Peter Huber Kältemaschinenbau DC31 — Immersion coolers.
The DC31 is an air-cooled, natural-refrigerant immersion cooler with no pump, rated for -30 to 50 °C and 0.4 kW cooling at 15 °C. It is the mid-size unit of the DC30/DC31/DC32 series, distinguished by its capacity and M16x1 male fluid connection. It serves direct bath cooling in laboratory and process circuits.
Peter Huber Kältemaschinenbau Durchflusskuehler Series — Model Comparison & Selection
Full specifications
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Temperature range | -30...50 °C |
| Dimensions (W x D x H) mfr | 250 x 310 x 415 mm |
| Refrigeration machine mfr | air-cooled, natural refrigerant |
| Pump characteristics mfr | no pump |
| Weight | 23 kg |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 2, 7 A (max.) |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 06 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 15:0, 4; 0:0, 35; -20:0, 1 kW (at °C bath temperature) |
| Safety class | I / NFL |
Description
Construction and series positioning
The DC31 is a refrigerating immersion cooler using an air-cooled condenser and R-290 (A3, H220) as the first-stage refrigerant. It has no circulation pump; the fluid connection is an M16x1 male thread, so the unit is placed in or connected to an existing bath or circuit rather than driving flow itself. Within the DC30/DC31/DC32 series it occupies the middle position; the DC30 and DC32 are the neighbouring variants, and selection between them should follow the required cooling capacity at the actual working temperature rather than nominal figures.
Performance behaviour
Cooling capacity is strongly temperature-dependent: 0.4 kW at 15 °C bath temperature, 0.35 kW at 0 °C and 0.1 kW at -20 °C. The published output is valid at 20 °C room temperature and may drop as ambient rises, so sizing must use the capacity at the lowest expected process temperature and with an ambient margin. The operating temperature range is -30 to 50 °C; the process must remain inside this window.
Installation and operation
- Electrical: 220-240 V, single/two-phase, 50/60 Hz, max. current 2.7 A, fuse 10-16 A. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
- Enclosure: IP20, EMC Class A Group 1, intended for non-residential environments; install in a dry indoor location.
- Ambient: 5 to 40 °C.
- Fluid connection: M16x1 male; mating hoses and fittings must match this thread.
- Delivery: single/two-phase devices come with a power cable and country-specific plug, which must be specified at ordering.
- Space requirements around the air-cooled condenser must be respected per the operating conditions.
Integration & Operating Notes
Cooling capacity falls from 0.4 kW at 15 °C to 0.1 kW at -20 °C, and the published data are valid at 20 °C room temperature. A unit sized on the 15 °C figure will be undersized for low-temperature duty or for operation in a warm room.
The condenser is air-cooled, so space requirements around the unit must be respected per the operating conditions; insufficient clearance impairs heat rejection and service access. The IP20 enclosure requires a dry indoor site protected from water and dust, with ambient held between 5 and 40 °C.
The first stage uses R-290 (A3, H220), a flammable natural refrigerant. Installation must follow flammable-refrigerant requirements for ventilation and ignition-source control, and the site safety classification must account for the refrigerant charge of 0.06 kg.
Single/two-phase devices are delivered with a power cable and a country-specific plug that must be stated when ordering. Omitting this detail delays commissioning because the plug type cannot be assumed from the model code.
The DC31 has no circulation pump and pump speed is not adjustable. Fluid movement must be provided by the bath or an external circuit, and the M16x1 male connection must be matched with compatible hoses and fittings.
Selection & Integration Q&A
Choose the DC31 when the required cooling capacity at your lowest process temperature falls within its 0.4 kW at 15 °C to 0.1 kW at -20 °C curve. Compare the DC30 and DC32 capacity curves at the same working temperature; pick the smallest unit that covers the duty with ambient margin.
The DC31 has no pump and no speed control, so circulation must come from the bath or an external circuit. The fluid connection is M16x1 male; mating hoses and fittings must match this thread, and the process temperature must stay within -30 to 50 °C.
Install indoors in a dry location: IP20 enclosure, ambient 5-40 °C, EMC Class A Group 1 for non-residential environments. Respect condenser clearance per the operating conditions. Supply must be 220-240 V 1~/2~ 50/60 Hz with 10-16 A fuse and voltage/frequency within the allowed combined envelope.
Delivery includes a power cable with a country-specific plug that must be specified at ordering. Hoses and fittings for the M16x1 male connection are not part of the base unit and must be sourced to match. No pump is fitted, so any circulation hardware is external.