Shimadzu MOC63u
Shimadzu ·Made in Japan · In Active Production

Shimadzu MOC63u

Shimadzu MOC63u — Moisture meters.

The MOC63u is a halogen (straight-tube) moisture analyzer with a 60.0 g capacity and 0.001 g / 0.01 % readability, featuring a 95 mm weighing pan and 400 W heat source. It offers standard, rapid, slow, and 3-stage step drying modes with automatic or timed ending, storing up to 10 sets of measurement conditions. Designed for laboratory moisture determination, it provides USB and RS-232C outputs for data logging and printer connectivity.

Full specifications

Core Performance & Functional Specifications
Resolution LCD with backlight
Data memory 100
Drying modes Standard drying mode (Automatic ending/timed ending) Rapid drying mode (Automatic ending/timed ending) Slow drying mode (Automatic ending/timed ending) Step (3-stage) drying mode (Automatic ending/timed ending)
Heat source Method / Power
Temperature settings 50 to 200 °C (1 °C interval) (up to 1 hour for settings over 180 °C) to 200 °C (1 °C interval)
Repeatability 0.15 % (2 g), 0.05 % (5 g), 0.02 % (10 g) % (2 g), 0.05 % (5 g), 0.02 %
Pan size 95 mm
Measured moisture content (105 °C, 5 hrs) mfr 0.27 % (105 °C, 5 hrs)
Measured moisture content (105 °C, 1 hrs) mfr 0.25 % (105 °C, 1 hrs)
Measured moisture content (150 °C, 15 min) mfr 0.27 % (150 °C, 15 min)
Indentation load (PP, dynamic ultra micro hardness) mfr 502.3 mN (PP, dynamic ultra micro hardness)
Hardness (PP, dynamic ultra micro hardness) mfr 103.1 N/mm2 (PP, dynamic ultra micro hardness)
Hardness deviation (PP, dynamic ultra micro hardness) mfr 1.59 (PP, dynamic ultra micro hardness)
Indentation load (HDPE, dynamic ultra micro hardness) mfr 502.2 mN (HDPE, dynamic ultra micro hardness)
Hardness (HDPE, dynamic ultra micro hardness) mfr 045.7 N/mm2 (HDPE, dynamic ultra micro hardness)
Hardness deviation (HDPE, dynamic ultra micro hardness) mfr 0.87 (HDPE, dynamic ultra micro hardness)
Indentation load (ABS/PMMA, dynamic ultra micro hardness) mfr 502.1 mN (ABS/PMMA, dynamic ultra micro hardness)
Hardness (ABS/PMMA, dynamic ultra micro hardness) mfr 193.8 N/mm2 (ABS/PMMA, dynamic ultra micro hardness)
Hardness deviation (ABS/PMMA, dynamic ultra micro hardness) mfr 8.74 (ABS/PMMA, dynamic ultra micro hardness)
Indentation load (PS, dynamic ultra micro hardness) mfr 502.2 mN (PS, dynamic ultra micro hardness)
Hardness (PS, dynamic ultra micro hardness) mfr 227.8 N/mm2 (PS, dynamic ultra micro hardness)
Hardness deviation (PS, dynamic ultra micro hardness) mfr 6.15 (PS, dynamic ultra micro hardness)
Minimum sample quantity (Easy start/Automatic end/Timed end) mfr 0.02 g (Easy start/Automatic end/Timed end)
Drying heater (Easy start/Automatic end/Timed end) mfr Straight type halogen heater (Easy start/Automatic end/Timed end)
External Output mfr -100
Storage of Measurement Conditions mfr 10 sets
Dimensions, Weight & Mechanical Design
Capacity . Sample Quantity Min. Sample Quantity Max
Weight Approx 4.2 kg
Width 202 × 336 × 157 mm
Indentation depth (PP, dynamic ultra micro hardness) mfr 16.4 μm (PP, dynamic ultra micro hardness)
Indentation depth (HDPE, dynamic ultra micro hardness) mfr 23.5 μm (HDPE, dynamic ultra micro hardness)
Indentation depth (ABS/PMMA, dynamic ultra micro hardness) mfr 12.1 μm (ABS/PMMA, dynamic ultra micro hardness)
Indentation depth (PS, dynamic ultra micro hardness) mfr 11.2 μm (PS, dynamic ultra micro hardness)
Electrical, Power & Drive Specifications
Power consumption 430 VA
Interfaces, Connections & Control
Data interface USB port
Standards, Safety & General Data
Number-average molecular weight Mn (PLA resin container, GPC) mfr 81,900 (PLA resin container, GPC)
Weight-average molecular weight Mw (PLA resin container, GPC) mfr 179,600 (PLA resin container, GPC)
Polydispersity Mw/Mn (PLA resin container, GPC) mfr 2.19 (PLA resin container, GPC)
Standard Accessories mfr 3 aluminum

Description

Construction and Positioning

The MOC63u is a benchtop moisture analyzer using a straight-tube halogen lamp as the heat source, rated at 400 W. Its weighing system has a 60.0 g capacity and a 95 mm diameter pan, with a minimum display of 0.001 g for mass and 0.01 % for moisture ratio. The unit measures 202 mm (W) × 336 mm (D) × 157 mm (H) and weighs 4.2 kg. It is positioned as a versatile model within Shimadzu's moisture analyzer line, offering multiple drying modes and data storage capabilities. ### Drying Behavior and Modes

The analyzer supports four drying modes: standard, rapid, slow, and step (3-stage) drying, each with automatic or timed ending. Time settings range from 1 to 240 minutes or continuous operation up to 12 hours. Repeatability is specified as 0.15 % (2 g), 0.05 % (5 g), and 0.02 % (10 g), indicating improved precision with larger sample sizes. The halogen lamp provides efficient heating, and the straight-tube design ensures uniform radiation across the sample. ### Installation and Operation

The MOC63u operates on 430 VA power and is intended for indoor laboratory use at ambient temperatures of 5 to 40 °C and relative humidity up to 85 %. It includes a backlit LCD for readout and stores up to 10 sets of measurement conditions and 100 data items. External interfaces include USB for data I/O and printer output (EP-100/EP-110) and RS-232C (D-sub 9P) for connection to a PC or peripheral. Standard accessories include sample pans, pan holder, windbreak, aluminum sheets, pan handler, power cable, spare fuses, protective display cover, and a hexagonal wrench.

Integration & Operating Notes

Mounting
Place on a stable, level surface free from vibration

The instrument should be installed on a sturdy bench that minimizes vibration and drafts, as airflow and mechanical disturbance can affect weighing accuracy. The built-in windbreak helps isolate the sample from ambient air currents, but the location should still avoid direct sunlight and heat sources.

Instruction Manual, Section 3 Installation
Electrical
Do not exceed 430 VA supply capacity

The MOC63u draws up to 430 VA during heating. Ensure the power outlet and any extension cords are rated for at least this apparent power to prevent overload or voltage drop, which could cause malfunction or reset during a drying cycle.

Factory Specification Sheet
Process Limits
Sample size affects repeatability and drying time

Repeatability is specified for sample masses of 2 g, 5 g, and 10 g, with better precision at higher masses. However, larger samples may require longer drying times. The user must balance sample size against required accuracy and time constraints, and ensure the sample fits within the 95 mm pan without overflowing.

Factory Specification Sheet
Control & I/O
External devices must be connected before starting a run

When using the USB or RS-232C interfaces for data logging or printer output, connect and configure the peripheral devices before initiating measurement. The instrument does not support hot-plugging during operation; doing so may cause data loss or communication errors.

Instruction Manual, Section 5 External Connections
Maintenance
Regular cleaning of the sample chamber and pan is required

After each use, allow the instrument to cool and clean the pan, pan holder, and chamber to prevent residue buildup that could affect future measurements. The halogen lamp and temperature sensor should be inspected periodically for contamination or damage; refer to the manual for replacement procedures.

Instruction Manual, Section 6 Maintenance

Selection & Integration Q&A

When should I choose the MOC63u over other moisture analyzers in the Shimadzu series?

Choose the MOC63u when you need a 60 g capacity with 0.001 g readability and multiple drying modes (standard, rapid, slow, step) plus data storage for 10 sets of conditions. It suits labs requiring flexible method development and USB/RS-232C connectivity. For simpler needs or different capacity, consider other models.

What are the requirements for connecting a printer or PC to the MOC63u?

The MOC63u provides a USB data I/O port for Shimadzu printers EP-100/EP-110 and an RS-232C (D-sub 9P) port for PC connection. Use appropriate cables (not included) and ensure the peripheral is compatible with the output format. Configure communication settings in the instrument menu before use.

What environmental conditions must be maintained for accurate measurements?

Operate the MOC63u at ambient temperatures of 5 to 40 °C and relative humidity up to 85 %. Avoid direct sunlight, drafts, and vibration. The built-in windbreak helps, but the location should be stable and free from rapid temperature fluctuations to ensure repeatable results.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
c10g-e105
Datasheet ·3,5 МБ ·24 pp.
View PDF
PDF
qn5042000001b2db
Datasheet ·1,4 МБ ·8 pp.
View PDF
PDF
Multi-Balance Collect PC connection simple manual
Manual ·1,7 МБ ·18 pp.
View PDF