No photo available yet
Shimadzu MOC-120H
Shimadzu MOC-120H — Moisture meters.
The Shimadzu Moc 120H is an infrared heat drying moisture analyzer using a mid-wave quartz heater and a 120 g capacity balance with 0.001 g resolution. It determines moisture content from 0.01% to 100.00% with 0.01% display, offering automatic or timed halting and rapid, slow, or step drying modes. This model suits benchtop quality control where samples up to 120 g require precise, reproducible loss-on-drying measurements.
Full specifications
| Resolution | Large backlit LCD |
| Drying modes | 9 combinations of drying and halting modes |
| Heat source | Mid-wave infrared quartz heater |
| Model | MOC-120H (321-63300-10) |
| Temperature settings | (Sample position temperature) |
| Stored procedures | 10 |
| Measurement principle (Texture Analyzer EZ-SX) mfr | high-precision load cell (Texture Analyzer EZ-SX) |
| Sample amount (Mayonnaise, Moisture Analyzer) mfr | 1 g (Mayonnaise, Moisture Analyzer) |
| Measurement time (Mayonnaise, Moisture Analyzer) mfr | 10:00 min (Mayonnaise, Moisture Analyzer) |
| Moisture ratio (Mayonnaise, Moisture Analyzer) mfr | 20.61 % (Mayonnaise, Moisture Analyzer) |
| Repeatability (Mayonnaise, Moisture Analyzer) mfr | 0.46 % (Mayonnaise, Moisture Analyzer) |
| Sample amount (Instant coffee, Moisture Analyzer) mfr | 1 g (Instant coffee, Moisture Analyzer) |
| Measurement time (Instant coffee, Moisture Analyzer) mfr | 10:00 min (Instant coffee, Moisture Analyzer) |
| Moisture ratio (Instant coffee, Moisture Analyzer) mfr | 7.43 % (Instant coffee, Moisture Analyzer) |
| Repeatability (Instant coffee, Moisture Analyzer) mfr | 1.18 % (Instant coffee, Moisture Analyzer) |
| Sample amount (Chocolate, Moisture Analyzer) mfr | 3 g (Chocolate, Moisture Analyzer) |
| Measurement time (Chocolate, Moisture Analyzer) mfr | 6:18 min (Chocolate, Moisture Analyzer) |
| Moisture ratio (Chocolate, Moisture Analyzer) mfr | 2.36 % (Chocolate, Moisture Analyzer) |
| Repeatability (Chocolate, Moisture Analyzer) mfr | 1.49 % (Chocolate, Moisture Analyzer) |
| Sample amount (Corn starch, Moisture Analyzer) mfr | 5 g (Corn starch, Moisture Analyzer) |
| Measurement time (Corn starch, Moisture Analyzer) mfr | 7:54 min (Corn starch, Moisture Analyzer) |
| Moisture ratio (Corn starch, Moisture Analyzer) mfr | 12.94 % (Corn starch, Moisture Analyzer) |
| Repeatability (Corn starch, Moisture Analyzer) mfr | 0.16 % (Corn starch, Moisture Analyzer) |
| Sample amount (White rice, Moisture Analyzer) mfr | 5 g (White rice, Moisture Analyzer) |
| Measurement time (White rice, Moisture Analyzer) mfr | 12:30 min (White rice, Moisture Analyzer) |
| Moisture ratio (White rice, Moisture Analyzer) mfr | 15.12 % (White rice, Moisture Analyzer) |
| Heater operational life (mid-wave infrared quartz heater) mfr | 20,000 to 30,000 hours (mid-wave infrared quartz heater) |
| Ending modes mfr | Automatic ending mode; Timed ending mode |
| Alternate drying modes mfr | Rapid drying mode; Slow drying mode |
| Weighing sensor mfr | Shimadzu UniBloc cell |
| Power requirements mfr | AC 100 to 120 / 220 to 240 V |
| Power consumption mfr | 640 W maxi |
| Data interface | RS-232C (WindowsDirect) |
Description
Construction and positioning
- The Moc 120H combines a mid-wave infrared quartz heater with a 0.001 g resolution balance and a ø130 mm sample pan. - Maximum sample capacity is 120 g, allowing larger sample sizes than typical 50–100 g analyzers. ### Measurement behavior
- Moisture content range: 0.01% to 100.00%, with minimum display 0.01%. - Drying temperature settable from 30 °C to 200 °C in 1 °C increments. - Measurement modes include automatic or timed halting, plus rapid, slow, and step drying profiles and a predictive measuring mode. ### Installation and operation
- Operates in ambient 5 °C to 40 °C, up to 85% RH. - Power: 100–127 V or 220–240 V AC, 640 W maximum. - Standard accessories include two sample pans, pan stop, windbreak case, pan handler (2 pcs), 20 aluminum sheets, spoon, and spatula. - No PDF digests were provided, so connection, control, and detailed operating specifics cannot be covered.
Integration & Operating Notes
The unit accepts either 100–127 V AC or 220–240 V AC, but the specific configuration must be confirmed against the nameplate. Connecting a 100–127 V unit to a 220–240 V supply will damage the instrument. Maximum power draw is 640 W, so ensure the circuit can supply this continuously.
Operation outside this range may affect balance stability and moisture results. Condensation at high humidity can corrupt readings; allow the instrument to equilibrate if moved from a cold environment. The 85% RH maximum is a non-condensing limit.
The windbreak case is supplied to shield the sample from air currents. Install the pan stop and windbreak as per the instruction manual. Avoid vibration and direct sunlight, which can affect the 0.001 g resolution balance.
Selection & Integration Q&A
It requires 100–127 V or 220–240 V AC, 640 W max, and operates in 5–40 °C, ≤85% RH. Ensure the supply matches the unit's configuration and that the bench is stable and draft-free.
Mount on a level, vibration-free surface away from drafts. Ambient must be 5–40 °C and ≤85% RH. No duty-cycle limit is specified, but continuous high-temperature operation may require cooling periods; refer to the manual.