Shimadzu GPC
Shimadzu ·Made in Japan · In Active Production

Shimadzu GPC

Shimadzu GPC — Chromatographs.

The Shimadzu GPC is a modular gel permeation chromatography system built around the RID-20A refractive index detector, available in integrated Prominence-i or modular Nexera configurations. It delivers retention time reproducibility of 0.025–0.066 %RSD over five days for polystyrene, with flow rate precision ≤0.06 %RSD. The system is used for polymer molecular weight distribution analysis, with LabSolutions GPC software supporting multiple calibration curve approximations.

Direct drop-in replacements

0%
match
⚖️ Compare
0%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Detector types Refractive Index Detector (RID-20A), Photodiode Array Detector (PDA) (Prominence-i GPC System)
Retention time reproducibility mfr 0.066 %RSD (n=10) (Day 1, polystyrene)
Retention time reproducibility mfr 0.034 %RSD (n=10) (Day 2, polystyrene)
Retention time reproducibility mfr 0.065 %RSD (n=10) (Day 3, polystyrene)
Retention time reproducibility mfr 0.025 %RSD (n=10) (Day 4, polystyrene)
Retention time reproducibility mfr 0.057 %RSD (n=10) (Day 5, polystyrene)
Retention time mfr 7.427 min (Day 1, polystyrene)
Retention time mfr 7.444 min (Day 2, polystyrene)
Retention time mfr 7.455 min (Day 3, polystyrene)
Retention time mfr 7.418 min (Day 4, polystyrene)
Retention time mfr 7.434 min (Day 5, polystyrene)
System configuration mfr Modular (Nexera GPC System standard)
Detector cell volume mfr 9 μL (RID-20A detector)
Calibration curve approximation formulae mfr Linear, 3rd-order, 3rd-order + hyperbolic curve, (LabSolutions GPC Software)
Data output method mfr ASCII format (LabSolutions GPC Software)
Solvent usage savings mfr 30 % (RID-20A GPC analysis)
Stabilization time mfr After 30 Minutes (Power ON, RID-20A)
Electrical, Power & Drive Specifications
Mobile phase flow rate accuracy mfr No more than 1% or 2 μL/min, whichever is greate (Nexera GPC System standard)
Mobile phase flow rate precision mfr No more than 0.06% RSD or 0.02 min SD, whichever (Nexera GPC System standard)
Mobile phase containers monitored mfr up to twelve containers (1-liter containers; up to 4 with 2- to 5-liter containers)

Description

Construction and positioning

Performance under load

Installation and operation

Integration & Operating Notes

Process Limits
Mobile phase flow rate must stay within 0.01–2 mL/min

Operating outside this range compromises flow accuracy (specified as ≤1% or 2 µL/min) and may affect retention time reproducibility. Ensure the LC-40D solvent delivery unit is calibrated for the intended flow rate before running samples.

Nexera GPC System standard specifications
Control & I/O
Allow 30 minutes for RID-20A stabilization after power-on

The refractive index detector requires a 30-minute warm-up to reach stable operating conditions. Starting analysis before stabilization can cause baseline drift and invalidate retention time reproducibility. Incorporate this delay into the startup sequence.

RID-20A detector specifications
Maintenance
Monitor mobile phase container levels to prevent pump cavitation

The system can monitor up to twelve 1-liter containers or four 2- to 5-liter containers. Low solvent levels can lead to cavitation and flow interruptions. Set appropriate alerts in LabSolutions and refill containers before they run dry.

LabSolutions GPC Software features

Selection & Integration Q&A

When should I choose the integrated Prominence-i GPC system versus the modular Nexera GPC system?

Choose Prominence-i for a compact, integrated HPLC with PDA and RID-20A, ideal for routine GPC with compositional analysis. Choose Nexera for modular flexibility, higher throughput, and independent component upgrades.

What are the requirements for connecting the RID-20A detector to the LC system?

The RID-20A requires mobile phase flow from the LC-40D pump, with flow rate within 0.01–2 mL/min and precision ≤0.06% RSD. Ensure proper tubing connections and temperature control (30–60 °C) for stable baseline.

What maintenance is required and what consumables are not included?

Regular maintenance includes replacing mobile phase filters, checking for leaks, and calibrating the detector. Consumables like columns (Shim-pack GPC series), vials, and solvents are not included. Refer to the manual for service intervals.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
C190E182
Datasheet ·144,8 КБ ·2 pp.
View PDF
PDF
c190-e092
Datasheet ·2,0 МБ ·12 pp.
View PDF