Peter Huber Kältemaschinenbau HB720 Umwälz-Heizgerät
Peter Huber Kältemaschinenbau HB720 Umwälz-Heizgerät — Thermostatic baths.
The HB720 is a 72 kW three-phase immersion heating circulator for external open operation, delivering 60–250 °C with a magnetically coupled stainless-steel pressure pump (200 l/min, 5.5 bar). It has no expansion tank and no active refrigeration, so minimum temperature depends on ambient and application heat loss. It serves high-capacity external heating loops in industrial processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Umwaelz Heizgeraete Series — Model Comparison & Selection
| Model | Dimensions w d h | Power supply | Pump connection | Pump type | Heating capacity, W | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| HB720 Umwälz-Heizgerät Current | 800 x 1007 x 1607 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 72000 | 5–40 |
| HB120 Umwälz-Heizgerät | 479 x 539 x 695 | 380-460V 3~ 50/60Hz | M30x1, 5 AG male | magnetically coupled, st | — | 5–40 |
| HB240 Umwälz-Heizgerät | 570 x 700 x 1171 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 24000 | 5–40 |
| HB480 Umwälz-Heizgerät | 570 x 700 x 1171 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 48000 | 5–40 |
| HB60 Umwälz-Heizgerät | 323 x 451 x 498 | 400V 3~N 50Hz | M30x1, 5 AG male | Pressure pump, speed not | 6000 | 5–40 |
Full specifications
| Dimensions (w × D × H) | 800 x 1007 x 1607 mm |
| Heating capacity | 72 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 60...250 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | II / FL |
Description
Construction and series positioning
- The HB720 is the largest heating-capacity modification in the HB series (siblings HB45, HB120, HB240, HB480), with 72 kW heating power and a 60–250 °C working range.
- It is built exclusively for external open operation and has no expansion tank; the connected application must accommodate fluid volume changes.
- The pump is a variable-speed, magnetically coupled stainless-steel pressure pump with M38x1.5 male connections, rated up to 200 l/min and 5.5 bar.
- Control is via the Pilot ONE controller with 0.01 K display resolution; interfaces include Ethernet, USB (host and device) and RS232, plus a Pt100 external sensor connection.
- Electrical supply is 400 V 3~ 50 Hz; protection rating IP20; safety class II / FL with adjustable overtemperature protection per DIN 12876.
Behavior under load and media
- Because there is no active refrigeration, the 60 °C minimum temperature is reachable only when ambient is below 60 °C and application heat loss falls between 0.1 kW and 2.8 kW (model-dependent).
- The 72 kW heating power sets heat-up capability and must be matched to the thermal load; the pump curve (200 l/min, 5.5 bar) must overcome the external circuit pressure drop.
- No digitized curve points are available; hydraulic verification must rely on the stated maximum delivery and head.
Installation, connection and operation
- Three-phase units below 63 A are supplied with cable but no plug; units above 63 A are supplied without cable and without plug, requiring full customer-side cabling.
- Voltage and frequency tolerances of ±10% are permitted only if they do not act in opposite directions; supply quality must be checked.
- EMC classification is EN55011 Class A Group 1, intended for industrial environments; interference in residential areas is possible.
- Ambient temperature must remain 5–40 °C; IP20 requires a dry indoor location protected from water and dust.
Integration & Operating Notes
The unit has no active refrigeration, so 60 °C is reachable only if ambient is below 60 °C and the application heat loss lies between 0.1 kW and 2.8 kW depending on model. Size the external circuit and insulation so the heat loss stays in this band, otherwise the minimum temperature cannot be attained.
Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the customer must provide and fit a suitable plug. Devices drawing more than 63 A are supplied without cable and without plug, requiring full customer-side power cabling and planning.
The unit provides adjustable overtemperature protection according to DIN 12876. The trip limit must be configured for the specific external application so that the connected circuit is protected against overheating; this is a safety setting, not a fixed factory value.
The circulation pump delivers up to 200 l/min at up to 5.5 bar. Verify that this head and flow are sufficient for the pressure drop of the external circuit, and match hoses or adapters to the M38x1.5 male pump connection.
Selection & Integration Q&A
Pick the HB720 when the process needs up to 72 kW heating and the 60–250 °C range with a 200 l/min, 5.5 bar pump. Smaller siblings suit lower heat loads; the HB720 is the largest heating-capacity modification in the series.
The application must be an external open circuit that accommodates fluid volume changes, since there is no expansion tank. Hoses or adapters must match the M38x1.5 male pump connection and handle up to 200 l/min at 5.5 bar.
Ambient must stay 5–40 °C, the location must be dry indoor due to IP20, and the 400 V 3~ 50 Hz supply must not combine voltage and frequency deviations in opposite directions beyond ±10%.
Depending on current draw, the unit ships with cable but no plug (below 63 A) or without cable and plug (above 63 A), so the customer must provide the plug and, above 63 A, the full power cabling.