °C ambient temperature. It offers no water protection, so it must be sited away from splashing water and dust. Voltage and frequency tolerances of ±10% apply only if they do not act in opposite directions.

What is not included in the base delivery and must be ordered or provided separately?

The country-specific plug must be specified at ordering; the power cable is supplied but the plug type is not automatic. External branch circuit protection per UL 489, bath fluid, and hoses/adapters for the M16x1 male pump connections are not part of the base unit.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-308B Technical Data Sheet
Datasheet ·87,8 КБ ·2 pp.