Peter Huber Kältemaschinenbau CC-308B Wärme-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-308B Wärme-Badumwälzthermostat — Thermostatic baths.
The CC-308B is a 7.6-litre stainless-steel bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated to 300 °C. Its operating range depends on the cooling method: 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator. It serves as a temperature-control source for external circuits and immersed samples in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 300C Series — Model Comparison & Selection
| Model | Current consumption, A | Dimensions w d h | Fuse rating, A | Weight, kg | Ambient temperature limit, °C | Controller |
|---|---|---|---|---|---|---|
| CC-308B Wärme-Badumwälzthermostat Current | 14 | 242 x 404 x 392 | 16 | 18 | 5–40 | Pilot ONE |
| CC-304B Wärme-Badumwälzthermostat | 14 | 210 x 335 x 392 | 16 | 14 | 5–40 | Pilot ONE |
| CC-315B Wärme-Badumwälzthermostat | 16 | 335 x 382 x 433 | 20 | 22 | 5–40 | Pilot ONE |
Full specifications
| Dimensions (w × D × H) | 242 x 404 x 392 mm |
| Maximum delivery flow rate | 25 l/min (pressure pump; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 28...300 °C with cooling unit: -20...300 °C · with water cooling: 20...300 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 2, 2 - 3 kW |
| Delivery capacity suction pump max. mfr | 18, 5 l/min (suction pump; 0, 4 bar) |
| Bath opening WxD/BD mfr | 130 x 110 / 155 mm |
| Bath volume mfr | 7, 6 l |
| Filling volume min. mfr | 5, 8 l (minimum) |
| Weight | 18 kg |
| Current consumption | 14 A (max.) |
| Fuse rating | 16 A |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The CC-308B is a warm bath circulator with a 7.6-litre stainless-steel bath (minimum filling 5.8 l; 5.2 l with displacement rack) and a pressure/suction pump with M16x1 male connections.
- Within the CC-30xB series, the CC-304B and CC-315B are the neighbouring sizes; the CC-308B sits between them in bath volume. The 308 code denotes the 7.6-litre bath, not an electrical rating.
- Heating capacity is 2.2–3 kW depending on supply voltage (3 kW at 230–240 V, 2.2 kW at 200 V). Temperature stability is stated as ±0.02 K.
Behaviour across media and modes
- The achievable temperature range is set by the cooling configuration: 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator. The unit is therefore a warm bath by default and requires an external cooling source for sub-ambient work.
- The pump is a variable-speed pressure and suction type: maximum delivery is 25 l/min at 0.7 bar head on the pressure side and 18.5 l/min at 0.4 bar on the suction side. Pump speed is adjustable, allowing the external circuit pressure drop to be matched.
- Bath fluid must have a kinematic viscosity no higher than 50 mm²/s; this limits fluid choice at low temperatures where viscosity rises.
Installation, control and operation
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and accepts an external Pt100 sensor.
- Electrical supply is 200–240 V, 1~ or 2~, 50/60 Hz, with maximum current 14 A and a 16 A fuse. External branch circuit protection according to UL 489 is required. Voltage and frequency tolerances are ±10% only if they do not act in opposite directions.
- Single- and two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering.
- The unit is IP20 and must be operated between 5 °C and 40 °C ambient. Adjustable overtemperature protection per DIN 12876 is built in, with safety classification III/FL.
Integration & Operating Notes
The datasheet requires external branch circuit protection according to UL 489 and specifies a 16 A fuse with a maximum current of 14 A. The supply must be 200–240 V, 1~ or 2~, 50/60 Hz; voltage and frequency tolerances of ±10% apply only if they do not act in opposite directions.
Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified at ordering. Without this specification the unit may arrive without a usable mains connection, delaying commissioning.
The operating range is 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator. The required process temperature therefore dictates whether water cooling or a refrigerator must be provided; the bath cannot go below 28 °C unaided.
The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s. At low temperatures viscosity rises, so the fluid must be selected to stay within this limit across the whole operating range, otherwise pump delivery and heat transfer degrade.
The bath volume is 7.6 l and the minimum filling capacity is 5.8 l; with a displacement rack the usable bath capacity drops to 5.2 l. The fluid charge must be set accordingly, and sample or displacement inserts reduce the available volume.
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature. IP20 offers no water protection, so the installation site must be free of splashing water and excessive dust, and the ambient limits must be respected for reliable operation.
The unit has adjustable overtemperature protection according to DIN 12876 with safety classification III/FL. The cut-out must be set for the fluid and process so that it trips before the fluid's upper temperature limit is reached, protecting both the sample and the equipment.
The pump connections are M16x1 male threads. Hoses and adapters for the external circuit must match this thread; the pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, so the external loop pressure drop must be checked against these figures.
Selection & Integration Q&A
The CC-308B has a 7.6-litre bath (5.8 l minimum fill). Choose it when your sample or external loop needs that volume; the CC-304B and CC-315B are the smaller and larger siblings in the same series. All share the 300 °C upper limit and Pilot ONE control.
Without cooling the bath is limited to 28–300 °C; water cooling extends it to 20–300 °C and a refrigerator to -20–300 °C. The supply is 200–240 V, 1~ or 2~, 50/60 Hz, max. 14 A, with a 16 A fuse and UL 489 branch protection.
The unit is IP20 and must be operated between 5 °C and 40 °C ambient temperature. It offers no water protection, so it must be sited away from splashing water and dust. Voltage and frequency tolerances of ±10% apply only if they do not act in opposite directions.
The country-specific plug must be specified at ordering; the power cable is supplied but the plug type is not automatic. External branch circuit protection per UL 489, bath fluid, and hoses/adapters for the M16x1 male pump connections are not part of the base unit.