Waters Xevo TQ-S micro
Waters ·Made in USA · In Active Production

Waters Xevo TQ-S micro

Waters Xevo TQ-S micro — Mass spectrometers.

The Xevo TQ-S micro is a benchtop tandem quadrupole mass spectrometer with two high-resolution quadrupole analyzers and a T-Wave collision cell, covering 2–2048 m/z. It is distinguished within the series by its compact 60×35.6×93 cm frame and standard tool-free ESI/APCI/ESCi multi-mode source. It serves as the MS/MS detector in quantitative LC-MS workflows, delivering up to 555 MRM transitions per second.

Direct drop-in replacements

Shimadzu
Maldimini 1
94%
match
⚖️ Compare
Shimadzu
Nexera Mikros
52%
match
⚖️ Compare
Shimadzu
MALDI 8020
33%
match
⚖️ Compare
Shimadzu
Axima Assurance
0%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Detector types Low noise, off axis, long life photomultiplier detector
Ion source High performance ZSpray dual-orthogonal API sources: Multi-mode source – tool-free ESI/APCI/ESCi (standard); UniSpray ion source (optional); Tool-free APCI probe (optional); nanoFlow ESI source (optional); ASAP (optional); APGC ion source (optional); ionK
Collision cell mfr T-Wave enabled
Vacuum system mfr One split-flow air-cooled vacuum turbomolecular
Regulatory approvals/marks mfr CE, CB, NRTL (CAN/US), RCM
MRM speed mfr 555 MRM/sec
Software mfr waters_connect and MassLynx version 4.2 or later
Dimensions, Weight & Mechanical Design
Mass range 2 to 2048 m/z
Height 60.0 cm
Length 93 cm
Width 35.6 cm
Depth mfr 93 cm (36.6 in)
Mass analyzer mfr Two high resolution, high stability quadrupole a

Description

Construction and positioning

The Xevo TQ-S micro is a tandem quadrupole (MS/MS) instrument built around two high-resolution, high-stability quadrupole analyzers (MS1 and MS2) with pre-filters and a T-Wave-enabled collision cell. The vacuum system uses one split-flow air-cooled turbomolecular pump backed by one vacuum backing pump. Its 60 cm height, 35.6 cm width and 93 cm depth make it the compact member of the TQ-S family; the standard ion source is a tool-free ZSpray dual-orthogonal multi-mode ESI/APCI/ESCi source, with UniSpray, APCI probe, nanoFlow ESI, ASAP, APGC and ionKey sources listed as optional. Detection is by a low-noise, off-axis, long-life photomultiplier. ### Behavior under load

Integration & Operating Notes

Control & I/O
MassLynx 4.2 or waters_connect required before first acquisition

The instrument will not operate with earlier software revisions. The workstation must have MassLynx version 4.2 or later, or waters_connect, installed and licensed before the system can be tuned or run; plan the software installation and PC specification as part of the instrument order, not after delivery.

Factory Specification Sheet, Software
Mounting
Bench must accommodate 93 cm depth and air-cooled vacuum pump

The 60 × 35.6 × 93 cm frame plus the split-flow air-cooled turbomolecular pump and separate backing pump require a bench with rear clearance for pump exhaust and airflow. Restricting rear ventilation raises pump temperature and can trigger vacuum faults; allow free space behind the instrument and route the backing pump exhaust away from the source intake.

Factory Specification Sheet, Dimensions and Vacuum System
Process Limits
Standard source covers ESI, APCI and ESCi without tools

The tool-free ZSpray dual-orthogonal multi-mode source supports ESI, APCI and combined ESCi operation. UniSpray, APCI probe, nanoFlow ESI, ASAP, APGC and ionKey are optional and must be ordered separately; confirm the ionization mode required by the application before specifying the configuration, because the optional sources are not part of the base delivery.

Factory Specification Sheet, Ion Source
Maintenance
Source and pre-filter cleaning interval depends on matrix load

The tool-free source design allows the probe and source housing to be removed for cleaning without tools. Pre-filter and source cleaning frequency is set by sample matrix and throughput; high-matrix samples require more frequent cleaning to preserve sensitivity. Establish a cleaning schedule from observed response drop rather than a fixed calendar interval.

Factory Specification Sheet, Ion Source and Mass Analyzer

Selection & Integration Q&A

When should I choose the Xevo TQ-S micro rather than a larger TQ-S series model?

Choose the micro when bench space is the constraint: 60 × 35.6 × 93 cm and 555 MRM/s cover routine quantitative LC-MS/MS. Larger TQ-S variants are warranted only when higher sensitivity or source options beyond the standard tool-free ESI/APCI/ESCi are required.

What must be in place on the workstation and bench before the instrument can run?

MassLynx 4.2 or later, or waters_connect, must be installed and licensed on the control PC. The bench must provide 93 cm depth plus rear clearance for the air-cooled split-flow turbomolecular pump and a location for the separate backing pump.

Can I switch between ESI and APCI on the same instrument without tools?

Yes. The standard ZSpray dual-orthogonal multi-mode source is tool-free for ESI, APCI and ESCi. UniSpray, nanoFlow ESI, ASAP, APGC and ionKey are optional sources ordered separately and are not included in the base configuration.

What maintenance and accessories are not included in the base delivery?

Optional ion sources (UniSpray, APCI probe, nanoFlow ESI, ASAP, APGC, ionKey) and the control workstation with MassLynx 4.2 or waters_connect are not part of the base instrument. Source and pre-filter cleaning consumables must be planned separately.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve

Click drawing to inspect vector dimensions

Technical documentation