Peter Huber Kältemaschinenbau DC32 — Recirculating chillers.

The DC32 is a recirculating chiller with a rated operating range of -30 to 50 °C and no integrated pump, so it functions as a cooling stage in an externally pumped loop. Its cooling capacity falls from 0.6 kW at 15 °C to 0.12 kW at -20 °C, making it suitable only for low-load process cooling within that envelope. It uses R-290 refrigerant and an ON-OFF controller.

Peter Huber Kältemaschinenbau Durchflusskuehler Series — Model Comparison & Selection

Model Ambient temperature limit, °CControllerCooling capacity, WCurrent consumption, APower supplyProtection rating
DC32 Current 5–40ON-OFF-Regler1203220-240V 1~/2~ 50/60HzIP20
DC30 ——————
DC31 ——————

Full specifications

Core Performance & Functional Specifications
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type no pump
Temperature range -30...50 °C
Dimensions (W x D x H) mfr 280 x 340 x 465 mm
Dimensions, Weight & Mechanical Design
Weight 30 kg
Electrical, Power & Drive Specifications
Current consumption 3 A (max.)
Power supply 220-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller ON-OFF-controller
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 0.08 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling capacity 0.6 kW (15 °C)
0 °C: 0.47 kW · -20 °C: 0.12 kW
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and positioning

Behavior under load

Installation and operation

Integration & Operating Notes

Process Limits
Cooling capacity at -20 °C is only 0.12 kW

The refrigeration circuit delivers 0.6 kW at 15 °C but only 0.12 kW at -20 °C. Selecting the DC32 for a load that requires more than 0.12 kW at -20 °C will result in insufficient cooling and possible process deviation. Always size using the capacity at the lowest intended operating temperature.

DC32 Datasheet, Specifications
Electrical
Upstream fuse must be between 10 A and 16 A

The datasheet specifies a minimum fuse of 10 A and a maximum fuse of 16 A. A lower-rated breaker may trip during compressor start-up, while a higher-rated one may not protect the unit. Verify the supply voltage is 220–240 V and frequency 50/60 Hz within the allowed tolerances.

DC32 Datasheet, Specifications
Mounting
Install only in a dry indoor room between 5 and 40 °C

The IP20 enclosure is not protected against water or dust ingress, so the installation site must be dry and sheltered. Ambient temperature must remain between 5 and 40 °C; outside this range reliable operation is not guaranteed and cooling capacity may drop further.

DC32 Datasheet, Specifications
Process Limits
Fluid connection is M16x1 male; no pump is included

The DC32 has no integrated pump and terminates in an M16x1 male thread. The customer must provide an external pump, hoses, and fittings matching this thread. Pump speed cannot be adjusted from the chiller because no pump control is present.

DC32 Datasheet, Specifications
Electrical
Country-specific plug must be ordered explicitly

Single/two-phase devices are delivered with a power cable and a country-specific plug. The plug variant must be specified at the time of ordering; otherwise the unit may arrive with a plug unsuitable for the site.

DC32 Datasheet, Installation
Cooling & Media
R-290 refrigerant requires flammable-gas awareness

The refrigerant is R-290 (A3, H220), classified as flammable. Installation and service personnel must be aware of the safety measures for A3 refrigerants, including avoiding ignition sources and ensuring adequate ventilation during service.

DC32 Datasheet, Safety

Selection & Integration Q&A

When should I choose the DC32 instead of the DC30 or DC31?

Choose the DC32 when the loop already has an external pump and you need cooling down to -30 °C with up to 0.6 kW at 15 °C. The DC30 and DC31 are siblings in the same series; select them if their published configurations better match your required temperature range or pump arrangement.

What external pump and fittings are required to operate the DC32?

The DC32 has no pump and provides an M16x1 male fluid connection. You must supply an external pump and hoses or fittings with a matching M16x1 female thread. Pump speed cannot be controlled by the chiller, so the external pump must be set independently.

What is included in the base delivery and what must be ordered separately?

The base delivery includes the chiller with a power cable and a country-specific plug that must be specified at ordering. No pump, hoses, or fittings are included. The R-290 refrigerant charge is factory-filled; service consumables are not part of the scope.