Peter Huber Kältemaschinenbau CC-315B Wärme-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-315B Wärme-Badumwälzthermostat — Thermostatic baths.
The CC-315B is a 15.6-litre stainless-steel bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for up to 300 °C. Its base range is 28–300 °C, extended to 20–300 °C with water cooling or −20–300 °C with a refrigerator. It serves as a precise (±0.02 K) thermal source for external circuits and bath immersion.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 300C Series — Model Comparison & Selection
| Model | Current consumption, A | Dimensions w d h | Fuse rating, A | Weight, kg | Ambient temperature limit, °C | Controller |
|---|---|---|---|---|---|---|
| CC-315B Wärme-Badumwälzthermostat Current | 16 | 335 x 382 x 433 | 20 | 22 | 5–40 | Pilot ONE |
| CC-304B Wärme-Badumwälzthermostat | 14 | 210 x 335 x 392 | 16 | 14 | 5–40 | Pilot ONE |
| CC-308B Wärme-Badumwälzthermostat | 14 | 242 x 404 x 392 | 16 | 18 | 5–40 | Pilot ONE |
Full specifications
| Dimensions (w × D × H) | 335 x 382 x 433 mm |
| Maximum delivery flow rate | 25 l/min (pressure pump; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 28...300 °C with cooling unit: -20...300 °C · with water cooling: 20...300 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 3 - 3, 5 kW |
| Delivery capacity suction pump max. mfr | 18, 5 l/min (suction pump; 0, 4 bar) |
| Bath opening WxD/BD mfr | 270 x 145 / 200 (WxD/depth) |
| Bath volume mfr | 15, 6 l |
| Filling volume min. mfr | 11, 5 l (minimum) |
| Weight | 22 kg |
| Current consumption | 16 A (max.) |
| Fuse rating | 20 A |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-315B is the 15.6-litre member of the CC-B series of stainless-steel bath circulators rated to 300 °C. It uses a pressure-and-suction pump with variable speed, a Pilot ONE controller, and a bath opening of 270 × 145 mm (200 mm deep). Compared with the smaller CC-304B and CC-308B siblings, it offers the largest bath volume and the highest heating capacity (3–3.5 kW), making it suited to larger sample loads or higher heat-demand external loops. The wetted circuit is stainless steel; the unit is IP20 and weighs 22 kg.
Thermal behaviour and media
Temperature stability is ±0.02 K. The achievable range is configuration-dependent: 28–300 °C as standard, 20–300 °C with water cooling, and −20–300 °C with an integrated refrigerator. The bath fluid must have a kinematic viscosity no higher than 50 mm²/s. Minimum filling is 11.5 l; with a displacement rack the usable bath capacity drops to 8.5 l. The pressure pump delivers up to 25 l/min at 0.7 bar; the suction pump up to 18.5 l/min at 0.4 bar. Adjustable overtemperature protection per DIN 12876 is provided.
Installation, control and operation
The unit accepts 200–240 V single- or two-phase 50/60 Hz supply, drawing up to 16 A with a 20 A fuse; external branch protection to UL 489 is required. Voltage and frequency tolerances are ±10% and must not oppose each other. Pump ports are M16×1 male. Control and data interfaces include Ethernet, USB host and device, RS232, and a Pt100 external sensor input. Ambient temperature must remain 5–40 °C in a dry indoor location.
Integration & Operating Notes
The 200–240 V 1~/2~ supply draws up to 16 A and requires a 20 A fuse plus external UL 489 branch protection. Undersized wiring or breakers will trip under full heating load, and the ±10% voltage/frequency tolerance must not run in opposite directions.
Without auxiliary cooling the unit operates 28–300 °C; water cooling extends the low end to 20 °C and a refrigerator to −20 °C. The cooling option must be chosen at selection because the base unit cannot reach sub-28 °C temperatures on its own.
Higher-viscosity fluids overload the pressure/suction pump and reduce circulation. Select a thermofluid rated for the full 300 °C top temperature and verify its kinematic viscosity at the lowest operating temperature stays within 50 mm²/s.
Minimum filling is 11.5 l; the bath holds 15.6 l. A displacement rack reduces usable capacity to 8.5 l, so sample volume and expansion space must be accounted for to avoid starving the pump or overflowing at 300 °C.
The DIN 12876 overtemperature protection setpoint must be set above the process temperature but below the fluid's limit. It acts as an independent safety cut-out if the Pilot ONE control or sensor fails, protecting the bath and external circuit.
Selection & Integration Q&A
Pick the CC-315B when you need the largest bath volume (15.6 l) and highest heating capacity (3–3.5 kW) of the series, e.g. for larger sample racks or high-heat external loops. The smaller CC-304B/CC-308B suit lower-volume or lower-heat-demand work.
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump 18.5 l/min at 0.4 bar. Ports are M16×1 male, so hoses and adapters must match that thread; check the external loop's pressure drop against these heads.
A country-specific plug is supplied only if specified at ordering; three-phase units under 63 A ship with cable but no plug. External UL 489 branch protection, bath fluid, hoses/adapters for M16×1 ports, and any cooling-water supply or refrigerator option are not part of the base unit.