Peter Huber Kältemaschinenbau HB60 Umwälz-Heizgerät
Peter Huber Kältemaschinenbau HB60 Umwälz-Heizgerät — Thermostatic baths.
The HB60 is a 6 kW circulating heating unit for external open operation, delivering up to 90 l/min at 2.5 bar over a 60–250 °C range. It has no expansion tank, so the connected application must absorb thermal volume changes. It serves as the mid-capacity member of the HB series, between the HB45 and HB120.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Umwaelz Heizgeraete Series — Model Comparison & Selection
| Model | Dimensions w d h | Power supply | Pump connection | Pump type | Heating capacity, W | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| HB60 Umwälz-Heizgerät Current | 323 x 451 x 498 | 400V 3~N 50Hz | M30x1, 5 AG male | Pressure pump, speed not | 6000 | 5–40 |
| HB120 Umwälz-Heizgerät | 479 x 539 x 695 | 380-460V 3~ 50/60Hz | M30x1, 5 AG male | magnetically coupled, st | — | 5–40 |
| HB240 Umwälz-Heizgerät | 570 x 700 x 1171 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 24000 | 5–40 |
| HB480 Umwälz-Heizgerät | 570 x 700 x 1171 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 48000 | 5–40 |
| HB720 Umwälz-Heizgerät | 800 x 1007 x 1607 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 72000 | 5–40 |
Full specifications
| Dimensions (w × D × H) | 323 x 451 x 498 mm |
| Heating capacity | 6 kW |
| Maximum delivery flow rate | 90 l/min (pressure pump max.; also 2, 5 bar) |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure pump, speed not variable |
| External sensor connection | Pt100 |
| Temperature range | 60...250 °C |
| Weight | 49 kg (net) |
| Power supply | 400V 3~N 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C (permissible ambient temperature) |
| Protection rating | IP20 |
| Safety class | II / FL |
Description
Construction and series positioning
The HB60 is a pressure-pump circulating heater with a fixed-speed pump (no speed adjustment) and a Pilot ONE controller. Its 6 kW heating capacity and 90 l/min maximum delivery place it between the HB45 and HB120 in the HB series; the larger HB240 and HB480 extend capacity further. The unit is designed exclusively for external open operation and has no expansion tank, so the connected application must accommodate fluid filling and thermal volume changes. Minimum filling capacity is 3.5 l.
Hydraulic and thermal behavior
The pump delivers up to 90 l/min at a maximum pressure of 2.5 bar, with 75 l/min at 0.5 bar. The external circuit pressure drop must be checked against this curve. The temperature range is 60–250 °C with a display resolution of 0.01 K. Heating power is 6 kW, so heat-up time and electrical supply capacity must be planned for this load.
Installation and control
The unit requires a 400 V 3~N 50 Hz supply with a 3x16 A fuse. Three-phase devices below 63 A are supplied with cable but without a plug, so the installer must provide a country-specific plug. Voltage and frequency tolerances are ±10% only if they do not run in opposite directions. Pump connections are M30x1.5 male with a 60° Y arrangement to DIN 3863; pipework and flexible hoses must use a union to DIN 3863. The unit is rated IP20 and must be installed in a dry, protected location. Ambient temperature must remain between 5 °C and 40 °C. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and an external Pt100 sensor connection. Adjustable overtemperature protection according to DIN 12876 must be set for the application. The unit is Class A, Group 1 to EN55011, and leakage current from external temperature control must be considered for protective earthing and RCD selection.
Integration & Operating Notes
The HB60 is designed exclusively for external open operation and has no expansion tank. The connected application must provide volume compensation for fluid filling and thermal expansion, and must hold at least the 3.5 l minimum filling capacity to avoid pump cavitation and loss of thermal buffer.
The unit requires a three-phase 400 V 3~N 50 Hz supply protected by a 3x16 A fuse. Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so supply quality must be checked before commissioning. Leakage current from external temperature control must be considered for protective earthing and RCD selection.
Pump connections are M30x1.5 male in a 60° Y arrangement to DIN 3863. Pipework and flexible tempering hoses must be fitted with a union to DIN 3863 to ensure correct sealing and avoid adapters or leaks at the circulation loop interface.
The unit is rated IP20 and must not be exposed to water or dust ingress. Ambient temperature must remain between 5 °C and 40 °C; operating outside this range can reduce performance, trip protections or damage components. Class A, Group 1 to EN55011 may require additional EMC measures in sensitive environments.
Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug. The installer must provide a suitable country-specific plug and connection, which affects the bill of materials for commissioning.
The fixed-speed pressure pump delivers up to 90 l/min at 2.5 bar, with 75 l/min at 0.5 bar. The external circuit pressure drop must be checked against this curve to ensure the required flow is achieved; the pump speed is not adjustable.
The unit has adjustable overtemperature protection according to DIN 12876. It must be set appropriately for the application before running the process to provide a protective limit and safe operation.
Selection & Integration Q&A
Choose the HB60 when the process needs up to 6 kW heating and up to 90 l/min at 2.5 bar over 60–250 °C. The HB45 offers lower capacity; the HB120 is for higher heat demand. All require external open operation without an expansion tank.
A 400 V 3~N 50 Hz supply with a 3x16 A fuse. The unit is supplied with cable but without a plug, so a country-specific plug must be provided. Voltage and frequency tolerances are ±10% only if they do not run in opposite directions; leakage current must be considered for RCD selection.
The base delivery includes a cable but no plug for the three-phase supply. Hoses or piping must be fitted with unions to DIN 3863 for the M30x1.5 male pump connections. The external application must provide volume compensation and at least 3.5 l filling capacity, since no expansion tank is included.