Peter Huber Kältemaschinenbau DC31
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau DC31

Peter Huber Kältemaschinenbau DC31 — Immersion coolers.

The DC31 is an air-cooled, natural-refrigerant immersion cooler with no pump, rated for -30 to 50 °C and 0.4 kW cooling at 15 °C. It is the mid-size unit of the DC30/DC31/DC32 series, distinguished by its capacity and M16x1 male fluid connection. It serves direct bath cooling in laboratory and process circuits.

Peter Huber Kältemaschinenbau Durchflusskuehler Series — Model Comparison & Selection

Model Weight, kgAmbient temperature limit, °CPower supplyProtection ratingPump connectionSafety class
DC31 Current 235–40220-240V 1~/2~ 50/60HzIP20M16x1 AG maleI / NFL
DC30 165–40220-240V 1~/2~ 50/60HzIP20M16x1 AG maleI / NFL
DC32 ——————

Full specifications

Core Performance & Functional Specifications
Pump connection M16x1 AG male
Pump speed adjustable no
Temperature range -30...50 °C
Dimensions (W x D x H) mfr 250 x 310 x 415 mm
Refrigeration machine mfr air-cooled, natural refrigerant
Pump characteristics mfr no pump
Dimensions, Weight & Mechanical Design
Weight 23 kg
Electrical, Power & Drive Specifications
Power supply 220-240V 1~/2~ 50/60Hz
Current consumption max. mfr 2, 7 A (max.)
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 1st stage mfr 0, 06 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 15:0, 4; 0:0, 35; -20:0, 1 kW (at °C bath temperature)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The DC31 is a refrigerating immersion cooler using an air-cooled condenser and R-290 (A3, H220) as the first-stage refrigerant. It has no circulation pump; the fluid connection is an M16x1 male thread, so the unit is placed in or connected to an existing bath or circuit rather than driving flow itself. Within the DC30/DC31/DC32 series it occupies the middle position; the DC30 and DC32 are the neighbouring variants, and selection between them should follow the required cooling capacity at the actual working temperature rather than nominal figures.

Performance behaviour

Cooling capacity is strongly temperature-dependent: 0.4 kW at 15 °C bath temperature, 0.35 kW at 0 °C and 0.1 kW at -20 °C. The published output is valid at 20 °C room temperature and may drop as ambient rises, so sizing must use the capacity at the lowest expected process temperature and with an ambient margin. The operating temperature range is -30 to 50 °C; the process must remain inside this window.

Installation and operation

Integration & Operating Notes

Process Limits
Size capacity at the lowest process temperature, not at 15 °C

Cooling capacity falls from 0.4 kW at 15 °C to 0.1 kW at -20 °C, and the published data are valid at 20 °C room temperature. A unit sized on the 15 °C figure will be undersized for low-temperature duty or for operation in a warm room.

DC31 Datasheet, Specifications
Mounting
Plan condenser clearance and a dry indoor location

The condenser is air-cooled, so space requirements around the unit must be respected per the operating conditions; insufficient clearance impairs heat rejection and service access. The IP20 enclosure requires a dry indoor site protected from water and dust, with ambient held between 5 and 40 °C.

DC31 Datasheet, Specifications
Process Limits
R-290 is flammable; respect refrigerant safety requirements

The first stage uses R-290 (A3, H220), a flammable natural refrigerant. Installation must follow flammable-refrigerant requirements for ventilation and ignition-source control, and the site safety classification must account for the refrigerant charge of 0.06 kg.

DC31 Datasheet, Specifications
Electrical
Specify the country-specific plug at ordering

Single/two-phase devices are delivered with a power cable and a country-specific plug that must be stated when ordering. Omitting this detail delays commissioning because the plug type cannot be assumed from the model code.

DC31 Datasheet, Specifications
Control & I/O
No pump and no speed control; flow must come from elsewhere

The DC31 has no circulation pump and pump speed is not adjustable. Fluid movement must be provided by the bath or an external circuit, and the M16x1 male connection must be matched with compatible hoses and fittings.

DC31 Datasheet, Specifications

Selection & Integration Q&A

When should I choose the DC31 rather than the DC30 or DC32?

Choose the DC31 when the required cooling capacity at your lowest process temperature falls within its 0.4 kW at 15 °C to 0.1 kW at -20 °C curve. Compare the DC30 and DC32 capacity curves at the same working temperature; pick the smallest unit that covers the duty with ambient margin.

What must I provide on the fluid side, given the DC31 has no pump?

The DC31 has no pump and no speed control, so circulation must come from the bath or an external circuit. The fluid connection is M16x1 male; mating hoses and fittings must match this thread, and the process temperature must stay within -30 to 50 °C.

What site and duty limits apply to installing the DC31?

Install indoors in a dry location: IP20 enclosure, ambient 5-40 °C, EMC Class A Group 1 for non-residential environments. Respect condenser clearance per the operating conditions. Supply must be 220-240 V 1~/2~ 50/60 Hz with 10-16 A fuse and voltage/frequency within the allowed combined envelope.

What is included in delivery and what must be ordered or planned separately?

Delivery includes a power cable with a country-specific plug that must be specified at ordering. Hoses and fittings for the M16x1 male connection are not part of the base unit and must be sourced to match. No pump is fitted, so any circulation hardware is external.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
DC31 Datasheet
Datasheet ·27,9 КБ ·1 pp.
View PDF